TECHNICAL NOTE

Comparison of combined oximetry and cutaneous capnography using a digital sensor with arterial blood gas analysis

, , , , , & show all
Pages 60-64 | Received 03 Aug 2009, Accepted 12 Oct 2009, Published online: 03 Dec 2009
 

Abstract

Background: Cutaneous carbon dioxide tension (PcCO2) is a promising non-invasive surrogate measure of arterial partial pressure of carbon dioxide (PaCO2). Objectives: To compare values of PcCO2 and oxygen saturation (SpO2) with arterial blood gas (ABG) analysis. Methods: SpO2 and PcCO2 were measured with a v-Sign™-sensor (Sentec AG, Therwil, Switzerland) and the values compared with simultaneously obtained SaO2 and PaCO2 obtained from ABG analysis (ABL 725, Radiometer, Copenhagen, Denmark) in 275 adult patients referred to the lung function laboratory. Results: Median of the PcCO2 was 4.7 kPa (interquartile range [IQR] 0.9 kPa). Median of the SpO2 was 97% (IQR 3%). Bland-Altman analysis for comparison of PcCO2 with PaCO2 showed a bias of −0.1 kPa with a precision of ± 0.9 kPa with 3.7% outlying values. Bland-Altman analysis for the comparison of SpO2 and SaO2 showed a bias of 20.1 % with a precision of ± 3.5%. There were no complications. Conclusion: There is a good agreement between combined cutaneous capnography and oximetry values with ABG analysis. Due to the excellent safety profile and the short time to get a continuous measurement, this technique should be examined in settings where it can complement repeated ABG analysis when ventilatory disturbances are suspected or non-invasive monitoring of ventilation is needed.

Declaration of interest: Prashant N. Chhajed is on the medical advisory board of Sentec AG.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 200.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.