Radiofrequency ablation with a high-power generator: Device efficacy in an in vivo porcine liver model

, PhD, , , , &
Pages 387-394 | Received 07 Dec 2006, Accepted 15 Apr 2007, Published online: 09 Jul 2009
 

Abstract

Purpose: The purpose of this study was to test the feasibility and efficacy of using a high-power generator with nondeployable electrodes to create large zones of coagulation in an in vivo porcine liver model.

Methods: With approval from our institution's research animal care and use committee, 12 female swine (mean weight = 55 kg) were anesthetized and received RF ablation at laparotomy. Twenty-nine ablations were performed in four groups using: (i) a conventional 200-W generator and cluster electrode (n = 4), or an experimental prototype 250-W generator and (ii) a single, 17-gauge electrode (n = 9), (iii) a cluster electrode (n = 8) or (iv) three electrodes spaced 2.0 cm apart in a triangular configuration (n = 8). In the three-electrode group, power was applied by switching between electrodes using a prototype switching device. All electrodes were internally cooled. Ablation zone size, shape and generator data from each group were compared using a mixed-linear model with animals modeled as random effects.

Results: The high-power generator was able to increase significantly the zone of coagulation. Mean (±SD) ablation diameter was largest in the switched group (4.31 ± 0.7 cm) followed by the cluster (3.98 ± 0.5 cm) and single-electrode (3.26 ± 0.5 cm) groups. Mean diameter in the high-power single-electrode group was no different than the low-power cluster group (3.25 ± 0.4 cm, p = 0.98). Circularity measures were high (>0.75) in all groups.

Conclusions: Coupling a high-power generator and switching device is feasible. At higher powers, the switching device creates larger zones of ablation than cluster or single electrodes. Single-electrode ablations created with the prototype high-power generator were equivalent to those produced with the cluster electrode at conventional lower powers.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.