Abstract
A method for determining serum creatinine avoiding protein precipitation is described. This method has proved well suited for automation. The creatinine is determined by its reaction with alkaline picrate at pH 10 and pH 12, whereupon a correction is made for a pseudocrcatinine derived from serum protein.