A Signal Processor for a Multiple-electrode Hearing Prosthesis

, , , , &
Pages 135-139 | Published online: 08 Jul 2009
 

Abstract

A 22-electrode implantable hearing prosthesis uses a wearable speech processor which estimates three speech signal parameters. These are voice pitch, second formant frequency and flattened spectrum amplitude. The signal is monitored continuously for periodicity in the range 80400 Hz and, if this is present, stimulation occurs at the same rate. Otherwise, as in the case of unvoiced sounds, it occurs at the random rate of fluctuation of the signal envelope. The second formant is obtained by filtering to extract the dominant peak in the midband region and by continuous measurement of the zero crossing rate. The amplitude measured is that of the whole speech spectrum, pre-emphasized by differentiation. The values that are presented to the patient are the parameter estimates immediately prior to the stimulation pulse. Second formant frequency is coded by selection of an appropriate electrode in the cochlea and amplitude by a suitably controlled current. Automatic gain control is used to keep the dynamic range of the amplitude estimate within the 30 dB range of the circuitry.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.