Fetal transabdominal pulse oximeter studies using a hypoxic sheep model

, , , , &
Pages 393-399 | Published online: 07 Jul 2009
 

Abstract

Objective. This study investigates the validity of transabdominal pulse oximetry using a sheep fetal hypoxia model with fetal arterial hemoglobin saturation.

Methods. Four pregnant ewes were anaesthetized and cannulated through the brachial artery to measure direct arterial blood saturation, SaO2. Next, the transabdominal pulse oximeter was used to measure indirect measurement of the arterial saturation of the fetus, SpO2, from the maternal abdomen. Hypoxia was induced by a balloon placed in the maternal aorta.

Results. There is a linear relationship between SaO2, arterial blood saturation values of the fetus, and SpO2, the values measured by the transabdominal pulse oximetry with a slope of 0.75 (r2 = 0.76).

Conclusion. This information can be used to calibrate the transabdominal pulse oximeter as a measurement of fetal arterial saturation. With these results, we can advance the accurate, no-risk, noninvasive transabdominal fetal pulse oximeter for human use. This research may contribute to the more accurate diagnosis of the diseases of the fetus including Hypoxic Ischemic Encephalopathy.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.