Comparison of Pre- and Post-DMEK Keratometry and Total Keratometry Values for IOL Power Calculations in Eyes Undergoing Triple DMEK

, ORCID Icon, , , , ORCID Icon, ORCID Icon & ORCID Icon show all
Pages 477-486 | Received 24 Oct 2023, Accepted 05 Jan 2024, Published online: 22 Jan 2024
 

Abstract

Purpose

To evaluate prediction accuracy of pre- and post-DMEK keratometry (K) and total keratometry (TK) values for IOL power calculations in Fuchs endothelial corneal dystrophy (FECD) eyes undergoing DMEK with cataract surgery (triple DMEK).

Methods

Retrospective cross-sectional multicenter study of 55 FECD eyes (44 patients) that underwent triple DMEK between 2019 and 2022 between two centers in USA and Europe. Swept-source optical coherence tomography biometry (IOLMaster 700) was used for pre- and post-DMEK measurements. K and TK values were used for power calculations with ten formulae (Barrett Universal II (BUII), Castrop, Cooke K6, EVO 2.0, Haigis, Hoffer Q, Hoffer QST, Holladay I, Kane, and SRK/T). Mean error, mean absolute error (MAE), standard deviation, and percentage of eyes within ±0.50/±1.00 diopters (D) were calculated. Studied formulae were additionally adjusted using a method published previously (IOLup1D Method), which increases the IOL power by 1D. While both eyes from the same patient were considered for descriptive statistics, we restricted to one eye per individual (44 eyes for statistical comparisons.

Results

MAEs for all formulae were lower for post-DMEK K and TK than pre-DMEK K and TK by an average of 0.24 and 0.47 D, respectively. The lowest MAE was 0.49 D for Kane using post-DMEK TK, and the highest MAE was 1.05 D for BUII using pre-DMEK TK. Most IOLup1D formulae had lower MAEs than pre-DMEK K and TK formulae.

Conclusions

The IOLup1D Method should be used instead of pre-DMEK K and TK values for triple DMEK in FECD eyes. Using post-DMEK TK values for cataract surgery after DMEK provides better refractive accuracy than any of the three studied methods used for triple DMEK procedures.

Disclosure statement

No potential conflict of interest was reported by the author(s).

Data availability statement

The data that support the findings of this study are available from the corresponding author, KMR, upon reasonable request.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.