Polyethylene wear with an entirely HA-coated total hip replacement 79 hips followed for 5 years

&
Pages 253-258 | Received 29 Sep 1997, Accepted 23 Feb 1998, Published online: 08 Jul 2009
 

Abstract

We measured the eccentricity of the femoral head in the metal backing annually during 5 years in 79 consecutive total hip replacements (73 patients). The mean age of the patients was 57 (32–73) years and the female/male ratio 63/22. The prostheses were entirely coated with hydroxyapatite. Modular 32 mm stainless steel heads and hemispherical, self-tapping screw cups with polyethylene liners were used. We found accelerating eccentricity throughout the observation period. The mean eccentricity at 5 years was 0.71 mm (95% CI 0.53–0.90), resulting in a mean eccentricity rate of 0.14 mm (0.11 -0.18) per year. The 5-year eccentricity was 0.5 mm or less in 45 hips and more than 1.5 mm in 13 hips, 2 in the latter group apparently having worn through the polyethylene liner. The true wear may be twice as great.

The use of 32 mm stainless steel heads and thin polyethylene inlays may have aggravated the wear problems. These HA-coated prostheses must all be checked regularly, so that cases with excessive polyethylene wear can be reoperated on before wear-through of the liner.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.