Volumetric Coronary Angiography Using the 256-Detector Row Computed Tomography Scanner: Comparison in vivo and in vitro with Porcine Models

, , , , &
Pages 186-191 | Accepted 13 Oct 2005, Published online: 09 Jul 2009
 

Abstract

Purpose: To investigate the feasibility of volumetric cine imaging in human cardiac studies by comparing in vivo and in vitro coronary angiography using a 256-detector row computed tomography (CT) without ECG gating.

Material and Methods: The left and right coronary arteries of two domestic pigs were scanned in vivo and in vitro in cine mode using the 256-detector row CT. The device scanned approximately 100 mm in the cranio-caudal direction with one rotation, with a slice thickness of 0.5 mm.

Results: The coronary arteries could be observed to the third-degree branches in vitro, but could be visualized clearly only to the proximal portion (first-degree or second-degree branches) in vivo.

Conclusion: Application of cardiac volumetric cine imaging with 256-detector row CT may be a promising means of obtaining diagnostic information and has potential for adoption to human studies.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.