Large bore transhepatic tract dilatation in pigs: Results and implications for human procedures

, , , , &
Pages 5-11 | Published online: 10 Jul 2009
 

Summary

Large bore transhepatic tracts were created in six anesthetised pigs. A peripheral tract (entering the gall-bladder bed) and a central tract were created in each pig. After balloon dilatation to 30 F and passage of a 30 F inner diameter sheath, bleeding was monitored with the sheath in place and after sheath removal. Blood loss was quantified and tracts were examined microscopically. Peripheral tracts were shorter than central tracts (mean 1.9 versus 3.6 cm). Bleeding occurred more frequently and to a greater degree from central tracts. Injury to small arteries and veins was identified microscopically in two tracts that bled heavily. Although significant bleeding, or lack thereof after dilatation of large transhepatic tracts, was variable, recommendations for large tract dilatation include: selection of a peripheral tract route, use of a working sheath with an outside diameter not exceeding that of the inflated balloon, particular care to avoid loss of access and not removing sheaths prematurely.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.