Difference between two exhaled nitric oxide analyzers, NIOX VERO® electrochemical hand-held analyzer and NOA280i® chemiluminescence stationary analyzer

, MD, , MD, PhD, , MD, PhD, , MD, , MD, , PhD, , BS, , BS, , MD, PhD, , MD, PhD, , MD, PhD, , MD, PhD & , MD, PhD show all
Pages 167-172 | Received 05 Oct 2017, Accepted 07 Feb 2018, Published online: 18 Oct 2018
 

Abstract

Background: Fractional exhaled nitric oxide (FENO) is useful for the evaluation of eosinophilic airway inflammation, including that seen in asthma. Although a new electrochemical hand-held FENO analyzer, the NIOX VERO® (Aerocrine AB, Solna, Sweden), is clinically convenient to use, it has not been fully compared with the chemiluminescence stationary electrochemical analyzer NOA280i® (Sievers Instruments, Boulder, CO, USA) in terms of the level of measured FENO. The aim of this study was to determine whether there is a difference between the two analyzers. Methods: The FENO levels measured with both NIOX VERO® and NOA280i® were evaluated in 1,369 adults at Juntendo University Hospital from May 2016 to October 2016. Results: The median FENO level measured with the NIOX VERO® was significantly lower than that measured with the NOA280i® (41 ppb, range 5–368 ppb vs. 29 ppb, range 5–251 ppb; p < 0.001). There was a strong positive correlation in the measurement of FENO level between the NOA280i® and the NIOX VERO® (r = 0.942, p < 0.001). The following conversion equation was calculated: FENO (NOA280i®) = 1.362 (SE, 0.661) + 1.384 (SE, 0.021) × FENO (NIOX VERO®). Conclusions: To our best knowledge, we have provided the first report showing that the measured FENO level with the NIOX VERO® was approximately 30% lower than that with the NOA280i® and that there was a significant correlation between the measurements of these two devices. The correction equation that we provided may help assess the data obtained by these two analyzers.

Abbreviations
ATS=

American Thoracic Society

BMI=

Body mass index

ERS=

European Respiratory Society

FENO=

Fractional exhaled nitric oxide

GINA=

Global Initiative for Asthma

NO=

Nitric oxide

ppb=

Parts per billion

ROC=

Receiver operating characteristic

SD=

Standard deviation

Declaration of interest

KT received honoraria from AstraZeneca K.K., Boehringer Ingelheim Japan, Inc., Eli Lilly Japan K.K., MSD K.K., and Pfizer Inc. JO is an employee of Shino-Test Corporation. RA received honoraria from Kyorin Pharmaceutical Co., Ltd. The rest of the authors have no conflict of interest.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 1,078.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.