Accuracy of portable glucose meters for rapid determination of amniotic fluid glucose levels

, , &
Pages 183-189 | Received 05 Sep 1997, Accepted 24 Apr 1998, Published online: 07 Jul 2009
 

Abstract

This study was designed to evaluate the accuracy and feasibility of use of three commercially available portable blood glucose meters to measure amniotic fluid glucose(AFG) levels as compared to an accepted laboratory standard. A prospective study of amniotic fluid from 101 consecutive amniocenteses was performed. Glucose concentration in the amniotic fluid was assessed by hexokinase method in our hospital laboratory (control) and by using three portable meters: Advantage (ADV) (Boehringer Mannheim), Glucometer Elite (ELT) (Bayer), and One Touch II Hospital (T-2) (Lifescan). Twenty consecutive amniotic fluid samples were sent to the laboratory in two vials, the first without additive and the second with potassium oxalate to prevent metabolic activity, to assess the effect of cellular metabolism and time delay on amniotic fluid glucose concentrations. Data are reported as mean ±SE and were assessed by one-way ANOVA. Of the 101 patients studied, 29 were of gestational age <20 wks. The remaining 72 patients were < 20 wks. All three ambulatory meters demonstrated a linear relationship with control (all P < 0.001). Given a slope of almost 1.0 (m = 0.94) and a y-intercept approaching zero (b = 4.3), the OT2 proved to correlate best with control. ELT: (r2 = 0.55, m = 0.79, b = 22.2) and ADV: (r2 = 0.74, m = 1.45, b = 16.9) both overestimated amniotic fluid glucose. When AFG was < 30 mg/dl via laboratory standard, OT2: (r2 = 0.78, m = 1.05, and b = -2.20, P < 0.001), ADV: (m = 1.02, b = 24.1, r2 = 0.12, P = 0.133). The One Touch II Hospital accurately predicted amniotic fluid glucose at the bedside with excellent correlation including with laboratory standard glucose levels < 30 mg/dl. ADV and ELT proved too inaccurate for clinical use. Control samples were not affected by additives or time delay. These findings confirm that AFG determinations can be obtained rapidly with the OT2 meter at the bedside.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.