Kiwi omnicup handheld versus Mityvac M-style conventional vacuum system: a retrospective observational study

, , , , &
Pages 3178-3182 | Received 28 Jul 2017, Accepted 08 Aug 2017, Published online: 24 Aug 2017
 

Abstract

Introduction: To improve the choice of vacuum-assisted delivery (VAD) system, we compared the outcomes of Kiwi handheld system and Mityvac M-style conventional system (both use disposable plastic cups).

Materials and methods: Retrospective observational study with data collection from electronic medical records. The study was conducted at a tertiary medical center, with approximately 7000 deliveries annually. Categorical and continuous variables were analyzed using chi-square test and t-test, respectively. p value < .05 was considered significant. The main outcomes assessed were the overall failure rate of each system, failure rates for occipito-anterior (OA) versus occipito-transverse/occipito-anterior (OT/OP) positions, +1 versus +2 fetal stations, and early maternal/neonatal outcomes.

Results: During a 10-month period, there were 507 (8.4%) attempted VADs, 36 failed (7.1%), and eight (1.5%) converted to cesarean section. Of these, 364 were Kiwi-assisted and 143 Mityvac-assisted. Background characteristics were similar. The handheld system had more failures overall (9.6 versus 0.7%), at OA (7.6 versus 0.9%), and non-OA positions (17.3% versus none), at +1 (13.25 versus 0.96%) and at +2/3 stations (6.1% versus none), than the conventional system did, respectively. There was a higher rate of early post-partum hemorrhage (15.3 versus 7.4%) in the conventional group. Both systems had similar rates of third/fourth degree perineal tears, shoulder dystocia and adverse neonatal outcomes.

Conclusions: Our results suggest more failures with Kiwi compared to Mityvac, overall and at any fetal position/station, without a significant difference in adverse outcome profile.

View correction statement:
Notice of duplicate publication: ‘Kiwi OmniCup Handheld vs. Mityvac M-Style Conventional Vacuum System: a Retrospective Observational Study’

Acknowledgements

We would like to thank Ms. Nava Jelin for data analysis and Ms. Faye Schreiber for editing. They are employees of Meir Medical Center. The study was approved by the Institutional Ethics Committee in July 2016, registration number 0116-16-MMC.

Disclosure statement

The authors report no conflicts of interest.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access
  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart
* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.