Tibiofemoral contact stress after total knee arthroplasty

Comparison of fixed and mobile-bearing inlay designs

, , &
Pages 638-646 | Received 29 Jul 2001, Accepted 14 Mar 2002, Published online: 22 Feb 2011
 

Abstract

We measured tibiofemoral contact stresses and the load-bearing contact area of fixed and mobile-bearing inlay knee prostheses under dynamic loading conditions. An electronic resistive pressure-measuring sensor was used to detect contact stresses and contact area in five cadaver knees. Stresses were measured with the tibial component aligned normally, as well as in internally- and externally-rotated positions. The average peak contact stresses measured on the fixed inlay were greater (medial 21 MPa and lateral 21 MPa) than those on the mobile inlay (medial/lateral 7.7/5.3 MPa, p = 0.04). Although the average peak contact stresses of the fixed standard inlay greatly exceeded the contact stresses of the other two inlay designs in each malrotated position tested, no statistically significant differences were seen. The data suggest that the ability of the inlay to translate on the tibial baseplate permits the inlay to align itself on the femoral component so that the contact surface area is maximized and contact stresses are reduced.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.