Time to think outside the box? Prothrombin time, international normalised ratio, international sensitivity index, mean normal prothrombin time and measurement of uncertainty: a novel approach to standardisation

, , , &
Pages 277-287 | Received 27 Dec 2006, Accepted 23 May 2007, Published online: 06 Jul 2009
 

Summary

Background: The prothrombin time (PT) assay is the most clinically ordered coagulation test, and most often used for monitoring of vitamin K antagonist therapy (e.g., warfarin), where results are expressed as an international normalised ratio (INR). The INR is in essence the patient's PT ‘mathematically adjusted’ to a standardised value taking into account the peculiarities of the test system as defined by an ISI (international sensitivity index) and MNPT (mean normal prothrombin time). Although some manufacturers provide assigned ISI values for specific PT reagents and instrumentation, it is still recommended practice that laboratories check or validate these ISIs, as well as estimate the MNPT. Where an ISI is not provided by a manufacturer, the laboratory needs to estimate its own value. Current recommendations suggest the use of commercial reference-plasma calibration sets, but there is limited information on the performance of these in the field.

Results: We report a comparative study that assessed the utility of three such commercial calibration plasma sets, used as recommended, as well as alternate or supplementary procedures for estimation of ISI and MNPT. The latter included one novel approach using comparative data of ‘existing’ versus ‘replacement’ reagent, as well as assessment of external quality assurance data. Although MNPT value estimates were not grossly disparate, a wide variety of ISI values (e.g., 1.12–1.30 for our primary instrument) was obtained with the different plasma sets.

Conclusion: Because of the above, further verification checks are required prior to acceptance of ISI and MNPT estimates generated from commercial plasma calibration sets. We also provide some recommendations regarding the process of standardisation of INR testing.

Abbreviations
DB=

Dade-Behring

H=

Helena

INR=

international normalised ratio

ISI=

international sensitivity index

LT=

Life Therapeutics

MNPT=

mean normal prothrombin time

MU=

measurement of uncertainty

PT=

prothrombin time

VKA(s)=

vitamin K antagonist(s)

Abbreviations
DB=

Dade-Behring

H=

Helena

INR=

international normalised ratio

ISI=

international sensitivity index

LT=

Life Therapeutics

MNPT=

mean normal prothrombin time

MU=

measurement of uncertainty

PT=

prothrombin time

VKA(s)=

vitamin K antagonist(s)

Log in via your institution

Log in to Taylor & Francis Online

There are no offers available at the current time.

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.