Performance of blood glucose meters in compliance with current and future clinical ISO15197 accuracy criteria

, , , , &
Pages 185-190 | Accepted 21 Oct 2013, Published online: 06 Nov 2013
 

Abstract

Since 2003, blood glucose meters for patient self testing are approved in Europe based on the accuracy performance criteria as defined by the ISO15197 guideline. A new draft ISO guideline is currently under regulatory review, which suggests more strict accuracy acceptance criteria, and which may not be entirely fulfilled by currently commercialized blood glucose meter systems. In order to investigate the compliance of BG*Star and iBG*Star and several other blood glucose meters with the new draft ISO guideline, we performed a post-hoc analysis of data obtained from a recently performed ISO-conforming clinical accuracy performance study. This study was performed with 106 patients, clinically presenting with blood glucose levels distributed over the entire measurement range and in line with the glucose distribution requirements as demanded by the guideline. The YSI 2300 STAT Plus analyzer (glucose oxidase) served as reference method.

While all tested meters had been in a high degree of compliance with the current ISO criteria, performance was lower when analyzed in accordance with the new acceptance criteria (95% of readings have to be within ±15 mg/dL for values <100 mg/dL, and within ±15% for values ≥100 mg/dL). The following meters met the new criteria: Accu-Chek Aviva (95.5%/98.6%), BG*Star (98.5%/97.3%), iBG*Star (98.5%/97.3%), FreeStyle Freedom Lite (95.5%/96.6%), and OneTouch Ultra2 (95.5%/96.5%). One meter failed with low blood glucose values (Contour: 90.9%/95.9%).

In conclusion, BG*Star and iBG*Star and several other branded meters met the new draft ISO15197 acceptance criteria, when tested in accordance with the instructions for use and with the ISO accuracy testing protocol in a clinical setting.

Transparency

Declaration of funding

The study was funded by Sanofi Aventis.

Declaration of financial/other relationships

AP, TF, and PBM have received consultancy fees, research grants and travel expenses from Sanofi Aventis. CH, FD, CS have no conflicts of interest.

CMRO peer reviewers may have received honoraria for their review work. The peer reviewers on this manuscript have disclosed that they have no relevant financial relationships.

Acknowledgements

The authors received medical writing support from Julia Heise, IKFE-CRO GmbH, funded by Sanofi Aventis.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.