A precise, sensitive and stable LC-MSMS method for detection of picomolar levels of serum aldosterone

&
Pages 379-385 | Received 15 Dec 2017, Accepted 20 May 2018, Published online: 09 Jul 2018
 

Abstract

Background: Accurate quantification of aldosterone is essential in the diagnosis of primary aldosteronism. Liquid chromatography-tandem mass spectrometry (LC-MSMS) analysis is increasingly being used to improve analytical sensitivity and specificity, since this technology reduces most of the interferences observed with immunological methods.

Methods: Serum samples with d7-aldosterone as internal standard were extracted with methyl tert-butyl ether, using liquid-liquid extraction (LLE). Chromatographic separation was performed on a C18 reverse phase column with a methanol-water gradient containing ammonium fluoride. Aldosterone detection was performed on an Agilent 6490 triple quadrupole using electro spray ionisation in positive mode.

Results: Multiple reaction monitoring transitions were m/z 361.2–315.1 for aldosterone, and 368.5–323.3 for d7-aldosterone. Chromatographic retention time was 2.7 min. The method’s total CVs at aldosterone concentrations of 45.4 and 1080 pmol/L were 7.0% and 4.8%, respectively. The intra-assay CVs at concentrations of 60.0 and 637 pmol/L were 4.0% and 2.6%, respectively. The method’s LOQ and LOD were 10 and 5 pmol/L, respectively, demonstrating an excellent analytical sensitivity. The upper limit of quantification was set to 5000 pmol/L, corresponding to the highest calibrator concentration. The long-term stability of the method was evident from repeated measurements of external control pools from UKNEQAS over a period of about 3 years, showing CVs between 2.0 and 7.0%.

Conclusions: We have described a precise, sensitive and stable LC-MSMS method for the measurement of serum aldosterone. In addition, due to the use of LLE and a short LC-column, the method is simple to perform, with a short chromatographic run time.

Disclosure statement

No potential conflict of interest was reported by the authors.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 200.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.