The validation of smartphone applications for heart rate measurement

, ORCID Icon, ORCID Icon & ORCID Icon
Pages 721-727 | Received 25 Jul 2018, Accepted 27 Sep 2018, Published online: 17 Nov 2018
 

Abstract

Background: The smartphone apps provide a user-friendly option for measurement of heart rate (HR) by detecting pulsatile photoplethysmographic signals with built-in cameras from the fingertips, however, the validation study is limited.

Methods: We compared HR detected by the smartphone apps (App1 = Instant HR, App2 = Cardiio: HR Monitor and App3 = Runtastic HR Monitor) with simultaneous standard ECG monitoring in the adult patients at the critical care unit.

Results: HR measurements were obtained from 140 patients with mean age 67.6 ± 15.3 years. Mean baseline HR was 89.1 ± 19.1 bpm (range, 32–136 bpm). Sinus rhythm was presented in 111 patients (79.3%), atrial fibrillation in 25 patients (17.9%), pacemaker rhythm in 3 patients (2.1%), and high-grade AV block in 1 patient (0.7%). The ECG-derived HR correlated well with App1 (r = 0.98), App2 (r = 0.97), and App3 (r = 0.92). In patients with regular rhythm, mean absolute deviation was 0.8 ± 1, 0.7 ± 0.9, 1.0 ± 1.3 bpm on App1, App2 and App3, respectively. In the patients with irregular rhythm, median absolute deviation (IQR) was 3 (2–5.5), 4 (1.5–11.5), and 6 (2–13) bpm. Skin colour did not affect with the HR measurement.

Conclusions: HR measurements from all applications were correlated well with ECG monitoring. However, it was less accurate in case of irregular rhythm such as atrial fibrillation.

    Key messages

  • Several reports on inaccuracy of mobile health apps have been published. We conducted the validation study in the real patients by using popular mobile apps.

  • Heart rate measurements from mobile apps were correlated well with standard ECG. The accuracy of HR from apps was worse at irregular rate and tachycardia.

Acknowledgements

We thank Umaporn Udomtrupayakul, PhD for her assistance in statistical analysis.

Disclosure statement

No potential conflict of interest was reported by the authors.

Data availability

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.