Abstract
In order to conserve material that is available in limited quantities, “directed-flow” nose-only exposure systems have at times been run at flow rates close to the minute ventilation of the animal. Such low-flow-rate conditions can contribute to a decrease of test substance concentration in inhaled air; near the animal nose, exhaled air and the directed flow of exposure air move in opposite directions. With a Cannon “directed-flow” nose-only exposure system (Lab Products, Maywood, NJ), we investigated the extent to which exposure air plus exhaled air can be inhaled by an animal. A mathematical model and a mechanical simulation of respiration were adopted to predict for a male Fischer 344 rat the concentration of test substance in inhaled air. The mathematical model was based on the assumption of instantaneous mixing. The mechanical simulation of respiration used a Harvard respirator. When the system was operated at an exposure air flow rate greater than 2.5 times the minute ventilation of the animal, the concentration of test substance in the inhaled air was reduced by less than 10%. Under these conditions, the circular jet of air exiting the exposure air delivery tube tended to reach the animal's nose with little dispersion. For exposure air flow rates less than 2 times the minute ventilation, we predict that the interaction of exhaled air and exposure air can be minimized by proportionally reducing the delivery tube diameter. These findings should be applicable to similar “directed-flow” nose-only exposure systems.