Innovative treatments for adults with obstructive sleep apnea

, , , , , , & show all
Pages 137-147 | Published online: 18 Nov 2014

Abstract

Obstructive sleep apnea (OSA) affects one in five adult males and is associated with significant comorbidity, cognitive impairment, excessive daytime sleepiness, and reduced quality of life. For over 25 years, the primary treatment has been continuous positive airway pressure, which introduces a column of air that serves as a pneumatic splint for the upper airway, preventing the airway collapse that is the physiologic definition of this syndrome. However, issues with patient tolerance and unacceptable levels of treatment adherence motivated the exploration of other potential treatments. With greater understanding of the physiologic mechanisms associated with OSA, novel interventions have emerged in the last 5 years. The purpose of this article is to describe new treatments for OSA and associated complex sleep apnea. New approaches to complex sleep apnea have included adaptive servoventilation. There is increased literature on the contribution of behavioral interventions to improve adherence with continuous positive airway pressure that have proven quite effective. New non-surgical treatments include oral pressure devices, improved mandibular advancement devices, nasal expiratory positive airway pressure, and newer approaches to positional therapy. Recent innovations in surgical interventions have included laser-assisted uvulopalatoplasty, radiofrequency ablation, palatal implants, and electrical stimulation of the upper airway muscles. No drugs have been approved to treat OSA, but potential drug therapies have centered on increasing ventilatory drive, altering the arousal threshold, modifying loop gain (a dimensionless value quantifying the stability of the ventilatory control system), or preventing airway collapse by affecting the surface tension. An emerging approach is the application of cannabinoids to increase upper airway tone.

Introduction

Obstructive sleep apnea (OSA) is a common problem. The second most diagnosed respiratory condition after asthma, 9%–28% of women and 24%–26% of males in the US population have apneic events at a treatable level, making this syndrome a serious public health issue.Citation1 Untreated OSA has a profound effect on health and behavior, resulting in excessive daytime sleepiness, impaired cognition and mentation, decreased daytime function, and alterations in mood.Citation1 OSA contributes to increased heart disease, stroke, hypertension, diabetes, and obesity, with a high rate of comorbidity. The economic burden of untreated OSA is enormous, having a cost similar in magnitude to diabetes, ie, 132 billion dollars in 2002, with an estimated $15.9 billion for the 800,000 crashes attributed to OSA.Citation2,Citation3 From the perspective of the health care system, for each quality-adjusted life year, $3,354 and $11.1 billion in collision costs would be saved with successful continuous positive airway pressure (CPAP) treatment, the primary intervention for OSA for over 25 years.Citation3 CPAP introduces a column of air that serves as a pneumatic splint for the upper airway, preventing the airway collapse that is the physiologic definition of this syndrome. Despite demonstrated efficacy and potential cost savings, the overall effectiveness of CPAP is significantly limited by poor adherence.Citation4 This limitation has prompted the development of new approaches to produce more tolerable designs for CPAP treatment, approaches to enhance adherence, and innovations in other interventions to treat OSA. With greater understanding of the physiologic mechanisms associated with OSA, novel interventions have emerged in the last 5 years. The purpose of this article is to describe new treatments for OSA and associated complex sleep apnea.

Advances in positive airway pressure treatment

Complex sleep apnea refers to a pattern of sleep-disordered breathing that includes both obstructive apneas/hypopneas and central apneas. Specifically, this occurs in patients without daytime hypercapnia. Central apneas are noted during non-rapid eye movement sleep and may “emerge” during CPAP titration. It is likely that these individuals have a high loop gain (a dimensionless value quantifying the stability of the ventilatory control system) in addition to a collapsible airway. Loop gain is measured by repeatedly dropping CPAP pressure during sleep to induce prolonged hypopneas, and then returning CPAP pressure to normal. The return of normal pressure causes an overshoot of ventilation, called the “response”. Loop gain is calculated by dividing “the response” by “the stimulus” (the difference between ventilation during the hypopnea and zero ventilation). A high loop gain approaching a value of 1 reflects relative instability in the ventilatory system and predisposes to recurrent apneas. Adaptive servoventilators are spontaneous-timed bilevel positive airway pressure devices that estimate the patient’s minute ventilation and tidal volume and/or flow, and respond to it by delivering variable pressure support. Theoretically, this minimizes the hyperpneic phase (which precipitates central apneas). These devices provide better control of complex sleep apnea than other CPAP devices, particularly in patients with heart failure.Citation6Citation8

The S9™ (ResMed Ltd, Bella Vista, Australia) is a novel CPAP device that includes a humidification system with a heated tube, reduced noise levels, and an adaptive positive airway pressure algorithm. Use of the S9 device resulted in longer average CPAP use by approximately 30 minutes after 28 days in 50 patients with established CPAP use.Citation9 A new CPAP modality, SensAwake® (Fisher and Paykel Healthcare Ltd, Auckland, New Zealand), was developed to monitor breathing patterns and reduce pressure with awakenings. When tested in a group of 42 CPAP-naïve patients with OSA, the SensAwake modality had results similar to those of traditional CPAP.Citation10

Interventions to promote CPAP adherence

Adherence is the major obstacle to effective CPAP treatment. A Cochrane Collaboration report reviewing 30 studies of educational, supportive, and behavioral interventions concluded that all three types of interventions had a positive impact on CPAP usage in 2,047 adults with OSA.Citation11 There was moderate-quality evidence to support a modest increase in adherence to CPAP using short-term educational interventions and low-quality evidence to support a large increase in CPAP adherence using behavioral interventions such as cognitive behavioral therapy and motivational interviewing. Recently, Bartlett et al conducted a randomized controlled study comparing a group cognitive behavioral therapy intervention with a social interaction control on CPAP adherence.Citation12 They found no difference between groups in CPAP adherence or self-efficacy using CPAP at 1 or 6 months. In another randomized controlled study, Deng et al compared a stage-matched intervention with usual care in newly diagnosed OSA patients.Citation13 Participants in the intervention group received care based on their stage of behavior change. Interventions ranged from education about OSA (stage 1) to setting and attaining goals for increased CPAP use at night (stage 4). They found significant differences between groups with regard to CPAP adherence at 1 and 3 months, and in the percentage of subjects using CPAP for more than 4 hours per night.Citation13 Sawyer et al tested a risk screening questionnaire, ie, the Index for Nonadherence to PAP, for CPAP nonadherence in a longitudinal study that included 97 subjects with newly diagnosed OSA. The final Index for Nonadherence to PAP model included 19 items with questions regarding health literacy, self-efficacy, body mass index (BMI), marriage status, presenting symptoms, and gender. The questionnaire successfully identified CPAP nonadherers and may provide a tool for early intervention development to decrease or prevent nonadherence.Citation14 Although there is growing evidence to support educational and behavioral interventions to enhance adherence with CPAP, further studies with stronger designs and larger samples are needed.

Nasal expiratory positive airway pressure

During the five distinct phases of the respiratory cycle, the final expiratory phase has the greatest risk of airway occlusion. At the end of expiration, there is neither positive pressure nor inspiratory phasic activation of the upper airway dilator muscles.Citation15 Moreover, there is significant narrowing of the upper airway during expiration preceding apnea,Citation16 and the lowest cross-sectional of the upper airway occurs during expiration.Citation17Citation19 The nasal expiratory positive airway pressure (nEPAP) device allows for low inspiratory resistance while increasing expiratory resistance to prevent upper airway collapse (see ).Citation18,Citation19 It comprises a single-use device containing a mechanical valve that can create high expiratory resistance, with very low inspiratory resistance applied to each nostril with an adhesive to provide a seal.Citation20,Citation21 There have been six clinical trials using nEPAP.Citation18,Citation20Citation24 In all studies, there was a significant decrease in apnea-hypopnea index (AHI). Four of the nEPAP studies were uncontrolled, proof-of-concept studies,Citation18,Citation20,Citation23,Citation24 one was a prospective, sham-controlled, randomized, double-blind, multicenter clinical trial,Citation21 and another was an extension of the latter clinical trial to observe long-term treatment durability and safety.Citation22

Figure 1 Photograph of the expiratory positive airway pressure device in place.

Notes: Reprinted from Sleep Med, 12(2), Walsh JK, Griffin KS, Forst EH, et al. A convenient expiratory positive airway pressure nasal device for the treatment of sleep apnea in patients non-adherent with continuous positive airway pressure, 147–152, Copyright © 2011, with permission from Elsevier.Citation24
Figure 1 Photograph of the expiratory positive airway pressure device in place.

In the proof-of-concept studies, two used nEPAP devices of varying expiratory resistances to increase tolerability.Citation20,Citation24 In these proof-of-concept studies, there was a 31%–49% decrease in AHI.Citation18,Citation20,Citation23,Citation24 Two studies showed no difference in sleep architecture,Citation18,Citation20 and one study showed significant changes in sleep architecture.Citation24 Two studies quantified and showed significant improvements in daytime sleepiness or sleep quality,Citation20,Citation24 and three studies quantified and showed significant improvement in oxygen saturation.Citation8,Citation20,Citation24 The results of the proof-of-concept trials are encouraging. However, as with CPAP,Citation25 tolerability and adherence to nEPAP was an issue. Although self-reported use was as high as 94%,Citation20,Citation24 15% of participants in one study failed to take the device home.Citation20 In another study, there was a dropout rate of 22% due to lack of tolerability of nEPAP, and many adverse events like difficulty breathing and sleeping, dry mouth, and nasal discomfort were reported.Citation24

A prospective, sham-controlled, double-blind, multicenter clinical trial randomized newly diagnosed or previously diagnosed but untreated OSA patients to either nEPAP or a sham group.Citation21 The sham treatment device was similar to the nEPAP device, with the only difference being that the sham device contained valves that did not provide expiratory resistance. The participants wore the nEPAP/sham devices for 3 months with two nonconsecutive night sleep studies a week into the study and two nonconsecutive night sleep studies at the end of the 3 month study period. The two nonconsecutive night sleep studies were done to evaluate on-device and off-device polysomnographs (randomly assigned). Significant median decreases in AHI after 1 week and 3 months of nEPAP were 53% and 43%, respectively, when compared with sham. There were also significant decreases in oxygen desaturation index, Epworth Sleepiness Scale, and arousal index, and significant changes in sleep architecture between the nEPAP and sham groups. Adherence was high in both the nEPAP and sham groups; however, there was significantly less adherence in the nEPAP group than in the sham group (88.2% versus 92.3%). Adverse events, like difficulty exhaling, nasal discomfort, dry mouth, headache, and insomnia, were not significantly different between the two groups; however, more participants dropped out of the nEPAP group due to adverse events. A follow-up study evaluated the long-term durability of treatment response and safety of nEPAP.Citation22 The positive results obtained from the original 3-month studyCitation21 were replicated after 12 months of wearing nEPAP.Citation22 As in the 3-month study, adverse events were still reported.Citation21,Citation22

Although AHI is an important metric to quantify improvements in treatment response, another outcome that is important to the user as well as his or her sleeping partner is that of snoring. In three studies, the percentage of total sleep time spent snoring decreased significantly by 57.8%–74.4% with use of nEPAP, providing a quieter environment for the sleeping partner.Citation18,Citation20,Citation22

The significant improvements in AHI, oxygen saturation, and snoring with nEPAP are very encouraging, and the improvement in snoring suggests that this treatment may be suitable for those with mild OSA. Moreover, given the challenges of traveling with a CPAP device, nEPAP may be an alternative for patients who travel frequently. As with CPAP, adherence and tolerability are an issue. However, considering the ease of application, compact design, low cost, and availability of these devices (ie, accessible without a prescription), nEPAP may become a second-line therapy to CPAP.

Oral pressure therapy

Oral pressure therapy, in which a vacuum is applied to the oropharynx to pull the soft palate forward and stabilize the tongue, has recently been studied in adult patients with OSA.Citation26 The system, such as the Winx device (see ), includes a mouthpiece in a number of sizes, a vacuum pump, a console, and tubing. While breathing normally through the nose, continuous negative pressure in the oral cavity moves the soft palate forward into contact with the tongue to reduce obstruction. The negative pressure is isolated from the nasopharyngeal airway by the natural seal that occurs between the soft palate and tongue. In a multicenter, prospective, randomized, crossover trial, Colrain et al evaluated the safety and efficacy of an oral pressure therapy system in subjects with mild to severe OSA.Citation27 The therapy was well tolerated and used on average for 6 hours at night. Oral pressure therapy was associated with clinically significant improvements in AHI, sleep quality and continuity, and daytime sleepiness.Citation27 Although there is preliminary evidence to support oral pressure therapy as a therapy for OSA, additional studies with strong designs and larger samples are needed.

Figure 2 (A) Winx device, an oral pressure device that uses gentle suction to anteriorly and superiorly displace the tongue and the soft palate. (B) Effect of oral pressure device on the tongue and soft palate displacement.

Notes: Copyright © 2013. Reproduced with permission from the American College of Chest Physicians. Park JG, Morgenthaler TM, Gay PC. Novel and emerging nonpositive airway pressure therapies for sleep apnea. Chest. 2013;144(6): 1946–1952.Citation19
Figure 2 (A) Winx device, an oral pressure device that uses gentle suction to anteriorly and superiorly displace the tongue and the soft palate. (B) Effect of oral pressure device on the tongue and soft palate displacement.

Oral appliance innovations

Although not a new concept, novel oral appliances, such as tongue repositioning or restraining devices, have recently been tested in people with OSA. A tongue-stabilizing device that uses suction to protrude the tongue was compared with a mandibular advancement device in a randomized crossover study of a small sample of patients recruited from a tertiary hospital sleep clinic.Citation28 Although the AHI was significantly improved with both the mandibular advancement device and the tongue-stabilizing device, compliance was poor for the tongue-stabilizing device, with participants preferring the mandibular advancement device. A full breath solution appliance was examined in adults diagnosed with OSA. This is an oral appliance that has a horizontal transpalatal bar with a vertical extension called the posterior tongue restrainer. The transpalatal bar and tissue make contact above the tongue and below the palatal mucosa when the person is supine during sleep, theoretically restraining and inhibiting the tongue from obstructing the upper airway. In this preliminary study, the mean AHI fell by 73% and the mean SaO2 significantly improved.Citation29 Dort and Remmers tested a combined approach using a mandibular repositioning appliance with an attached tongue-retaining device; the desired response (respiratory disturbance index <10 events per hour) was achieved in 34% of subjects, but many subjects (39%) were nonresponders.Citation30 The authors concluded that patients require several weeks to acclimate to the device and achieve an optimal response, which may explain the poor response rate in this study.

Ngiam and Kyung investigated a novel mandibular advancement device that involved mini-implants and a facemask as anchors.Citation31 This device may be particularly beneficial in cases where there are too few teeth to retain the oral appliances, or in patients who cannot tolerate a bulky intraoral device. Albeit based on only ten subjects, the results showed a significant reduction in AHI and snoring frequency, although design alterations in the face pads were necessary to avoid pressure sores. Almeida et al showed how short-term oral appliances could be a viable treatment option in patients who are adherent to CPAP but need to pause use of the machine in certain circumstances, such as traveling or lack of power supply.Citation32 Having both treatment options provided good outcomes and more convenience for the patients.

One factor enhancing the effectiveness of mandibular advancement device treatment is the ability to titrate the appliance to a patient’s specific needs. Titration protocols may be based on subjective parameters including patient comfort, objective parameters such as polysomnographic data, or both. To date, no consensus has been reached on a standardized titration protocol.Citation33

In order to measure the overall therapeutic effectiveness of an oral appliance completely, objective compliance data must be obtained. To date, most relevant studies have relied on self-reporting. A thermosensor embedded in an oral appliance has been reported to have a high correlation with corresponding subjective reporting.Citation34 This same method was studied long term, and after 1 year the correlation remained the same with respect to self-report.Citation35 Although not currently in widespread clinical use, this provides a safe and effective method for obtaining objective reports on patient adherence. Further studies need to be conducted to determine the optimal battery life, power consumption, and storage capacity of such compliance trackers. Recently, Smith and Verrett investigated a novel device that included a supplementary sensor to enable conservation of power and memory.Citation36 Their study also found a high correlation between subjective compliance reporting and data from their objective monitor. Objective compliance monitoring will be very valuable for long-term studies of oral appliance efficacy as well as interpreting outcomes of treatment.

Prediction of oral appliance treatment outcomes

Although oral appliance use for treating OSA has been studied extensively, it remains difficult to predict which patients will respond favorably to treatment with an oral appliance.Citation37 Moreover, there is a need for prospective long-term studies to evaluate success rates.Citation37 Oropharyngeal crowding and obesity were studied as predictive factors for response to oral appliance therapy. Patients with moderate OSA who have oropharyngeal crowding, identified via Mallampati scoring, and who are obese were identified as poor responders to oral appliance therapy.Citation38 Additional research has been conducted on craniofacial morphologic features as predictors for oral appliance treatment. Cephalometric measurements such as increase in mandibular plane to cranial base angle are related to poor response to oral appliance treatment, whereas shorter soft palate length, increased cranial base angulation, minimal retroglossal airway, short anterior face height, and mandibular retrusion have been related to successful oral appliance treatment. These findings support the view that successful oral appliance treatment is moderated by multiple anatomic factors.Citation39,Citation40 As dental professionals transition to three-dimensional imaging, further studies may be done utilizing three-dimensional landmarks and measurements to provide more accurate predictors of the outcome of oral appliance treatment. Recent studies utilizing magnetic resonance imaging, computed tomography, and cone-beam computed tomography have all provided valuable airway volumetric data that are not provided by two-dimensional cephalograms alone.Citation41Citation43 Future research using volumetric airway imaging may provide valuable insights into predicting individual responses to oral appliance therapy.

Another method that can evaluate lateral airway dimensions aside from anteroposterior ones is nasal endoscopy. This method is promising as it is able to evaluate dynamic airway changes with an oral appliance either during sleep or wakefulness.Citation44,Citation45 In 2014, Sasao et al evaluated a number of patients diagnosed with sleep apnea and subjected them to a nasoendoscopic procedure while advancing the mandible. His group proposed conducting this test during wakefulness due to its clinical practicality compared with performing it overnight during sleep. They found that the orohypopharynx widened in all patients but the velopharynx only widened in approximately 75% of subjects, and that after oral appliance treatment, the AHI scores of those without velopharyngeal widening did not decrease significantly. The use of nasoendoscopy with mandibular advancement may be valuable in the future to evaluate velopharynx widening as a predictor of the efficacy of treatment with an oral appliance.Citation45

Positional therapy

Sleep apnea in over half of patients with OSA is position- dependent,Citation46 defined as having at least double the AHI in the supine position compared with the lateral position.Citation47 Studies of the effect of positional therapy on OSA have shown positive results, ie, lower efficacy than CPAP, but often with better adherence, despite being potentially uncomfortable.Citation48 Recent studies examining the efficacy of position therapy based on position monitoring with a supine alarmCitation49 and a sleep position trainer device worn around the chestCitation50 showed that these new devices were well tolerated and markedly reduced supine sleeping time and AHI. Additional studies of improved positional therapy devices are indicated.

Surgical treatment approaches

Upper airway, facial skeletal, and bariatric surgeries for morbid obesity continue to offer important alternative treatment options for adult OSA. Surgical treatments may be selected based on patient preference and ability to tolerate or adhere to CPAP. Upper airway evaluation with a clinical staging system or dynamically with endoscopy during sleep/sedation, and with technologies such as cine magnetic resonance imaging, offer ways to localize upper airway anatomic compromise and tailor treatment to individual patients.Citation51 Nasal surgery (eg, septoplasty, turbinectomy) can improve tolerability of CPAP therapy in selected patients with OSA and reduce snoring. Minimally invasive surgeries such as palatal implants and radiofrequency or laser-assisted uvulopalatopharyngoplasty are not efficacious in adult OSA and may be considered for primary snoring.Citation52

Many adults do not respond to single operative interventions and benefit from planned multilevel upper airway surgical modification either simultaneously or in staged protocols.Citation53,Citation54 A common staged protocol involves utilizing uvulopalatopharyngoplasty with tongue suspension or with genioglossus advancement and hyoid myotomy (Phase I)Citation55 and maxillary and mandibular advancement osteotomy (Phase II). Several new techniques of upper airway surgical modifications have recently been described. Submucosal lingualplasty in combination with palatal surgery was found to be effective in improving symptoms and AHI in a case series of OSA patients with macroglossia.Citation56 A transcervical technique to access the retrolingual space and perform radiofrequency ablation of hypertrophic lingual tonsils under sedation and local anesthesia has been reported and its effectiveness remains to be established.Citation57 Other newly developed surgical approaches include modified uvulopalatopharyngoplasty and coblation channeling of the tongue, a coblation-assisted Lewis and MacKay operation, and transpalatal advancement.Citation58,Citation59 An investigational technique, typically used within a staged surgical approach, is transoral robotic surgery for tongue base reduction.Citation60

Mandibular advancement osteotomy results in a permanent increase in volume and reduction in length of the upper airway, and the extent of the anterior movements of the maxilla, soft palate, and hyoid are correlated with improvement in AHI.Citation61 The predictability of post-surgical airway change can be improved with virtual surgical planning, a technique that combines computed tomography imaging with software reconstruction and surgical expertise.Citation62 Notwithstanding effects on AHI and symptoms, upper airway soft tissue and skeletal surgical interventions may improve quality of life, driving performance, and mortality rates in appropriately selected adults with OSA.Citation63Citation66

Electrical stimulation

Remmers et al first suggested in 1978 that pharyngeal obstruction was related to loss of genioglossus muscle activity in OSA patients during sleep.Citation67 Since then, several experimental methods to treat OSA have focused on electrically stimulating upper airway muscle activity directly by transcutaneous, intraoral, or intramuscular electrodes (see ).Citation68Citation72 Although these methods decreased the severity of OSA, they also had the unfortunate consequence of induced arousals and concomitant sleep fragmentation related to the electrical stimuli.

Figure 3 Schematic representation of the Inspire™ I system for hypoglossal nerve stimulation in patients with obstructive sleep apnea. The stimulator (implantable pulse generator) is connected to the respiratory sensor by means of a sensor lead. A stimulation lead connects the stimulator to the half-cuff electrode around the hypoglossal nerve.

Notes: Reprinted from J Sleep Res, 23(1), Kezirian EJ, Goding GS Jr, Malhotra A, et al, Hypoglossal nerve stimulation improves obstructive sleep apnea: 12-month outcomes, 77–83, Copyright © 2014, with permission from Elsevier.Citation78
Figure 3 Schematic representation of the Inspire™ I system for hypoglossal nerve stimulation in patients with obstructive sleep apnea. The stimulator (implantable pulse generator) is connected to the respiratory sensor by means of a sensor lead. A stimulation lead connects the stimulator to the half-cuff electrode around the hypoglossal nerve.

A more promising method of stimulating upper airway muscles to treat OSA, which emerged over a decade ago but stalled due to lack of reliable devices,Citation73Citation76 is hypoglossal nerve stimulation (HNS).Citation77Citation81 In HNS, a silicone cuff with stimulating electrodes is surgically placed around a unilateral hypoglossal nerve, which includes motor neurons innervating both protrusor and retractor muscles of the tongue. Appropriate placement of the electrode cuff for activation of the protrusor muscles is confirmed by observing tongue protrusion during stimulation and by electromyographic monitoring during surgery. Stimulating leads are tunneled subplatysmally via the neck to the neurostimulator, which is placed subcutaneously inferior to the clavicle. From the neurostimulator, sensory leads are tunneled subcutaneously to intercostal muscles to monitor respiration.Citation77Citation81

The neurostimulator delivers either synchronousCitation77Citation80 or continuous stimulation.Citation81 In synchronous stimulation, sensory leads detect inspiratory efforts that are analyzed by the neurostimulator. The neurostimulator’s software algorithm predicts the onset of inspiratory effort, delivering stimulation pulses between the end of the expiratory effort and the beginning of the next expiratory phase of each respiratory cycle. Synchronous stimulation systems depolarize the entire hypoglossal nerve during inspiratory effort, causing bulk tongue protrusion.Citation76Citation80 In continuous stimulation, electrodes placed around the hypoglossal nerve are continuously and sequentially cycled from one electrode to the next in a predetermined fashion during the entire respiratory cycle. The focus of continuous stimulation is to increase tonic, rather than phasic, activation of the genioglossus.Citation76,Citation81 In both systems, stimulation periods are maintained on a duty cycle of <50% to prevent neuromuscular fatigue of the tongue.Citation76 After patients are implanted, stimulus titration for the patient occurs by gradually adjusting stimulus intensity (mA), frequency (Hz), and pulse width (μsec) to tolerable levels that consistently abolish inspiratory flow restriction during sleep.Citation76

Five recent HNS trials had similar inclusion and exclusion criteria. Two trials included participants with moderate to severe OSA (pretreatment AHI of 20–100) and BMI ≤40 and two trials included those with an AHI of 20–50 and BMI ≤32.Citation77Citation80 One trial included subjects with AHI ≥20 and BMI between 25 and 40.Citation81 The upper threshold of BMI was necessary because previous investigations showed that HNS had no therapeutic effect in individuals with elevated BMI.Citation78,Citation80 Exclusion criteria were similar in all trials, and included prior surgery of the head or neck, pronounced anatomic abnormalities preventing effective use of HNS, ≥5% of central or mixed apneas, cardiovascular disease, and sleep disorders other than OSA.Citation77Citation81 Two studies further excluded patients if palatal complete concentric collapse was observed during drug-induced sleep endoscopy (DISE) because absence of palatal complete concentric collapse during DISE has been shown to predict the therapeutic success of HNS.Citation34,Citation79,Citation80

In all five studies, AHI was decreased by ≥50% and the oxygen desaturation index significantly improved at various time points (3, 6, or 12 months) after implantation.Citation80 The apnea index and hypopnea index each were significantly decreased.Citation77,Citation78,Citation80 There were significant improvements in symptoms as quantified by improvements in daily functioning, excessive daytime sleepiness, mood, sleep quality, and quality of life.Citation77,Citation78,Citation80 In two trials, there were increases in sleep efficiency and positive changes in sleep architecture, while two other studies showed no differences.Citation77,Citation78,Citation80,Citation81 When OSA patients were taken off HNS therapy for at least 5 days, the AHI and oxygen desaturation index returned back to levels prior to implantation.Citation79 However, there is evidence that one-night off-stimulation resulted in a similar AHI and oxygen desaturation index as the previous on-stimulation night, suggesting that HNS therapy could have residual effects lasting for at least one night.Citation82 This insight suggests the intriguing possibility of alternate night treatment.Citation82

HNS therapy is initially invasive, requiring surgery, and adverse events related to the implantation procedure or stimulation therapy have been reported. A majority of participants reported minor adverse procedural-related events (eg, numbness/pain/swelling/infection at incision site, temporary tongue weakness) that resolved with time, pain medication, and/or antibiotic treatment, or minor adverse therapy-related events (eg, tongue abrasions cause by tongue movement over teeth, discomfort associated with stimulation) that resolved after acclimation.Citation77,Citation78,Citation80,Citation81 Serious adverse events happen infrequently and initially after implantation, and involve serious infection at the incision site requiring device removal or a subsequent surgical procedure to reposition/replace the electrode cuff or to repair malfunctioning equipment.Citation77,Citation78,Citation80,Citation81 There was also a very small percentage of participants who had HNS devices removed due to lack of objective and subjective effectiveness. The rate of system-related or procedural-related serious adverse events was between 2% and 29% after implantation.Citation77Citation80

Phenotyping of upper airway collapse can increase the probability of HNS success. The studies above did exclude OSA patients who had pronounced upper airway abnormalities, like enlarged tonsils or a modified Mallampati score of IV, which would prevent the success of HNS.Citation77Citation79,Citation81 Moreover, phenotyping for absence of palatal complete concentric collapse during DISE improved the success of HNS.Citation79,Citation80,Citation83 However, the lack of uniformity in sedation methods during DISE, which carries its own risks, and the lack of an established consensus on DISE scoring systems needs to be addressed before DISE can be widely used to phenotype OSA patients for HNS.Citation83

In view of mounting evidence for significant reductions in AHI and oxygen desaturation index, and the relatively low incidence of serious adverse events, HNS appears to be a promising alternative therapy for OSA patients with BMI <40 who cannot tolerate or refuse to adhere to CPAP. However, larger-scale clinical studies are needed to assess the long-term safety of HNS. In addition, it will be important to assess the efficacy and safety of HNS in subjects with more severe OSA, including comorbidities, as well as in individuals with BMI >40. A recent update from the Wisconsin Cohort Study has shown an increased prevalence in OSA that is correlated with an increase in BMI.Citation84 A growing number of individuals with OSA remain at a BMI >40, which may limit the use of HNS as a general therapeutic alternative in OSA.

Drug treatment

Oral drugs for treating OSA have focused on five strategies, ie, increasing ventilatory drive (eg, progestagens, theophylline, acetazolamide), increasing upper airway tone (eg, serotonergics, cholinergics), decreasing rapid eye movement sleep (eg, antidepressants, clonidine), increasing arousal threshold (eg, eszopiclone), and/or increasing the cross-sectional area or reducing the surface tension of the upper airway through topical therapy (eg, fluticasone, lubricant).Citation85 A recent Cochrane review assessed three decades of clinical trials of various drugs targeting these strategies and concluded that “some of the drugs may be helpful; however, their tolerability needs to be considered in long-term trials”.Citation85 The focus here is to present recent and innovative drug treatments for OSA.

A recent and novel approach towards increasing upper airway tone to alleviate OSA has been the use of cannabinoids. Dronabinol is a nonselective cannabinoid type 1 (CB1) and type 2 (CB2) receptor agonist used historically to treat anorexia/cachexia associated with acquired immunodeficiency syndrome and to control chemotherapy-induced nausea and vomiting. At doses of 2.5–10 mg per day, dronabinol significantly reduced AHI and improved subjective sleepiness in 17 adults with moderate to severe OSA, in a single-blind dose-escalation proof-of-concept clinical trial.Citation86 However, there were no changes in oxygen desaturation index, and sleep efficiency and arousal index showed improvements only during the first half of the night at the highest dose.Citation86 A more recent secondary analysis of this study demonstrated a shift in the sleep electroencephalogram toward lower frequencies and improved ultradian cycling that was predictive of improved alertness.Citation87 The improvement in AHI was unlikely due to any respiratory stimulant effect because animal studies have shown a decreased respiratory rate across all sleep/wake states after treatment with dronabinol.Citation88 The authors attributed the decrease in AHI to increased upper airway tone and observed a significant improvement in supine AHI.Citation86 This may reflect inhibition of afferent vagal nerve activity, leading to disinhibition of upper airway motor neurons. A recent report showing increased phasic activation of genioglossi, and attenuated serotonin-induced apnea in rats subjected to local nodose ganglion injections of dronabinol supports this interpretation.Citation89 There were no serious adverse events associated with daily oral dronabinol. Common minor adverse events included evening somnolence (after drug administration and before lights out) and increased appetite. Although weight gain was not observed during this 3-week treatment period, this will need to be confirmed by longer-term trials with cannabinoids.Citation86 Larger-scale, randomized controlled long-term studies will be needed to establish the efficacy and safety of cannabinoids in the treatment of OSA.

Another new therapeutic strategy to treat OSA is to pharmaceutically modify loop gain.Citation5 Two studies have shown that some patients with OSA have high loop gain, and therefore, unstable ventilatory systems prone to apneas.Citation90,Citation91 In a recent non-blinded study, patients taking acetazolamide 500 mg twice daily for 7 days exhibited significantly decreased loop gain and AHI when compared with baseline levels.Citation92 This improvement in ventilatory stability was attributed to a reduction in the magnitude of the ventilatory response to arousal. The ventilatory system was desensitized to disturbances contributing to ventilatory instability.Citation92 Unfortunately, long-term usage of acetazolamide is poorly tolerated and might not be a viable treatment option.Citation5 Randomized, double-blind, long-term studies are needed to verify the feasibility of using acetazolamide in the treatment of OSA.

OSA has been linked to gastroesophageal reflux disease (GERD), and several studies suggest that treatment of GERD may decrease AHI and daytime sleepiness.Citation93Citation95 However, these studies recruited individuals primarily for GERD. A recent study consecutively enrolled patients with OSA and investigated the effects of a proton pump inhibitor, lansoprazole, on OSA.Citation96 Ninety-one percent of participants had pathological reflux episodes, and 22% of participants had at least 50% of apneas preceded by a reflux event during polysomnography. Three months of twice-daily lansoprazole decreased AHI, but there were no changes in oxygen desaturation index or sleep efficiency. Proton pump inhibitors are well tolerated, with only weak observational associations between proton pump inhibitor therapy and fractures, pneumonia, mortality, and nutritional deficiencies.Citation97 Evaluation of GERD has the potential to become standard practice during OSA diagnosis.

The drug therapies listed above have potential in treating OSA. However, the long-term tolerability and efficacy of these drugs remain to be demonstrated. Further basic research needs to be conducted to better elucidate the neurobiological mechanisms underlying OSA; bringing fresh insights to suggest new (or to modify previous) treatment strategies in OSA. As progress is made in this area, the hope for safe and effective pharmacotherapies in OSA may become a reality.

Recommendations and conclusion

Few novel treatments have emerged that have proven as efficacious as CPAP treatment, especially for more severe disease. However, the modest level of treatment adherence makes CPAP less effective, underscoring the need for more innovative interventions. For the new approaches that have emerged, few have been robustly evaluated in randomized clinical trials and those that have had small sample sizes and limited outcomes. There is no universal definition of treatment success to guide determination of the efficacy of innovative treatments. However, outcomes that might be included to demonstrate treatment effectiveness would be adequate titration (CPAP and oral appliance), improved oxygenation, reduction in sleepiness (measured by the Epworth Sleepiness Scale), improved daily functioning (measured by the Functional Outcomes of Sleep Questionnaire), enhanced neurobehavior (metrics of sustained attention measured with the Psychomotor Vigilance Task), and improved mood (measured with a mood instrument such as the Profile of Mood States). In addition to providing the definition and outcomes of treatment success, larger clinical trials of novel interventions are needed with an evaluation of long-term treatment outcomes that reflect the variety of impairments associated with OSA.

Disclosure

Dr Carley owns common shares of Cortex Pharmaceuticals, a publicly traded company working to develop pharmacological treatments for sleep apnea syndrome and other breathing disorders, and is an inventor on patents related to pharmacological treatments for sleep apnea syndrome. The other authors report no conflicts of interest in this work.

References

  • YoungTPeppardPEGottliebDJEpidemiology of obstructive sleep apnea: a population health perspectiveAm J Respir Crit Care Med200216591217123911991871
  • Al GhanimNComondoreVRFleethamJMarraCAAyasNTThe economic impact of obstructive sleep apneaLung2008186171218066623
  • SassaniAFindleyLJKrygerMReducing motor-vehicle collisions, costs, and fatalities by treating obstructive sleep apnea syndromeSleep200427345345815164898
  • WeaverTEMaislinGDingesDFRelationship between hours of CPAP use and achieving normal levels of sleepiness and daily functioningSleep200730671171917580592
  • BurgessKRNew insights from the measurement of loop gain in obstructive sleep apnoeaJ Physiol2012590Pt 81781178222532646
  • DellwegDKerlJHoehnEWenzelMKoehlerDRandomized controlled trial of noninvasive positive pressure ventilation (NPPV) versus servoventilation in patients with CPAP-induced central sleep apnea (complex sleep apnea)Sleep20133681163117123904676
  • AuroraRNChowdhuriSRamarKThe treatment of central sleep apnea syndromes in adults: practice parameters with an evidence-based literature review and meta-analysesSleep2012351174022215916
  • SharmaBKBakkerJPMcSharryDGAdaptive servoventilation for treatment of sleep-disordered breathing in heart failure: a systematic review and meta-analysisChest201214251211122122722232
  • WimmsAJRichardsGNBenjafieldAVAssessment of the impact on compliance of a new CPAP system in obstructive sleep apneaSleep Breath2013171697622286779
  • DunganGC2ndMarshallNSHoyosCMYeeBJGrunsteinRRA randomized crossover trial of the effect of a novel method of pressure control (SensAwake) in automatic continuous positive airway pressure therapy to treat sleep disordered breathingJ Clin Sleep Med20117326126721677895
  • WozniakDRLassersonTJSmithIEducational, supportive and behavioural interventions to improve usage of continuous positive airway pressure machines in adults with obstructive sleep apnoeaCochrane Database Syst Rev20141CD00773624399660
  • BartlettDWongKRichardsDIncreasing adherence to obstructive sleep apnea treatment with a group social cognitive therapy treatment intervention: a randomized trialSleep201336111647165424179297
  • DengTWangYSunMChenBStage-matched intervention for adherence to CPAP in patients with obstructive sleep apnea: a randomized controlled trialSleep Breath201317279180122945541
  • SawyerAMKingTSHanlonARisk assessment for CPAP nonadherence in adults with newly diagnosed obstructive sleep apnea: preliminary testing of the Index for Nonadherence to PAP (I-NAP)Sleep Breath352014 [Epub ahead of print.]
  • SchwabRJGefterWBAnatomical factors: insights from imaging studiesPackAISleep Apnea: Pathogenesis, Diagnosis, and TreatmentNew York, NY, USAMarcel Dekker2002
  • MorrellMJArabiYZahnBBadrMSProgressive retropalatal narrowing preceding obstructive apneaAm J Respir Crit Care Med19981586197419819847295
  • YucelAUnluMHaktanirAAcarMFidanFEvaluation of the upper airway cross-sectional area changes in different degrees of severity of obstructive sleep apnea syndrome: cephalometric and dynamic CT studyAJNR Am J Neuroradiol200526102624262916286412
  • ColrainIMBrooksSBlackJA pilot evaluation of a nasal expiratory resistance device for the treatment of obstructive sleep apneaJ Clin Sleep Med20084542643318853699
  • ParkJGMorgenthalerTMGayPCNovel and emerging nonpositive airway pressure therapies for sleep apneaChest201314461946195224297128
  • RosenthalLMassieCADolanDCA multicenter, prospective study of a novel nasal EPAP device in the treatment of obstructive sleep apnea: efficacy and 30-day adherenceJ Clin Sleep Med20095653253720465019
  • BerryRBKrygerMHMassieCAA novel nasal expiratory positive airway pressure (EPAP) device for the treatment of obstructive sleep apnea: a randomized controlled trialSleep201134447948521461326
  • KrygerMHBerryRBMassieCALong-term use of a nasal expiratory positive airway pressure (EPAP) device as a treatment for obstructive sleep apnea (OSA)J Clin Sleep Med201175449B453B22003339
  • PatelAVHwangDMasdeuMJPredictors of response to a nasal expiratory resistor device and its potential mechanisms of action for treatment of obstructive sleep apneaJ Clin Sleep Med201171132221344051
  • WalshJKGriffinKSForstEHA convenient expiratory positive airway pressure nasal device for the treatment of sleep apnea in patients non-adherent with continuous positive airway pressureSleep Med201112214715221256800
  • SawyerAMGooneratneNSMarcusCLOferDRichardsKCWeaverTEA systematic review of CPAP adherence across age groups: clinical and empiric insights for developing CPAP adherence interventionsSleep Med Rev201115634335621652236
  • Farid-MoayerMSiegelLCBlackJA feasibility evaluation of oral pressure therapy for the treatment of obstructive sleep apneaTher Adv Respir Dis20137131223184709
  • ColrainIMBlackJSiegelLCA multicenter evaluation of oral pressure therapy for the treatment of obstructive sleep apneaSleep Med201314983083723871259
  • DeaneSACistulliPANgATComparison of mandibular advancement splint and tongue stabilizing device in obstructive sleep apnea: a randomized controlled trialSleep200932564865319480232
  • SinghGDKeropianBPillarGEffects of the full breath solution appliance for the treatment of obstructive sleep apnea: a preliminary studyCranio200927210911719455922
  • DortLRemmersJA combination appliance for obstructive sleep apnea: the effectiveness of mandibular advancement and tongue retentionJ Clin Sleep Med20128326526922701383
  • NgiamJKyungHMMicroimplant-based mandibular advancement therapy for the treatment of snoring and obstructive sleep apnea: a prospective studyAngle Orthod201282697898422594775
  • AlmeidaFRMulgrewAAyasNMandibular advancement splint as short-term alternative treatment in patients with obstructive sleep apnea already effectively treated with continuous positive airway pressureJ Clin Sleep Med20139431932423585745
  • DieltjensMVandervekenOMHeyningPHBraemMJCurrent opinions and clinical practice in the titration of oral appliances in the treatment of sleep-disordered breathingSleep Med Rev201216217718522033170
  • VandervekenOMDieltjensMWoutersKObjective measurement of compliance during oral appliance therapy for sleep-disordered breathingThorax2013681919622993169
  • DieltjensMBraemMJVroegopAVObjectively measured vs self-reported compliance during oral appliance therapy for sleep-disordered breathingChest201314451495150223928873
  • SmithYVerrettREvaluation of a novel device for measuring patient compliance with oral appliances in the treatment of obstructive sleep apneaJ Prosthodont2014231313823889695
  • ChenHLoweAAUpdates in oral appliance therapy for snoring and obstructive sleep apneaSleep Breath201317247348622562263
  • TsuikiSItoEIsonoSOropharyngeal crowding and obesity as predictors of oral appliance treatment response to moderate obstructive sleep apneaChest2013144255856323493981
  • ShenHLWenYWChenNHLiaoYFCraniofacial morphologic predictors of oral appliance outcomes in patients with obstructive sleep apneaJ Am Dent Assoc2012143111209121723115150
  • NgATDarendelilerMAPetoczPCistulliPACephalometry and prediction of oral appliance treatment outcomeSleep Breath2012161475821424283
  • ChoudhuryMPadmanabhanTVA preliminary report on the effect of a mandibular advancement device on obstructive sleep apnea using magnetic resonance imaging and polysomnographyInt J Prosthodont201225661361823101041
  • FuruhashiAYamadaSShiomiTEffective three-dimensional evaluation analysis of upper airway form during oral appliance therapy in patients with obstructive sleep apnoeaJ Oral Rehabil201340858258923692649
  • AlsufyaniNAAl-SalehMAMajorPWCBCT assessment of upper airway changes and treatment outcomes of obstructive sleep apnoea: a systematic reviewSleep Breath201317391192323344828
  • JohalABattagelJMKotechaBTSleep nasendoscopy: a diagnostic tool for predicting treatment success with mandibular advancement splints in obstructive sleep apnoeaEur J Orthod200527660761416049036
  • SasaoYNoharaKOkunoKNakamuraYSakaiTVideoendoscopic diagnosis for predicting the response to oral appliance therapy in severe obstructive sleep apneaSleep Breath2152014 [Epub ahead of print.]
  • OksenbergASilverbergDSAronsERadwanHPositional vs nonpositional obstructive sleep apnea patients: anthropomorphic, nocturnal polysomnographic, and multiple sleep latency test dataChest199711236296399315794
  • CartwrightRDEffect of sleep position on sleep apnea severitySleep1984721101146740055
  • RaveslootMvan MaanenJDunLde VriesNThe undervalued potential of positional therapy in position-dependent snoring and obstructive sleep apnea-a review of the literatureSleep Breath2013171394922441662
  • BignoldJJMercerJDAnticNAMcEvoyRDCatchesidePGAccurate position monitoring and improved supine-dependent obstructive sleep apnea with a new position recording and supine avoidance deviceJ Clin Sleep Med20117437638321897774
  • van MaanenJPMeesterKADunLNThe sleep position trainer: a new treatment for positional obstructive sleep apnoeaSleep Breath201317277177922927107
  • FriedmanMIbrahimHJosephNJStaging of obstructive sleep apnea/hypopnea syndrome: a guide to appropriate treatmentLaryngoscope2004114345445915091218
  • AuroraRNCaseyKRKristoDPractice parameters for the surgical modifications of the upper airway for obstructive sleep apnea in adultsSleep201033101408141321061864
  • CaplesSMRowleyJAPrinsellJRSurgical modifications of the upper airway for obstructive sleep apnea in adults: a systematic review and meta-analysisSleep201033101396140721061863
  • LinHCFriedmanMChangHWGurpinarBThe efficacy of multilevel surgery of the upper airway in adults with obstructive sleep apnea/hypopnea syndromeLaryngoscope2008118590290818300704
  • HandlerEHamansEGoldbergANMickelsonSTongue suspension: an evidence-based review and comparison to hypopharyngeal surgery for OSALaryngoscope2014124132933623729234
  • GunawardenaIRobinsonSMacKaySSubmucosal lingualplasty for adult obstructive sleep apneaOtolaryngol Head Neck Surg2013148115716523014996
  • BlumenMCoquilleFChabolleFLingual tonsil reduction in OSA: transcervical radiofrequency ablationEur Ann Otorhinolaryngol Head Neck Dis2012129633934223073498
  • ShineNPLewisRHTranspalatal advancement pharyngoplasty for obstructive sleep apnea syndrome: results and analysis of failuresArch Otolaryngol Head Neck Surg2009135543443819451461
  • MacKaySGCarneyASWoodsCModified uvulopalatopharyngoplasty and coblation channeling of the tongue for obstructive sleep apnea: a multi-centre Australian trialJ Clin Sleep Med20139211712423372463
  • LinHSRowleyJABadrMSTransoral robotic surgery for treatment of obstructive sleep apnea-hypopnea syndromeLaryngoscope201312371811181623553290
  • HsiehYJLiaoYFChenNHChenYRChanges in the calibre of the upper airway and the surrounding structures after maxillomandibular advancement for obstructive sleep apnoeaBr J Oral Maxillofac Surg201452544545124629456
  • BarreraJEVirtual surgical planning improves surgical outcome measures in obstructive sleep apnea surgeryLaryngoscope201412451259126624357526
  • CamachoMCertalVBrietzkeSETracheostomy as treatment for adult obstructive sleep apnea: a systematic review and meta-analysisLaryngoscope2014124380381124549987
  • WeaverEMWoodsonBTYuehBStudying Life Effects and Effectiveness of Palatopharyngoplasty (SLEEP) study: subjective outcomes of isolated uvulopalatopharyngoplastyOtolaryngol Head Neck Surg2011144462363121493246
  • HaraldssonPOCarenfeltCLysdahlMTornrosJLong-term effect of uvulopalatopharyngoplasty on driving performanceArch Otolaryngol Head Neck Surg1995121190947803027
  • BoydSBWaltersASSongYWangLComparative effectiveness of maxillomandibular advancement and uvulopalatopharyngoplasty for the treatment of moderate to severe obstructive sleep apneaJ Oral Maxillofac Surg201371474375123219145
  • RemmersJEdeGrootWJSauerlandEKAnchAMPathogenesis of upper airway occlusion during sleepJ Appl Physiol Respir Environ Exerc Physiol1978446931938670014
  • DeckerMJHaagaJArnoldJLAtzbergerDStrohlKPFunctional electrical stimulation and respiration during sleepJ Appl Physiol1993753105310618226511
  • EdmondsLCDanielsBKStansonAWSheedyPF3rdShepardJWJrThe effects of transcutaneous electrical stimulation during wakefulness and sleep in patients with obstructive sleep apneaAm Rev Respir Dis19921464103010361416392
  • GuilleminaultCPowellNBowmanBStoohsRThe effect of electrical stimulation on obstructive sleep apnea syndromeChest1995107167737813315
  • MikiHHidaWChonanTKikuchiYTakishimaTEffects of submental electrical stimulation during sleep on upper airway patency in patients with obstructive sleep apneaAm Rev Respir Dis19891405128512892817590
  • SteierJSeymourJRaffertyGFContinuous transcutaneous submental electrical stimulation in obstructive sleep apnea: a feasibility studyChest20111404998100721454399
  • EiseleDWSchwartzARSmithPLTongue neuromuscular and direct hypoglossal nerve stimulation for obstructive sleep apneaOtolaryngol Clin North Am200336350151012956097
  • IlomakiJBaerGAKarhuketoTTalonenPPuhakkaHPharyngeal patency caused by stimulation of the hypoglossal nerve in anaesthesia-relaxed patientsActa Otolaryngol Suppl19975292102119288312
  • OlivenAO’HearnDJBoudewynsAUpper airway response to electrical stimulation of the genioglossus in obstructive sleep apneaJ Appl Physiol20039552023202914555669
  • SchwartzARSmithPLOlivenAElectrical stimulation of the hypoglossal nerve: a potential therapyJ Appl Physiol2014116333734423813525
  • EastwoodPRBarnesMWalshJHTreating obstructive sleep apnea with hypoglossal nerve stimulationSleep201134111479148622043118
  • KezirianEJGodingGSJrMalhotraAHypoglossal nerve stimulation improves obstructive sleep apnea: 12-month outcomesJ Sleep Res2014231778324033656
  • StrolloPJJrSooseRJMaurerJTUpper-airway stimulation for obstructive sleep apneaN Engl J Med2014370213914924401051
  • Van de HeyningPHBadrMSBaskinJZImplanted upper airway stimulation device for obstructive sleep apneaLaryngoscope201212271626163322549513
  • MwengeGBRombauxPDuryMLengeleBRodensteinDTargeted hypoglossal neurostimulation for obstructive sleep apnoea: a 1-year pilot studyEur Respir J201341236036722599356
  • RodensteinDRombauxPLengeleBDuryMMwengeGBResidual effect of THN hypoglossal stimulation in obstructive sleep apnea: a disease-modifying therapyAm J Respir Crit Care Med2013187111276127823725628
  • VandervekenOMMaurerJTHohenhorstWEvaluation of drug-induced sleep endoscopy as a patient selection tool for implanted upper airway stimulation for obstructive sleep apneaJ Clin Sleep Med20139543343823674933
  • PeppardPEYoungTBarnetJHPaltaMHagenEWHlaKMIncreased prevalence of sleep-disordered breathing in adultsAm J Epidemiol201317791006101423589584
  • MasonMWelshEJSmithIDrug therapy for obstructive sleep apnoea in adultsCochrane Database Syst Rev20135CD00300223728641
  • PrasadBRadulovackiMGCarleyDWProof of concept trial of dronabinol in obstructive sleep apneaFront Psychiatry20134123346060
  • FarabiSSPrasadBQuinnLCarleyDWImpact of dronabinol on quantitative electroencephalogram (qEEG) measures of sleep in obstructive sleep apnea syndromeJ Clin Sleep Med2014101495624426820
  • CarleyDWPaviovicSJanelidzeMRadulovackiMFunctional role for cannabinoids in respiratory stability during sleepSleep200225439139812071539
  • CalikMWRadulovackiMCarleyDWIntranodose ganglion injections of dronabinol attenuate serotonin-induced apnea in Sprague-Dawley ratRespir Physiol Neurobiol2014190202424121138
  • EckertDJWhiteDPJordanASMalhotraAWellmanADefining phenotypic causes of obstructive sleep apnea. Identification of novel therapeutic targetsAm J Respir Crit Care Med20131888996100423721582
  • SalloumARowleyJAMateikaJHChowdhuriSOmranQBadrMSIncreased propensity for central apnea in patients with obstructive sleep apnea: effect of nasal continuous positive airway pressureAm J Respir Crit Care Med2010181218919319762565
  • EdwardsBAConnollyJGCampanaLMAcetazolamide attenuates the ventilatory response to arousal in patients with obstructive sleep apneaSleep201336228128523372276
  • BortolottiMGentiliniLMorselliCGiovanniniMObstructive sleep apnoea is improved by a prolonged treatment of gastrooesophageal reflux with omeprazoleDig Liver Dis2006382788116256447
  • SeniorBAKhanMSchwimmerCRosenthalLBenningerMGastroesophageal reflux and obstructive sleep apneaLaryngoscope2001111122144214611802013
  • SuurnaMVWelgeJSurdulescuVKushnerJStewardDLRandomized placebo-controlled trial of pantoprazole for daytime sleepiness in GERD and obstructive sleep disordered breathingOtolaryngol Head Neck Surg2008139228629018656731
  • ErmisFAkyuzFAriciSEffect of proton pump inhibitor (PPI) treatment in obstructive sleep apnea syndrome: an esophageal impedance-pHmetry studyHepatogastroenterology201158110–1111566157321940322
  • ReimerCSafety of long-term PPI therapyBest Pract Res Clin Gastroenterol201327344345423998981