Efficient Use of Automatic Exposure Control Systems in Computed Tomography Requires Correct Patient Positioning

, , , &
Pages 1035-1041 | Published online: 28 Oct 2009
 

Abstract

Background: Image quality and radiation dose to the patient are important factors in computed tomography (CT). To provide constant image quality, tube current modulation (TCM) performed by automatic exposure control (AEC) adjusts the tube current to the patient's size and shape.

Purpose: To evaluate the effects of patient centering on tube current–time product (mAs) and image noise.

Material and Methods: An oval-shaped acrylic phantom was scanned in various off-center positions, at 30-mm intervals within a 500-mm field of view, using three different CT scanners. Acquisition parameters were similar to routine abdomen examinations at each site. The mAs was recorded and noise measured in the images. The correlation of mAs and noise with position was calculated using Pearson correlation.

Results: In all three scanners, the mAs delivered by the AEC changed with y-position of the phantom (P<0.001), with correlation values of 0.98 for scanners A and B and −0.98 for scanner C. With x-position, mAs changes were 4.9% or less. As the phantom moved into the y-positions, compared with the iso-center, the mAs varied by up to +70%, −34%, and +56% in scanners A, B, and C, respectively. For scanners A and B, noise in two regions of interest in the lower part of the phantom decreased with elevation, with correlation factors from −0.95 to −0.86 (P<0.02). In the x-direction, significant noise relationships (P<0.005) were only seen in scanner A.

Conclusion: This study demonstrates that patient centering markedly affects the efficacy of AEC function and that tube current changes vary between scanners. Tube position when acquiring the scout projection radiograph is decisive for the direction of the mAs change. Off-center patient positions cause errors in tube current modulation that can outweigh the dose reduction gained by AEC use, and image quality is affected.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.