How precise is the identification of the center of the femoral head during total knee arthroplasty?

, , , &
Pages 53-58 | Received 02 May 2011, Accepted 05 Aug 2011, Published online: 23 Nov 2011
 

Abstract

Background and purpose Identification of the center of the femoral head in the coronal plane is essential during total knee arthroplasty. We evaluated a new method for localization of the center of the hip, thereby detecting the neutral mechanical axis using inter-femoral head center distances (X) measured from a radiograph. Our proposed method was compared with 3 commonly used methods using landmarks that are estimated to be 2 finger-breaths medial to the anterosuperior iliac spine (method A), 2.5 cm perpendicular to the mid-inguinal point (method I), and 1.5 cm lateral to the femoral artery (method F).

Methods 114 patients undergoing total knee arthroplasty were prospectively enrolled in the study. Four landmarks were marked and conventional anterior-posterior pelvic radiographs were taken. On the radiograph, the distance between the estimated FHC and the neutral mechanical axis was measured.

Results The median value (mm) of the measured distance was 9 in A, 7 in I, 8.5 in F, and 5 in X. When an error of more than 3° from neutral alignment was defined as an outlier, 15% of measurements in A, 6% of measurements in I, 14% in F, and 2% in X would fall in the outlier zone.

Interpretation The method detecting the neutral mechanical axis using inter-femoral head center distances (X) showed the least variability and the lowest percentage of outliers.

JGS: study design. HM: statistical analysis. YWM: evaluation of clinical relevance concerning methods. SHP: evaluation of clinical relevance concerning statistical analysis. SMK: preparation of manuscript.