ABSTRACT
Background: Dart-throwing motion (DTM) is an important component of wrist function and, consequently, has the potential to become an evaluation tool in rehabilitation. However, no measurement method is currently available to reliably measure range of motion (ROM) of the wrist in the DTM plane. Objectives: To determine the reliability and responsiveness of a goniometric device to measure wrist ROM in the DTM plane. Methods: ROM of the wrist in the DTM plane was measured in 70 healthy participants. The intra-class correlation coefficient (ICC) was used to evaluate the relative reliability of measurement, and a Bland–Altman analysis conducted to establish its absolute reliability, including the 95% limits of agreement (95% LOA). The standard error of the measurement (SEM) and minimal detectable change at the 95% confidence level (MDC95) were calculated as measures of responsiveness. Results: The intra-rater ICC was 0.87, and an inter-rater ICC of 0.71. There was no evidence of a fixed or proportional bias. For intra- and inter-rater reliability, 95% LOA ranged from −13.83 to 11.12 and from −17.75 to 16.19, respectively. The SEM and MDC95 were 4.5° and 12.4°, respectively, for intra-rater reliability, and 6.0° and 16.6°, respectively, for inter-rater reliability. Conclusion: The ROM of the wrist in the DTM plane was measured with fair-to-good reliability and responsiveness and, therefore, has the potential to become an evaluation tool for rehabilitation.