Technical Note

Filling accuracy and imprecision of commercial evacuated sodium citrate coagulation tubes

ORCID Icon, , , , , & show all
Pages 276-279 | Received 18 Jan 2019, Accepted 16 Apr 2019, Published online: 29 Apr 2019
 

Abstract

Current recommendations advocate that blood tubes for coagulation testing should be filled not less than 90% of their nominal filling volume, since under- or over-filling >10% may generate unreliable results of some hemostasis assays. This study was hence aimed to explore filling accuracy and precision of commercial blood tubes. Between-lot variations of 3 different lots (20 tubes per lot) of 3.2% citrate blood tubes manufactured by Becton Dickinson, Greiner and Kima were studied. One additional lot from each manufacturer was assessed in triplicate (three series of 20 tubes), to assess within-lot variation. All tubes were first weighed empty and then filled with distilled water by a syringe, under ideal filling conditions. Filled tubes were weighed again, in duplicate. For each 20 tubes series, mean bias (deviation from the ideal tube filling volume) and imprecision (coefficient of variation; CV%) were calculated. All biases were within ±10%. Within-lot and between-lot variation in filling volume was acceptable, and comprised between 0.4 and 2.4%. Greiner tubes were the most accurate (bias, −1.0 to 2.4%), followed by Kima (bias, −7.8 to −5.9%) and Becton Dickinson (bias, −9.6 to 3.3%) tubes. The highest between-lot difference was noted for Becton Dickinson tubes (up to 12.9%), followed by Greiner and Kima tubes (up to 3.4 and 1.8%, respectively). Although coagulation tubes filling accuracy was within ±10% for all three tested manufacturers, the overall bias was found to be variable among manufacturers and lots. Major effort shall be made by blood tube manufacturers for improving standardization of their products.

Acknowledgments

All blood tubes used in this experimental study were purchased from the manufacturers with proprietary funding. The companies were not aware of the purposes of this study and had no role in directing the study or influencing the outcomes.

Disclosure statement

No potential conflict of interest was reported by the authors.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 200.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.