Bacterial Contamination of Needles Used for Intravitreal Injections: Comparison between 27-gauge and 30-gauge Needles

, MD, , , MD, , MD, , MD & , MD
Pages 366-370 | Received 17 Oct 2012, Accepted 30 Apr 2013, Published online: 22 Jul 2013
 

Abstract

Purpose: To compare the contamination rate between 27-gauge and 30-guage needles used for intravitreal injection (IVT).

Methods: Patients undergoing IVT injections were enrolled prospectively. Injections were performed with 27- or 30-gauge needles. All needle tips were collected and placed in brain–heart infusion broth. The contamination rates of needles were compared.

Results: A total of 109 patients participated in the study and a total of 126 IVT injections were performed. Injections were performed by 27-gauge (49%) and 30-gauge (51%) needle. No patient developed endophthalmitis. The overall contamination rate of the used needles were 13% for 27-guage and 29% for 30-guage (p = 0.022). However, this difference was nonsignificant after Bonferroni's correction was applied. The most common bacteria isolated from the used needles are coagulase-negative Staphylococcus (CNS).

Conclusion: The results suggest that the needle bore size seems not to be a risk factor for contamination during IVT injection.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.