Compensating for bone interfaces and respiratory motion in high-intensity focused ultrasound

, , , , &
Pages 141-151 | Received 10 Oct 2006, Accepted 10 Jan 2007, Published online: 09 Jul 2009

Figures & data

Figure 1. (a) A 300-element HIFU array with a water-cooling and coupling system and the 300 electronic channels driving the HIFU brain array. A curved array (128 elements–3 MHz) is embedded in the vicinity of the HIFU array in order to provide a low-quality echographic image of the skull surface and brain for a rough positioning of the complete system with respect to the head. (b) Quasi Random distribution of the 300 elements into a spherical mould. (c, d) 2D scan of the transcranial focal spot: Distribution of the acoustic energy deposition in the focal plane of the array (z = 140 mm depth) without (c) and with (d) time reversal correction of the skull distortions.

Figure 1. (a) A 300-element HIFU array with a water-cooling and coupling system and the 300 electronic channels driving the HIFU brain array. A curved array (128 elements–3 MHz) is embedded in the vicinity of the HIFU array in order to provide a low-quality echographic image of the skull surface and brain for a rough positioning of the complete system with respect to the head. (b) Quasi Random distribution of the 300 elements into a spherical mould. (c, d) 2D scan of the transcranial focal spot: Distribution of the acoustic energy deposition in the focal plane of the array (z = 140 mm depth) without (c) and with (d) time reversal correction of the skull distortions.

Figure 2. The different steps of the brain HIFU therapy guided by Ctscans: (a) the monkey is imaged into a CT scanner. The head is fixed on a house-made stereotactic frame. (b) Treatment software allows for the selection of the targeted spot. From these data, numerical simulation of the 3D transcranial ultrasonic propagation enables the estimation and correction of skull aberrations for the HIFU treatment. (c) One week after the CT scan exam, the monkey's head is again fixed into the house-made stereotactic frame. (d) The HIFU array is coupled to the monkey's head via the stereotactic frame, thus ensuring the concordance of CT scan reference frames and the HIFU experimental reference frame. The HIFU treatment is then performed using the calculated emission signals.

Figure 2. The different steps of the brain HIFU therapy guided by Ctscans: (a) the monkey is imaged into a CT scanner. The head is fixed on a house-made stereotactic frame. (b) Treatment software allows for the selection of the targeted spot. From these data, numerical simulation of the 3D transcranial ultrasonic propagation enables the estimation and correction of skull aberrations for the HIFU treatment. (c) One week after the CT scan exam, the monkey's head is again fixed into the house-made stereotactic frame. (d) The HIFU array is coupled to the monkey's head via the stereotactic frame, thus ensuring the concordance of CT scan reference frames and the HIFU experimental reference frame. The HIFU treatment is then performed using the calculated emission signals.

Figure 3. (a) Principle of the transcranial time reversal focusing technique guided by CT scans. (b) 3D simulation of the wave propagation (finite differences): wave field distribution at different time steps (at t = 0 µs, t = 8 µs, t = 16 µs, t = 24 µs, t = 32 µs, t = 40 µs, t = 48 µs, t = 56 µs, t = 64 µs) in a selected plane. (c) The monkey was sacrificed 1 week after treatment: the 7T MR images of the fixed brain clearly show the thermal necrosis at the targeted location using a steady state free precession (SSFP) sequence. The histological examination clearly shows thermal acidophilic necrosis in the area treated with time reversal correction and normal pattern (neurons, oligodendrocytes, capillaries) in the area treated with a non-corrected HIFU beam.

Figure 3. (a) Principle of the transcranial time reversal focusing technique guided by CT scans. (b) 3D simulation of the wave propagation (finite differences): wave field distribution at different time steps (at t = 0 µs, t = 8 µs, t = 16 µs, t = 24 µs, t = 32 µs, t = 40 µs, t = 48 µs, t = 56 µs, t = 64 µs) in a selected plane. (c) The monkey was sacrificed 1 week after treatment: the 7T MR images of the fixed brain clearly show the thermal necrosis at the targeted location using a steady state free precession (SSFP) sequence. The histological examination clearly shows thermal acidophilic necrosis in the area treated with time reversal correction and normal pattern (neurons, oligodendrocytes, capillaries) in the area treated with a non-corrected HIFU beam.

Figure 4. Spatial directivity pattern of the transcostal ultrasonic beam in the focal plane using (a) conventional focusing and (b) time reversal focusing. (c) Necrosis obtained using time reversal focusing through the rib cage in liver samples for a 5-s duration at 1000 W · cm−2 and for a 5-s duration at 1600 W · cm−2. (d) Temperature elevation at several locations on the rib cage with conventional and time reversal focusing. The temperature elevation is estimated using implanted thermocouples (Iron-constantan, 40 µm diameter, PhysiTemp Corp.).

Figure 4. Spatial directivity pattern of the transcostal ultrasonic beam in the focal plane using (a) conventional focusing and (b) time reversal focusing. (c) Necrosis obtained using time reversal focusing through the rib cage in liver samples for a 5-s duration at 1000 W · cm−2 and for a 5-s duration at 1600 W · cm−2. (d) Temperature elevation at several locations on the rib cage with conventional and time reversal focusing. The temperature elevation is estimated using implanted thermocouples (Iron-constantan, 40 µm diameter, PhysiTemp Corp.).

Figure 5. Basic principle of the real time motion correction. (a) Four subapertures successively send short ultrasonic impulses focusing at the tracking location. (b) Standard 1D cross-correlations between successive sets of acquisitions permit to recover the 3D displacement vector. (c) Displacement trajectory of a liver sample mounted on three-axis linear motors. Necrosis obtained in the moving liver sample (d) without and (e) with real time motion correction.

Figure 5. Basic principle of the real time motion correction. (a) Four subapertures successively send short ultrasonic impulses focusing at the tracking location. (b) Standard 1D cross-correlations between successive sets of acquisitions permit to recover the 3D displacement vector. (c) Displacement trajectory of a liver sample mounted on three-axis linear motors. Necrosis obtained in the moving liver sample (d) without and (e) with real time motion correction.

Figure 6. (a) In vivo investigation: the anesthetized pig is placed in front of the water cooling and coupling balloon of the 300-element system. (b) Real-time recording of the 3D respiratory motion using the HIFU array (red: superior-inferior motion, blue: lateral motion, green: anterior posterior motion).

Figure 6. (a) In vivo investigation: the anesthetized pig is placed in front of the water cooling and coupling balloon of the 300-element system. (b) Real-time recording of the 3D respiratory motion using the HIFU array (red: superior-inferior motion, blue: lateral motion, green: anterior posterior motion).

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.