Use of recombinant human cardiac Troponin T for standardization of third generation Troponin T methods

, &
Pages 128-131 | Published online: 17 Mar 2010
 

Abstract

The Elecsys® Troponin T third generation assay uses recombinant human cardiac troponin T as standard material. The new assay has a linear calibration curve. Thus, linearity problems observed with the second generation assay have been eliminated. The assay has high precision, especially at the low end of measuring range (inter-assay CV <10 % at 0.1 μg/L). The new standardization does not change the cut-off value of 0.1 μg/L. The use of recombinant human cardiac troponin T as standard material enables a reproducible and reliable standardization of troponin T assays.