358
Views
193
CrossRef citations to date
0
Altmetric
Original Paper: Otoneurology

Conservation of low-frequency hearing in cochlear implantation

, , , , , & show all
Pages 272-280 | Received 05 Feb 2003, Accepted 24 Apr 2004, Published online: 17 Nov 2014
 

Abstract

Objective As results with cochlear implants have continued to improve, patients with some remaining cochlear function have become eligible for cochlear implantation. Thus, preservation of acoustic hearing after implantation has gained importance. Hearing preservation can be considered a benchmark for atraumatic implantation preventing neural degeneration from loss of residual hair cells or subsequent to local trauma. In this prospective study, the possibility of preserving low-frequency hearing in cochlear implantation using a modified surgical technique has been explored.

Materials and Methods In a prospective study design, 14 subjects with considerable low-frequency hearing of 2–60 dB in the frequency range 125–500 Hz but with unsatisfactory speech understanding with hearing aids of < 35% monosyllabic word understanding were implanted with a MED-EL COMBI-40+ cochlear implant. The insertion depth was intentionally limited to 19–24 mm to prevent damage to low-frequency regions of the cochlea. Pre- and postoperative pure-tone thresholds were measured.

Results Hearing was conserved within 0-10 dB in 9/14 subjects and within 11–20 dB in 3/14; in 2/14 subjects hearing was completely lost in the implanted ear. Thus hearing could at least partially be conserved in 12/14 subjects (86%). Median threshold values decreased by 10, 15, 17.5 and 5 dB at 125, 250, 500 and 1000 Hz, respectively. Even high levels of hearing, e.g. 30 dB at 500 Hz, could be maintained after implantation in some subjects.

Conclusions This study reports successful conservation of hearing after cochlear implantation using a modified surgical technique. Even high levels of hearing could be maintained, showing that implantation of an intracochlear electrode can be performed atraumatically with preservation of functional structures.

Kiefer J, Gstoettner W, Baumgartner W, Pok SM, Tillien J, Ye Q, von Ilberg C. Conservation of low-frequency hearing in cochlear implantation. Acta Otolaryngol 2004; 124: 272–280.

Kiefer J, Gstoettner W, Baumgartner W, Pok SM, Tillien J, Ye Q, von Ilberg C. Conservation of low-frequency hearing in cochlear implantation. Acta Otolaryngol 2004; 124: 272–280.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.