367
Views
13
CrossRef citations to date
0
Altmetric
Cornea

Comparison of Keratometry Obtained by a Swept Source OCT-Based Biometer with a Standard Optical Biometer and Scheimpflug Imaging

ORCID Icon, , &
Pages 882-888 | Received 02 Oct 2017, Accepted 25 Mar 2018, Published online: 09 Apr 2018
 

ABSTRACT

Purpose: To assess agreement of a swept source-optical coherence tomography (SS-OCT) based Biometer with a standard IOLMaster device and Scheimpflug Imaging (SI) to acquire keratometric measurements in cataract patients.

Methods: In this prospective comparative study, 101 eyes of 101 cataract surgery candidates, aged 24–81 years, were sequentially examined using three devices. Keratometry values at the flat (K1) and steep (K2) axis, mean corneal power (Km) and magnitude of corneal astigmatism as well as J0 and J45 vectoral components of astigmatism obtained with the SS-OCT based biometer (IOLMaster 700) were compared with those obtained with the IOLMaster 500 and SI. The agreement between measurements was evaluated by the Bland–Altman method, intraclass correlation coefficients (ICCs) and repeated-measures analysis of variance.

Results: Mean K1 values from the three devices were similar (p = 0.09). Mean K2 and Km values of IOLMaster 700 were higher than SI and lower than IOLMaster 500 (p = 0.04 for K2 and p = 0.02 for Km). There was a strong correlation between K1, K2, Km and magnitude of astigmatism obtained with all devices (r ≥ 0.80 and p < 0.01). The 95% limits of agreement (LoA) width for each keratometric value were highest for the comparison between IOLMaster 500 and SI and lowest for the comparison between IOLMaster 700 and 500. The mean differences (width of 95% LoA) for J0 and J45 vectoral components were 0.005 (2.19) and 0.12 (2.92) for the measurements obtained by IOLMaster 700 vs IOLMaster 500 and 0.06 (1.79) and 0.02 (1.58) for the measurements obtained by IOLMaster 700 vs SI, respectively.

Conclusions: With ICCs close to 1, the agreement between all devices was excellent for keratometric measurements. Mean K2, Km and astigmatism measurements from IOLMaster 700 were lower than IOLMaster 500 and higher than SI. However, the differences were quite small and are not expected to affect the final IOL power.

Acknowledgments

None.

Declaration of interest

The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper.

Additional information

Funding

No funding was received for this work.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.