62
Views
4
CrossRef citations to date
0
Altmetric
Review

Standardisation of cardiac troponin I measurement: past and present

, , , , , , , & show all
Pages 402-408 | Received 17 May 2010, Accepted 19 May 2010, Published online: 15 Jul 2010
 

Abstract

The laboratory measurement of cardiac troponin (cTn) concentration is a critical tool in the diagnosis of acute myocardial infarction (MI). Current cTnI assays produce different absolute troponin numbers and use different clinical cut-off values; hence cTnI values cannot be interchanged, with consequent confusion for clinicians. A recent Australian study compared patient results for seven cTnI assays and showed that between-method variation was approximately 2- to 5-fold.

A major reason for poor method agreement is the lack of a suitable common reference material for the calibration of cTnI assays by manufacturers. Purified complexed troponin material lacks adequate commutability for all assays; hence a serum-based secondary reference material is required for cTnI with value assignment by a higher order reference measurement procedure.

There is considerable debate about how best to achieve comparability of results for heterogeneous analytes such as cTnI, whether it should be via the harmonisation or the standardisation process. Whereas harmonisation depends upon consensus value assignment and uses those commercial methods which give the closest agreement at the time, standardisation comes closer to the true value through a reference measurement system that is based upon long-term calibration traceability.

The current paper describes standardisation efforts by the International Federation of Clinical Chemistry and Laboratory Medicine Working Group on Standardization of cTnI (IFCC WG-TNI) to establish a reference immunoassay measurement procedure for cTnI of a higher order than current commercial immunoassay methods and a commutable secondary reference material for cTnI to which companies can reference their calibration materials.

Acknowledgements

Thanks to Dr Graham Jones and Ms Janice Gill for permission to use the RCPA Quality Assurance Programs Pty Ltd troponin data.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.