Measurement uncertainty in clinical chemistry: ISO 20914 versus nordtest or intermediate precision versus bias

ORCID Icon, ORCID Icon & ORCID Icon
Received 08 Dec 2023, Accepted 01 Apr 2024, Published online: 14 May 2024
 

Abstract

Aim

Measuring uncertainty (MU) is crucial to ensure the accuracy and precision of laboratory results. This study compares the ISO 20914 and Nordtest guidelines to analyze the MU values for 20 clinical chemistry analytes over six months.

Methods

The researchers calculated MU components, including within-laboratory reproducibility (Rw), laboratory analytical performance bias (u(bias)), and combined standard uncertainty (uc), based on internal quality control and external quality assessment data. The final expanded uncertainty (U) values were determined by multiplying the combined uncertainty with a coverage factor (k = 2 for 95% Confidence Interval), following each guideline’s respective procedures. Clinical chemistry analytes were analyzed on Roche Cobas 6000 c501 auto analyzer (Roche Diagnostics, Mannheim, Germany) and manufacturer’s kits were used analysis.

Results

The results show that 11 out of 20 clinical chemistry analytes met the targeted maximum allowable measurement uncertainty (MAU) values when calculated according to ISO 20914 guideline. Also, 11 out of 20 clinical chemistry analytes’ MU values met the MAU values with the Nordtest guideline’s recommended calculations. However, some tests met the MAU in the ISO 20914 approach but not in the Nordtest guideline, and vice versa.

Conclusions

The study found that intermediate precision (uRw) in the ISO 20914 approach and performance bias (u(bias)) in the Nordtest approach significantly impacted MU values. The research highlights the importance of standardization in MU calculation approaches across clinical laboratories. These findings have implications for patient care and clinical decision-making, emphasizing the importance of selecting appropriate laboratory guidelines for routine use.

Disclosure statement

No potential conflict of interest was reported by the author(s).

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 200.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.