Remote monitoring of cardiovascular implanted electronic devices: a paradigm shift for the 21st century

&
Pages 367-376 | Published online: 09 Jan 2014

Abstract

Traditional follow-up of cardiac implantable electronic devices involves the intermittent download of largely nonactionable data. Remote monitoring represents a paradigm shift from episodic office-based follow-up to continuous monitoring of device performance and patient and disease state. This lessens device clinical burden and may also lead to cost savings, although data on economic impact are only beginning to emerge. Remote monitoring technology has the potential to improve the outcomes through earlier detection of arrhythmias and compromised device integrity, and possibly predict heart failure hospitalizations through integration of heart failure diagnostics and hemodynamic monitors. Remote monitoring platforms are also huge databases of patients and devices, offering unprecedented opportunities to investigate real-world outcomes. Here, the current status of the field is described and future directions are predicted.

When Michel Mirowski developed the first implantable cardioverter defibrillator (ICD) in 1980 Citation[1], he would scarcely have imagined that 30 years later, millions of such devices implanted in patients all over the world would be transmitting data on device and patient status to physicians’ offices via global networks. From technologically primitive transtelephonic monitoring (TTM) of pacemakers from the 1970s onwards, rapid evolution to automatic continuous remote monitoring sets the future standard of care Citation[2].

Definition

Remote monitoring of cardiovascular implantable electronic devices (CIED) involves the transmission of data regarding the status of the device, patient variables gathered by the device and sometimes additional disease-related data collected by the patient or caregiver (blood pressure and weight), over a network from the patient’s location via a central database to a hospital or physician’s office. A distinction should be made between true remote monitoring involving the automated transmission of data at regular intervals, and remote interrogation, in which scheduled device interrogations are carried out by the patient in their home, and the data are transmitted. Remote programming of devices, while technically feasible with existing systems, has not yet been implemented.

A paradigm shift

Traditional follow-up of recipients of CIEDs involves an office visit during which the device is interrogated and reprogrammed if necessary. The frequency at which patients are followed-up is arbitrary, requiring adjustment according to individual need, but is typically every 6–12 months for pacemakers, and every 3–6 months for ICDs and cardiac resynchronization therapy (CRT) devices Citation[3]. There are two problems with this episodic approach. First, with expanding indications for ICD and CRT therapy, this leads to an unsustainable burden on device clinics. Second, the majority of visits involve data collection without reprogramming or change in patient medications or investigations (‘nonactionable’). More significantly, patient or device events (lead malfunction or tachyarrhythmia therapy) may remain undetected for a prolonged period between device interrogations.

The safety and efficacy of remote monitoring to replace most routine office visits was demonstrated by the the TRUST trial Citation[4]. In this multicenter study of the BIOTRONIK Home Monitoring® system (Biotronik AG, Berlin, Germany), 1339 recipients of single- or dual-chamber ICDs were randomized to Home Monitoring, with office visits at 3 and 15 months, or conventional care with office visits only. Overall, 85.8% of the 6-, 9- and 12-month follow-ups were performed remotely only. Given that following physicians were permitted to follow-Home Monitoring with in-person visits if desired, this may represent a minimum estimate of the potential for remote monitoring to reduce clinic burden. Reinforcing the concept that the vast majority of device follow-ups involve only data download without programming changes or other alterations to the patient’s therapy, only 6.6% of the 3-monthly scheduled checks in TRUST were actionable Citation[4]. These data from US practice were supported by a recent European study indicating that 78% of scheduled ICD follow-up clinic visits did not involve reprogramming, medication change or any other intervention – in other words, they purely involved purely data download from the ICD, and those data were non-actionable Citation[5].

The potential for remote monitoring to allow near-real-time diagnosis and treatment of disease or device malfunction was also strikingly illustrated in TRUST. The time from event onset to physician evaluation for combined atrial fibrillation, ventricular tachycardia and ventricular fibrillation was 1 day in the Home Monitoring group versus 35.5 days in the conventional group Citation[4]. The early detection of silent arrhythmic or device events can facilitate action and has the potential to positively affect patient outcomes. This early notification function of remote monitoring depends on reliable transmission pathways. Current systems use different proprietary technologies. In an early study (the PREFER trial), Citation[6], remote follow-up using the wanded CareLink® platform (Medtronic Inc., Minneapolis, MN, USA) was compared with transtelephonic monitoring (TTM) of pacemakers. The median time to diagnosis of a clinically actionable arrhythmic event was 5.7 months in the remote arm versus 7.7 in the conventional (TTM with in-office follow-up) arm. A difference of this magnitude is not very useful for clinical purposes, and the comparator, TTM, is primarily useful for determining battery status. In a later trial (CONNECT) Citation[7], utilizing a wireless CareLink system with automatic clinician alerts incorporated into ICDs, the time from event onset to a clinical decision being made improved substantially, but the median 4.6 days from event to decision in the remote monitoring group indicates residual inefficiencies in workflow, or perhaps reluctance on the physician’s part to take action based on device-detected events. Of more concern, only 31% of alert messages were successfully triggered and transmitted Citation[7]. This modest ability questions whether this system is sufficiently robust for follow-up, since the lack of face–face encounters mandates technology that can be relied on to provide reliable alert transmission when problems arise. This role may be filled by systems that are used in TRUST: Home Monitoring was associated with a transmission success rate of (i.e., the proportion of daily device transmission reaching the clinic) 91% Citation[8].

Available platforms

Remote monitoring is currently available on five different platforms in Europe and four in the USA . These vary in their technical specifications and other details, but share a common basic design. This involves transfer of data from the CIED to a transmitter unit, which then sends data over either a landline or global system for mobile communications (GSM) network to a central server. There, the data are formatted for display on a secure internet website, which can be accessed by the following physician or device specialist. The data available are comparable to those available in the office, and include comprehensive diagnostics, device performance and intracardiac electrograms. Clinicians can also be alerted via other means, such as e-mail, short message service or fax, for critical events.

Home Monitoring

BIOTRONIK pioneered automatic remote monitoring in the late 1990s and early 2000s with its Home Monitoring system. This is available in all of its current pacemakers, ICDs and CRT devices. Wireless transmission from a miniature antenna in the device occurs daily to a receiver apparatus (CardioMessenger®), which is usually placed on the patient’s bedside table, to enable transmission at night. The CardioMessenger can use either a landline or the GSM network to transmit these data to the central server, where they are processed and made available to the following physician or device specialist through a password-protected website. To date, Home Monitoring is the only system that allows daily device transmission independent of any patient or physician interaction with the system.

LATITUDE®

Boston Scientific’s LATITUDE system incorporates wireless radiofrequency (RF) telemetry and is available on Boston Scientific’s ICDs and cardiac resynchronization therapy-defibrillation (CRT-D) (but not cardiac resynchronization therapy-pacing [CRT-P]) devices, with pacemaker support planned in the future. The LATITUDE Communicator transmits via a landline only. Transmissions can be either automatic or manual (patient initiated). Additional blood pressure and weight scales can also transmit data to the Communicator, and hence to the physician’s office, via this platform. The potential for early intervention in heart failure patients using this system is under investigation. The extensive data available from the LATITUDE platform are analyzed by a physician panel and have provided important insights into several clinical questions Citation[9–11].

Merlin.net™

St. Jude Medical’s platform also uses RF telemetry to communicate with the Merlin@home monitor, which transmits data over either a landline or GSM network. Automatic or manual transmissions are possible from current ICDs, CRT devices and newer pacemakers, and data from office visits can also be integrated into the system. The Merlin.net system includes a programmable feature that can call patients with prerecorded messages when transmissions have not been made, to encourage compliance and reduce the time spent by the device clinic following up missed transmissions. Alerts, including electrograms, are also available to read directly on several mobile devices.

CareLink®

Medtronic’s CareLink system is the only platform to still utilize an inductive system where it is necessary for the patient at home to place a wand over the device, similar to a programmer in a face-to-face clinic visit, as well as wireless transmission for newer device series. Data are then transmitted from the CareLink monitor over a landline or GSM network to the central server. Medtronic pacemakers, antitachycardia devices and implantable loop recorders are compatible with this system. The disadvantage of an inductive system is that patient interaction, and therefore compliance, is necessary. It can, however, be used to download data from devices lacking wireless capability. Automatic RF transmissions for programmable alert conditions are available in newer ICDs and CRT-D devices, but were found to have suboptimal transmission success when tested in a clinical trial (CONNECT) Citation[7]. Once triggered, an alert can only be reset during an in-office evaluation. In addition, frequent transmissions impose conspicuous battery drain Citation[12].

Smartview™

Sorin is the latest entrant to the remote monitoring field. Their Smartview platform received the Conformité Europeene mark in May 2011, and is available in Europe but not currently in the USA. It is compatible with current-generation ICDs, and operates on either a landline or GSM network. Data are downloaded from the device automatically at night, and scheduled as well as on-demand interrogations are available. Following physicians can access portable document format reports online.

Emerging uses of remote monitoring: beyond device interrogation

While remote monitoring has the potential to replace much in-office device follow-up, other uses have the potential to offer improved, timely interventions to patients. Successful application to these purposes demand daily automatic transmission as tested in TRUST, rather than scheduled intermittent remote interrogation.

Device integrity

Analysis of the TRUST trial showed that ICD generator and lead malfunction, while rare, are underestimated by conventional in-office follow-up and are detected late. By contrast, these are detected rapidly in devices monitored remotely, enabling rapid clinical intervention Citation[13]. Indicators of a device problem, such as lead impedance out of range, trigger alerts, which are transmitted to the following physician or device specialist, may avoid symptomatic presentation with, for instance, inappropriate therapy Citation[14]. Detection of asymptomatic device failure is especially facilitated by platforms capable of automatic daily transmission.

Existing estimates of device failure rates in the population at large are based on conventional in-office follow-up and symptomatic presentation, and differing definitions of failure further cloud the picture Citation[15]. Remote monitoring provides the opportunity for long-term surveillance of device performance in the population at large, which may lead to earlier identification of increased failure rates and more accurate definition of the magnitude of the problem.

Atrial fibrillation

Atrial fibrillation greatly increases the risk of stroke or systemic thromboembolism, which is common and often asymptomatic in CIED recipients, and can be detected by dual-chamber CIEDs. Early physician notification of this through remote monitoring permits appropriate anticoagulation, cardioversion and other management decisions Citation[16,17]. Such action was taken a median of 148 days before the next scheduled in-office follow-up in a recent Italian study of 166 patients using the Home Monitoring system Citation[17]. Computer modeling suggests that in patients monitored daily, with anticoagulation commenced on identification of atrial fibrillation, a reduction in stroke of 9–18% over 2 years is possible Citation[18]. The ongoing IMPACT trial is a prospective, multicenter trial which will test the hypothesis that initiation and withdrawal of oral anticoagulant therapy guided by remote monitoring of atrial intracardiac electrograms will improve clinical outcomes by reducing the combined rates of stroke, systemic embolism and major bleeding compared with conventional management in patients with a CHADS2 score ≥1 Citation[19].

Heart failure monitoring

Telemonitoring of heart failure patients, using a variety of devices, has recently been reviewed Citation[20,21]. The term ‘telemonitoring’ is preferred when referring to devices that are designed purely to monitor patient status, as opposed to remote monitoring of CIEDs, which while developed primarily to assess device performance, may also incorporate patient status indicators.

Several physiological variables, which are measured easily by CIEDs, have been found to predict impending heart failure exacerbation. These include patient activity, mean heart rate, heart rate variability and arrhythmias Citation[22,23]. Most manufacturers now include a suite of these measurements as part of the device interrogation, either in-office or remotely. With frequent remote monitoring, it should in theory be possible to detect impending decompensation and adjust therapy accordingly to avoid hospitalization. However, any single measurement or combination of measurements has not been specific enough to be useful. Medtronic ICDs and CRT devices measure the intrathoracic impedance, which should decrease if significant pulmonary congestion is present. This is reported as the OptiVol Fluid Index™, and can trigger an audible alert if it exceeds a prespecified threshold. However, intrathoracic impedance can be affected by other conditions, such as pericardial or pleural effusion, pneumonia and pocket hematoma, and it has low sensitivity and low positive-predictive value for predicting worsening heart failure Citation[24]. In fact, in the recent DOT-HF trial, patients who had the audible alert turned on had a higher rate of heart failure hospitalization and office visits Citation[25]. The use of the CareLink network to remotely follow a newer iteration of OptiVol to adjust therapy is being examined in the ongoing Phase IV OptiLink HF study Citation[101]. Boston Scientific’s LATITUDE platform includes optional blood pressure and weight sensors, data from which can be transmitted along with symptom status to the following heart failure specialist. The utility of this data was studied by the as-yet unpublished observational DECODE study Citation[26]. These systems, while useful, have not been shown to impact patient outcomes, and may be more useful to trigger an unscheduled evaluation rather than to remotely guide therapy.

Remote monitoring may be an ideal technology to combine with implantable hemodynamic monitors, several of which are currently in development. The HeartPOD™ device (St. Jude Medical) is implanted through the interatrial septum and directly measures left atrial pressure. It is currently under evaluation in a Phase III trial Citation[102]. It can be implanted as a stand-alone device, or integrated with a CRT or ICD device (Promote® left atrial pressure system). Several stand-alone hemodynamic monitors are also in development, which use various telemonitoring systems to communicate with the patient and physician.

Data fidelity & security

Of the currently available proprietary remote monitoring systems, published data on transmission characteristics are available for the Home Monitoring and CareLink networks Citation[2,7,8]. These show variable rates of successful data transmission, as discussed earlier. A mobile communicator that transmits on the GSM network is a clear advantage when a patient travels. Currently, international GSM access varies greatly by manufacturer and location. Interference from sources of electromagnetic interference, such as mobile phones, has not been reported Citation[27]. With inductive systems, most missing transmissions are as a result of the patient missing a transmission, or of the communicator not being plugged into the phone line or otherwise incorrectly set up Citation[7].

Of more concern is data security. While we are not aware of any reports of actual data loss to theft or hacking, the potential for this exists, despite encrypted transmission and password-protected internet platforms. While one group effected programming changes in a Medtronic Maximo® DR ICD under laboratory conditions, this would be difficult to replicate in the real world without the patient’s knowledge Citation[28].

Workflow

Remote monitoring appears to be a solution to the ever-increasing burden of follow-up encountered by device clinics worldwide. However, the reality may be more complex, and may depend on the system used and the healthcare infrastructure within which it is applied. Even though remote monitoring reduces in-office follow-up, data transmitted do require review by the following physician or device specialist Citation[4–7]. It is intuitive that this is accomplished more rapidly; however, conclusive data on the overall effect on workflow are lacking. With the development of automated pace capture threshold determination by many devices, remote transmissions deliver almost exactly the same data that are seen during an office interrogation. Remote programming is not currently implemented; therefore, direct comparison with office follow-up, which does permit programming changes, is not possible. Certainly, review of a nonactionable remote transmission is faster than a similar interrogation in-office. A European survey found that scheduled visits lasted 27 min Citation[29], compared with a mean of 11.5 min for scheduled or unscheduled remote transmissions at our institution Citation[30]. Our data also suggest that while review of remote monitoring transmissions can be rapidly performed, telephone contact with patients to schedule office follow-up, and especially to troubleshoot and encourage compliance among patients from whom transmissions have not been received, is time-consuming Citation[30]. It is possible that mechanisms for feedback via the communicator to patients to encourage compliance would be useful to ameliorate this limitation.

It is crucial to have a clear and efficient system in position to triage and act upon the large amount of information transmitted from CIEDs. At the Cleveland Clinic (OH, USA), a trained device nurse is rostered to review all four remote monitoring platforms each day for scheduled and unscheduled transmissions. Issues requiring the physician’s attention are directed to the patient’s attending electrophysiologist. This removes any uncertainty regarding whether action has been taken on a particular issue, as the remote monitoring websites can only record whether a transmission has been reviewed, but not by whom or whether the issue has been resolved. Platforms include the option to personalize alerts for each patient, and this should be used to tailor the data presented to the most useful for each individual patient.

An issue that frequently limits the utility of remote monitoring in our experience is patient compliance. For example, Movsowitz et al. cite data from Medtronic showing that out of 265,024 patients implanted with a wireless device, only 66% had been enrolled in the CareLink system, 25% of these had never made a transmission and only 5.2% had completed the scheduled quarterly transmissions Citation[31].

Patient acceptance & satisfaction

One factor that might decrease patient compliance with remote monitoring is satisfaction. Although it has been assumed that patients would appreciate the convenience of remote follow-up, this has been questioned and might not be the case in all cultures, patient populations or at all times post-CIED implant Citation[32]. In a pilot study of remote monitoring using the CareLink system versus standard in-office follow-up in patients implanted for a mean of 1.5 years at enrollment, health-related quality of life and patient satisfaction measures were worse at 6 months but the difference had disappeared by 12 months Citation[33]. It is possible that personal contact is important to satisfaction shortly after CIED implant, and declines thereafter. However, the vast majority of the evidence available to date indicates a high level of patient satisfaction with remote monitoring. Overall, 98% of the patients in the remote monitoring arm in TRUST elected to continue this mode of follow-up after completion of the trial Citation[34]. Other studies reporting patient satisfaction in diverse populations with CareLink Citation[35–38], Home Monitoring Citation[39] and Housecall II (the forerunner of Merlin.net) Citation[40] are universally positive, though each has used different methodologies. Physician satisfaction has also been high, where assessed Citation[35–38].

Economic impact & cost–effectiveness

Cost–effectiveness studies on remote monitoring are few. Unique healthcare system organization, costs, reimbursement rates and populations mean that generalization from one country to another may be inaccurate. Studies from France Citation[41] and Finland Citation[37] estimate savings of US$2149 over a projected 5-year lifetime of an ICD, and €524 over 9 months, respectively. However, a US pilot study using the CareLink system, which randomized 151 patients to remote monitoring or quarterly in-office interrogations, did not show any reduction in unscheduled hospital visits or costs Citation[33]. Indeed, the remote monitoring strategy was slightly more expensive. The much larger CONNECT trial examined healthcare utilization as a secondary end point. While there were no differences in mortality, emergency room visits, hospitalizations or unscheduled clinic visits between the arms, length of stay was shorter in remotely followed patients who were hospitalized during the study. This led to an estimated US$1793 reduction in cost per hospitalization Citation[7]. Remote monitoring involves an initial outlay for the transmitter and installation, and due to the earlier and more frequent detection of device problems and patient deterioration, it has the potential to increase in-person evaluations, medications and interventions in the short term Citation[4,6,7,13,14,16]. While this should logically be offset by improved patient outcomes and reduced costs over the longer-term, this remains to be demonstrated in practice. The economic analyses of the recently completed EVATEL and ECOST trials from France should provide further information on cost–effectiveness Citation[103,104]. Multinational data on healthcare economic aspects of remote monitoring will also be provided by the ongoing EuroEco trial Citation[105]. Ultimately, analyses specific to each health system are needed to accurately gauge the cost–effectiveness of remote monitoring. The complex economic implications of remote monitoring have been reviewed in depth by Burri et al. Citation[42], and the current state of reimbursement across Europe for both in-person and remote follow-up of CIEDs by Boriani et al. Citation[43].

Reimbursement contributes to cost, but can also either drive or inhibit implementation of remote monitoring. While reimbursement is established in the USA, it is available in only a few countries in Europe, such as the UK, Germany and Portugal, where it is similar to that offered for standard follow-up visits. Proof of cost–effectiveness, as discussed above, may be required before reimbursement for remote monitoring is universally implemented.

Research

Remote monitoring platforms can be viewed as vast databases of device characteristics, programming, therapies, and patient data including physiologic and disease state. As such, they are a treasure trove for researchers seeking ‘real-world’ outcome data. In the USA, the ALTITUDE study group coordinates research into relevant clinical questions using the LATITUDE database. Survival is determined from the Social Security Death Index, and device electrograms are adjudicated by an expert panel, with good interobserver agreement Citation[44]. The striking finding of the ALTITUDE survival study was that remote monitoring was associated with a 50% relative reduction in the risk of death (ICD hazard ratio: 0.56; CRT-D hazard ratio: 0.45; p < 0.0001) versus office follow-up only Citation[9]. Whether this is due to improved surveillance of clinical status, improved compliance or subtle differences between remotely and conventionally monitored patients (confounding) remains to be elucidated. It is difficult to imagine a randomized clinical trial being large enough, or with long enough follow-up, to detect findings such as this. The ALTITUDE group has also provided insights into the effect of programming on appropriate and inappropriate shocks, the influence of percentage biventricular pacing on outcome and several other issues Citation[10,11].

Remote monitoring also provides a more convenient and efficient method of following patients enrolled in clinical trials of CIEDs, as seen in the DAVID II trial Citation[45]. Patients may be more likely to enroll in studies if some follow-up can be performed remotely, lessening the burden of office follow-up visits.

Guidelines

While specific professional society guidelines do not exist for remote monitoring, the topic is covered by the 2008 Heart Rhythm Society/European Heart Rhythm Association expert consensus on the monitoring of CIEDs Citation[3]. These are necessarily consensus-rather than evidence-driven, and were written before the publication of TRUST or CONNECT. Remote interrogation rather than remote monitoring was the objective of discussion. These guidelines recommend that remote monitoring can replace office follow-up in patients whose condition is stable and device reprogramming is not required. In-office visits are still recommended for the immediate post-implant assessment and early surveillance period (4–12 weeks post-implant), and once yearly thereafter. As remote monitoring does not provide an opportunity to take a cardiovascular history or to examine the patient or the wound, these in-office visits are essential to monitor the patient, educate the patient about their device, and assess wound healing.

Expert commentary

Remote monitoring represents a paradigm shift in the follow-up of CIEDs. Intermittent data download, which may miss significant events and uncover others several months after they have occurred, can now be replaced with near-real-time monitoring of device function and patient status. This has significant potential for monitoring both atrial and ventricular arrhythmia occurrence, surveillance for generator and lead integrity, and heart failure status. Although remote monitoring has the potential to improve the device clinic workflow and efficiency, and is probably cost saving, these variables are healthcare system specific and it is difficult to extrapolate from an experience in one area to all others. Lack of reimbursement for remote device follow-up is currently inhibiting the expansion of this technology in several areas of the world. Further economic analyses will be necessary to clarify the economic impact of remote follow-up of CIEDs and provide the stimulus for reimbursement and widespread, worldwide adoption. Remote monitoring networks provide unparalleled opportunities for ‘real-world’ outcomes research regarding patients with CIEDs. Perhaps most intriguingly, remote monitoring of patients with ICDs and CRT-D devices was associated with a 50% reduction in mortality in a large unselected population Citation[9].

Five-year view

Remote monitoring will become the usual method of follow-up for patients with CIEDs in most health systems. Manufacturers will move towards automatic transmission without the need for patient interaction with the system to simplify the interrogation, and improve information flow and compliance. Cellular transmission will also be favored over landline-based communicators given the declining use of landlines and to allow portability. Remote programming, probably in a simple form, will become available for clinical use. However, complex programming changes will still require in-person consultation, as will early post-implant and annual visits. Analysis of remote monitoring databases will continue to provide novel insights into outcomes of patients with CIEDs that would not be possible with conventional research methodology.

Table 1. Technical aspects of currently available remote monitoring platforms.

Table 2. Ongoing studies of remote monitoring registered on ClinicalTrials.gov as of January 2012.

Key issues

  • • Remote monitoring is the transmission of cardiac implantable electronic device performance and patient data over a network from the patient’s home to their physician’s office. Data transfer can be automated or manual.

  • • This represents a paradigm shift from episodic in-office device interrogation, which largely consists of analysis of nonactionable data and may miss important clinical or device issues.

  • • Advantages include earlier detection of patient events such as arrhythmia onset and device therapy, programming issues such as therapies programmed off, and device integrity problems such as lead fracture.

  • • Remote monitoring platforms are marketed by all the major cardiac electronic implantable device manufacturers, including Medtronic (CareLink®), St. Jude (Merlin.net™), Boston Scientific (LATITUDE®), BIOTRONIK (Home Monitoring®) and Sorin (SMARTVIEW™).

  • • The effect on workflow is complex but probably results in significant time and cost savings, depending on the healthcare system and platforms available.

  • • Remote monitoring is associated with patient and physician satisfaction. Data security appears to be robust. However, technical performance is not uniform across different proprietary technologies.

  • • Although limited available data suggest cost savings, large-scale economic analyses are needed to confirm this before widespread acceptance and reimbursement is implemented.

  • • Remote monitoring platforms provide an unparalleled dataset of real-world patients, which can be utilized to assess outcomes and improve practice.

  • • In a large unselected population with implantable cardioverter defibrillators and cardiac resynchronization therapy defibrillation devices, remote monitoring was associated with significantly improved survival.

Financial & competing interests disclosure

EM Cronin has received minor educational support from Boston Scientific. N Varma has received minor consulting fees from Boston Scientific, BIOTRONIK and St. Jude Medical. The authors have no other relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript apart from those disclosed.

No writing assistance was utilized in the production of this manuscript.

References

  • Mirowski M, Reid PR, Mower MM et al. Termination of malignant ventricular arrhythmias with an implanted automatic defibrillator in human beings. N. Engl. J. Med. 303(6), 322–324 (1980).
  • Wallbrück K, Stellbrink C, Santini M, Gill J, Hartmann A, Wunderlich E. The value of permanent follow-up of implantable pacemakers – first results of an European trial. Biomed. Tech. 47(Suppl. 1 Pt 2), 950–953 (2002).
  • Wilkoff BL, Auricchio A, Brugada J et al.; Heart Rhythm Society; European Heart Rhythm Association; American College of Cardiology; American Heart Association; European Society of Cardiology; Heart Failure Association of ESC; Heart Failure Society of America. HRS/EHRA expert consensus on the monitoring of cardiovascular implantable electronic devices (CIEDs): description of techniques, indications, personnel, frequency and ethical considerations. Heart Rhythm 5(6), 907–925 (2008).
  • Varma N, Epstein AE, Irimpen A, Schweikert R, Love C; TRUST Investigators. Efficacy and safety of automatic remote monitoring for implantable cardioverter-defibrillator follow-up: the Lumos-T Safely Reduces Routine Office Device Follow-up (TRUST) trial. Circulation 122(4), 325–332 (2010).
  • Heidbüchel H, Lioen P, Foulon S et al. Potential role of remote monitoring for scheduled and unscheduled evaluations of patients with an implantable defibrillator. Europace 10(3), 351–357 (2008).
  • Crossley GH, Chen J, Choucair W et al.; PREFER Study Investigators. Clinical benefits of remote versus transtelephonic monitoring of implanted pacemakers. J. Am. Coll. Cardiol. 54(22), 2012–2019 (2009).
  • Crossley GH, Boyle A, Vitense H, Chang Y, Mead RH; CONNECT Investigators. The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) trial: the value of wireless remote monitoring with automatic clinician alerts. J. Am. Coll. Cardiol. 57(10), 1181–1189 (2011).
  • Varma N, Pavri B, Stambler B, Michalski J, TRUST investigators. Effect of transmission reliability on remote follow up in ICD patients: automatic home monitoring in the TRUST trial. Europace 13(3), abstract P1026 (2011).
  • Saxon LA, Hayes DL, Gilliam FR et al. Long-term outcome after ICD and CRT implantation and influence of remote device follow-up: the ALTITUDE survival study. Circulation 122(23), 2359–2367 (2010).
  • Hayes DL, Boehmer JP, Day JD et al. Cardiac resynchronization therapy and the relationship of percent biventricular pacing to symptoms and survival. Heart Rhythm 8(9), 1469–1475 (2011).
  • Gilliam FR, Hayes DL, Boehmer JP et al. Real world evaluation of dual-zone ICD and CRT-D programming compared to single-zone programming: the ALTITUDE REDUCES study. J. Cardiovasc. Electrophysiol. 22(9), 1023–1029 (2011).
  • Medtronic Inc. Clinician Manual Supplement Protecta™ XT / Protecta™. Projected service life information related to remote monitoring. Manual no.: M945739A001A, Medtronic Inc., MN, USA (2010).
  • Varma N, Michalski J, Epstein AE, Schweikert R. Automatic remote monitoring of implantable cardioverter-defibrillator lead and generator performance: the Lumos-T Safely RedUceS RouTine Office Device Follow-Up (TRUST) trial. Circ. Arrhythm. Electrophysiol. 3(5), 428–436 (2010).
  • Varma N. Remote monitoring for advisories: automatic early detection of silent lead failure. Pacing Clin. Electrophysiol. 32(4), 525–527 (2009).
  • Hauser RG, Hayes DL. Increasing hazard of Sprint Fidelis implantable cardioverter-defibrillator lead failure. Heart Rhythm 6(5), 605–610 (2009).
  • Varma N, Stambler B, Chun S. Detection of atrial fibrillation by implanted devices with wireless data transmission capability. Pacing Clin. Electrophysiol. 28(Suppl. 1), S133–S136 (2005).
  • Ricci RP, Morichelli L, Santini M. Remote control of implanted devices through Home Monitoring technology improves detection and clinical management of atrial fibrillation. Europace 11(1), 54–61 (2009).
  • Ricci RP, Morichelli L, Gargaro A, Laudadio MT, Santini M. Home monitoring in patients with implantable cardiac devices: is there a potential reduction of stroke risk? Results from a computer model tested through monte carlo simulations. J. Cardiovasc. Electrophysiol. 20(11), 1244–1251 (2009).
  • Ip J, Waldo AL, Lip GY et al.; IMPACT Investigators. Multicenter randomized study of anticoagulation guided by remote rhythm monitoring in patients with implantable cardioverter-defibrillator and CRT-D devices: rationale, design, and clinical characteristics of the initially enrolled cohort The IMPACT study. Am. Heart J. 158(3), 364–370.e1 (2009).
  • Hasan A, Paul V. Telemonitoring in chronic heart failure. Eur. Heart J. 32(12), 1457–1464 (2011).
  • Anker SD, Koehler F, Abraham WT. Telemedicine and remote management of patients with heart failure. Lancet 378(9792), 731–739 (2011).
  • Ellery S, Pakrashi T, Paul V, Sack S. Predicting mortality and rehospitalization in heart failure patients with home monitoring – the Home CARE pilot study. Clin. Res. Cardiol. 95(Suppl. 3), III29–III35 (2006).
  • Adamson PB, Smith AL, Abraham WT et al.; InSync III Model 8042 and Attain OTW Lead Model 4193 Clinical Trial Investigators. Continuous autonomic assessment in patients with symptomatic heart failure: prognostic value of heart rate variability measured by an implanted cardiac resynchronization device. Circulation 110(16), 2389–2394 (2004).
  • Conraads VM, Tavazzi L, Santini M et al. Sensitivity and positive predictive value of implantable intrathoracic impedance monitoring as a predictor of heart failure hospitalizations: the SENSE-HF trial. Eur. Heart J. 32(18), 2266–2273 (2011).
  • van Veldhuisen DJ, Braunschweig F, Conraads V et al.; DOT-HF Investigators. Intrathoracic impedance monitoring, audible patient alerts, and outcome in patients with heart failure. Circulation 124(16), 1719–1726 (2011).
  • Ewald GA, Gilliam FR, Sweeney RJ. Automated HF Decompensation Detection: results from the DECOmpensation DEtection Study (DECODE) [abstract]. J. Cardiac Fail. 15, S122 (2009).
  • Calcagnini G, Censi F, Floris M et al. Evaluation of electromagnetic interference of GSM mobile phones with pacemakers featuring remote monitoring functions. Pacing Clin. Electrophysiol. 29(4), 380–385 (2006).
  • Halperin D, Heydt-Benjamin TS, Ransford B et al. Pacemakers and implantable cardiac defibrillators: software radio attacks and zero-power defenses. Presented at: 2008 IEEE Symposium on Security and Privacy. Oakland, CA, USA, 18–21 May 2008.
  • Boriani G, Auricchio A, Klersy C et al.; European Heart Rhythm Association; Eucomed. Healthcare personnel resource burden related to in-clinic follow-up of cardiovascular implantable electronic devices: a European Heart Rhythm Association and Eucomed joint survey. Europace 13(8), 1166–1173 (2011).
  • Cronin EM, Ching EA, Varma N, Martin DO, Wilkoff BL, Lindsay BD. Remote monitoring of cardiovascular devices – a time and activity analysis. Heart Rhythm (2012) (In Press).
  • Movsowitz C, Mittal S. Remote patient management using implantable devices. J. Interv. Card. Electrophysiol. 31(1), 81–90 (2011).
  • Matlock DD. Big Brother is watching you: what do patients think about ICD home monitoring? Circulation 122(4), 319–321 (2010).
  • Al-Khatib SM, Piccini JP, Knight D, Stewart M, Clapp-Channing N, Sanders GD. Remote monitoring of implantable cardioverter defibrillators versus quarterly device interrogations in clinic: results from a randomized pilot clinical trial. J. Cardiovasc. Electrophysiol. 21(5), 545–550 (2010).
  • Varma N, Stambler BS. Letter by Varma and Stambler regarding article, Big Brother is watching you: what do patients think about ICD home monitoring? Circulation 123(7), e247 (2011).
  • Schoenfeld MH, Compton SJ, Mead RH et al. Remote monitoring of implantable cardioverter defibrillators: a prospective analysis. Pacing Clin. Electrophysiol. 27(6 Pt 1), 757–763 (2004).
  • Marzegalli M, Lunati M, Landolina M et al. Remote monitoring of CRT-ICD: the multicenter Italian CareLink evaluation – ease of use, acceptance, and organizational implications. Pacing Clin. Electrophysiol. 31(10), 1259–1264 (2008).
  • Raatikainen MJ, Uusimaa P, van Ginneken MM, Janssen JP, Linnaluoto M. Remote monitoring of implantable cardioverter defibrillator patients: a safe, time-saving, and cost-effective means for follow-up. Europace 10(10), 1145–1151 (2008).
  • Ando K, Koyama J, Abe Y et al. Feasibility evaluation of a remote monitoring system for implantable cardiac devices in Japan. Int. Heart J. 52(1), 39–43 (2011).
  • Ricci RP, Morichelli L, Quarta L et al. Long-term patient acceptance of and satisfaction with implanted device remote monitoring. Europace 12(5), 674–679 (2010).
  • Joseph GK, Wilkoff BL, Dresing T, Burkhardt J, Khaykin Y. Remote interrogation and monitoring of implantable cardioverter defibrillators. J. Interv. Card. Electrophysiol. 11(2), 161–166 (2004).
  • Fauchier L, Sadoul N, Kouakam C et al. Potential cost savings by telemedicine-assisted long-term care of implantable cardioverter defibrillator recipients. Pacing Clin. Electrophysiol. 28(Suppl. 1), S255–S259 (2005).
  • Burri H, Heidbüchel H, Jung W, Brugada P. Remote monitoring: a cost or an investment? Europace 13(Suppl. 2), ii44–ii48 (2011).
  • Boriani G, Burri H, Mantovani LG et al. Device therapy and hospital reimbursement practices across European countries: a heterogeneous scenario. Europace 13(Suppl. 2), ii59–ii65 (2011).
  • Powell BD, Cha YM, Asirvatham SJ et al. Implantable cardioverter defibrillator electrogram adjudication for device registries: methodology and observations from ALTITUDE. Pacing Clin. Electrophysiol. 34(8), 1003–1012 (2011).
  • Wilkoff BL, Kudenchuk PJ, Buxton AE et al.; DAVID II Investigators. The DAVID (Dual Chamber and VVI Implantable Defibrillator) II trial. J. Am. Coll. Cardiol. 53(10), 872–880 (2009).

Websites

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.