304
Views
0
CrossRef citations to date
0
Altmetric
Focus on Airway

A Practical Solution for Preoxygenation in the Prehospital Setting: A Nonrebreather Mask with Flush Rate Oxygen

, , , , , , , , , & show all
Pages 215-220 | Received 20 Mar 2023, Accepted 10 May 2023, Published online: 22 May 2023
 

Abstract

Objective

Prehospital clinicians need a practical means of providing adequate preoxygenation prior to intubation. A bag-valve-mask (BVM) can be used for preoxygenation in perfect conditions but is likely to fail in emergency settings. For this reason, many airway experts have moved away from using BVM for preoxygenation and instead suggest using a nonrebreather (NRB) mask with flush rate oxygen.

Literature on preoxygenation has suggested that a NRB mask delivering flush rate oxygen (on a 15 L/min O2 regulator, maximum flow, ∼50 L/min) is noninferior to BVM at 15 L/min held with a tight seal. However, in the prehospital setting, where emergency airway management success varies, preoxygenation techniques have not been deeply explored. Our study seeks to determine whether preoxygenation can be optimally performed with NRB at flush rate oxygen.

Methods

We performed a crossover trial using healthy volunteers. Subjects underwent 3-min trials of preoxygenation with NRB mask at 25 L/min oxygen delivered from a portable tank, NRB at flush rate oxygen from a portable tank, NRB with flush rate oxygen from an onboard ambulance tank, and BVM with flush rate oxygen from an onboard ambulance tank. The primary outcome was the fraction of expired oxygen (FeO2). We compared the FeO2 of the BVM-flush to other study groups, using a noninferiority margin of 10%.

Results

We enrolled 30 subjects. Mean FeO2 values for NRB-25, NRB-flush ambulance, NRB-flush portable, and BVM-flush were 63% (95% confidence interval [CI] 58–68%), 74% (95%, CI 70–78%), 78% (95%, CI 74–83%), and 80% (95%, CI 75–84%), respectively. FeO2 values for NRB-flush on both portable tank and ambulance oxygen were noninferior to BVM-flush on the ambulance oxygen system (FeO2 differences of 1%, 95% CI −3% to 6%; and 6%, 95% CI 1–10%). FeO2 for the NRB-25 group was inferior to BVM-flush (FeO2 difference 16%, 95% CI 12–21%).

Conclusions

Among healthy volunteers, flush rate preoxygenation using NRB masks is noninferior to BVM using either a portable oxygen tank or ambulance oxygen. This is significant because preoxygenation using NRB masks with flush rate oxygen presents a simpler alternative to the use of BVMs. Preoxygenation using NRB masks at 25 L/min from a portable tank is inferior to BVM at flush rate.

Disclosure Statement

None of the authors report any conflicts of interest.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 85.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.