34
Views
16
CrossRef citations to date
0
Altmetric
ORIGINAL ARTICLE

Experimental study of an adjustable-length titanium ossicular prosthesis in a temporal bone model

, &
Pages 33-37 | Received 08 Jan 2004, Accepted 18 Mar 2004, Published online: 08 Jul 2009
 

Abstract

Conclusions This prosthesis has the advantage of rapid adjustment at the time of insertion in order to achieve optimal tension and, as a result, optimal sound transmission.

Objective To test the acoustic performance of a new, adjustable incus replacement prosthesis in a human temporal bone model.

Material and methods Experiments were performed in seven human temporal bones, before and after removal of the incus and insertion of the prosthesis. The input comprised 406 pure tones ranging in frequency between 0.1 and 10 kHz at an intensity of 80 dB SPL at the tympanic membrane. The output measurement was stapes footplate displacement, determined by means of a laser Doppler vibrometer. Three lengths of the prosthesis were investigated: optimal, optimal +0.2 mm and optimal +0.4 mm.

Results The optimal-length prosthesis produced similar results to those of an intact middle ear. The slightly longer prostheses decreased middle ear sound transmission at all test frequencies, except those near 1.5 kHz.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.