The IAEA/WHO TL Dosimetry Service for Radiotherapy Centres 1969–1987

, &
Pages 461-467 | Accepted 31 May 1989, Published online: 08 Jul 2009
 

Abstract

The International Atomic Energy Agency (IAEA) started a postal dosimetry service in 1966 using lithium fluoride (LiF) thermoluminescent dosimeters (TLD). The World Health Organization (WHO) joined the programme in 1968. Dosimeters were sent on 29 occasions to primary standards laboratories or to the ‘Bureau International des Poids et Mesures' to guarantee the traceability to primary standards of absorbed dose. The mean deviation was −0.26 ± 0.18% (1% SD of the mean). One SD for a single result was about 1%. The number of intercomparisons made during the whole period is about 2000 from about 700 radiotherapy centres in 89 countries. To date only 60Co machines have been included. The mean deviation of all the results from the centres was −0.25, and 1% SD for a single result 6.7%. About 5% of the centres had deviations larger than 30%, at least on one occasion. It is shown that the accuracy in dosimetry improves considerably for those departments participating more than once in the TL dosimetry service.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.