174
Views
18
CrossRef citations to date
0
Altmetric
MIDDLE EAR

Intraoperative adjustments to optimize active middle ear implant performance

, , , , &
Pages 27-35 | Received 21 Apr 2010, Accepted 30 Jun 2010, Published online: 27 Sep 2010
 

Abstract

Conclusion: After initial contact of the active middle ear implant (AMEI) on the incus, significant increases in device performance can be achieved intraoperatively without affecting residual hearing by additional static loading of the incus with 62 μm (quarter turn) to 125 μm (half turn) increments via an adjustment screw. Objectives: To assess the performance gains of driving the incus with an AMEI under increasing static loads in cadaveric temporal bones. Methods: Incus drive efficacy was assessed using laser Doppler velocimetry measurements of stapes velocities over a frequency range of 0.25 to 8 kHz. Results were compared to stapes velocities following acoustic stimulation via insert earphone. Maximum equivalent ear canal sound pressure level (LEmax) and residual hearing loss after initial loading of the AMEI (first contact) were compared in each temporal bone. Additional increases in incus load were induced by turning an adjustment screw in quarter turn steps, corresponding to 62 μm increments per step. LEmaxand residual hearing loss were reassessed after each step. For each temporal bone, experiments were repeated for three different AMEIs. Results: On average across bones, incus stimulation upon initial contact produced an LEmaxof 125, 127, and 121 dB SPL and residual hearing losses of –2, –1, and –1 dB with respect to unloaded, unaided conditions for the three AMEIs, respectively. Across bones and transducers, increasing static transducer load by incrementing the AMEI up to 125 μm significantly improved performance without affecting residual hearing loss. Loading beyond 125 μm (half turn) did not improve performance but significantly increased residual hearing loss.

Acknowledgments

We would like to acknowledge the technical and material support provided by Laurent Badih, Brian Conn, and Jim Easter of Otologics LLC, Boulder, CO, USA. We appreciate the assistance of Dr. Michael Hall in constructing the custom experimental equipment (with support by NIH grant P30 NS041854-05). This study was supported in part by an Otologics LLC educational grant (H.A.J. and D.J.T.) and Department of Physiology and Biophysics funds (D.J.T.).

Declaration of interest: Partial funding for these experiments was provided by an education grant from Otologics LLC (Boulder, CO, USA).

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.