45
Views
1
CrossRef citations to date
0
Altmetric
Review

Dynamic contour tonometry in asymmetric glaucoma patients

, , , , , & show all
Pages 555-559 | Published online: 11 Apr 2012

Abstract

Background

The purpose of this study was to determine any difference in dynamic contour tonometry and ocular pulse amplitude in asymmetric glaucoma patients with the same applanation intraocular pressure.

Methods

This is a prospective, observational study of 30 glaucoma patients and 11 controls from June 2007 to February 2008. Most of the glaucoma patients were on prostaglandin analog treatment.

Results

Mean applanation intraocular pressure in the control group was 14.28 mmHg for the right eye and 14.10 mmHg for the left eye (P > 0.05). Corneal thickness was 519.10 μm for the right eye and 511.07 μm for the left eye (P > 0.05). Mean dynamic contour tonometry intraocular pressure was 17.28 mmHg for the right eye and 17.25 mmHg for the left eye (P > 0.05). Mean ocular pulse amplitude was 2.80 mmHg for the right eye and 2.92 mmHg for the left eye (P > 0.05).

Conclusion

No differences in ocular pulse amplitude were found between the two groups and between the worst and the best eye. In spite of there being no difference in ocular pulse amplitude, dynamic contour tonometry intraocular pressure was 2.44 mmHg higher in the worst eye than in the best eye in the glaucoma patients, even with the same applanation intraocular pressure. Further studies are needed to confirm if this difference is related to glaucoma progression or a worst prognosis and whether it can be considered to be a new risk factor.

Introduction

Glaucoma is a multifactorial disease associated with optic nerve damage, visual field defects, and often elevated intraocular pressure. Intraocular pressure remains the major risk factor for onset of glaucoma and its progression. Nevertheless, there are several aspects of intraocular pressure that have been researched to clarify its role in the pathology of glaucoma. Central cornea thickness, bioelasticity, and intraocular pulse amplitude are points of special interest.Citation1 Dynamic contour tonometry is a device capable of measuring intraocular pressure with less central cornea influence and ocular pulse amplitude.Citation2 The main purpose of this study is to determine if there can be any differences in dynamic contour tonometry and ocular pulse amplitude in asymmetric glaucoma patients with the same applanation intraocular pressure.

Materials and methods

This was a prospective, observational study of 30 glaucoma patients and 11 controls between June 2007 and February 2008. Inclusion criteria were: the same intraocular pressure on applanation tonometry in both eyes; a difference of ≥6 dB in mean deviation of visual field; a difference ≥0.2 in cup-to-disc ratio; and a dynamic contour tonometry measurement quality score reading of 1 or 2 in glaucoma patients. Dynamic contour tonometry intraocular pressure, ocular pulse amplitude, corneal thickness, applanation tonometry, and fundus examination was undertaken in both groups. Most of the glaucoma patients were on prostaglandin analog treatment. Statistical analyses were performed using the Kolmogorov-Smirnov and Shapiro-Wilk tests and Pearson’s linear correlation. Results with P < 0.05 were accepted as being statistically significant. The study was conducted in accordance with the clinical research regulations prevailing in Brazil (resolution 196/96) and with the ethical standards stated in the 1964 Declaration of Helsinki regarding ethical principles for medical research involving human subjects.

Results

The mean age of participants in this study was 61.23 years. Twenty-four female and 17 male patients were enrolled. Twenty-three patients were Hispanic and 18 were Black.

In the controls, mean applanation intraocular pressure was 14.28 mmHg for the right eye and 14.10 mmHg for the left eye (P > 0.05). Corneal thickness was 519.10 μm for the right eye and 511.07 μm for the left eye (P > 0.05). Mean dynamic contour tonometry intraocular pressure was 17.28 mmHg for the right eye and 17.25 mmHg for the left eye (P > 0.05). Mean ocular pulse amplitude was 2.80 mmHg for the right eye and 2.92 mmHg for the left eye (P > 0.05, ).

Table 1 Ocular parameters in control subjects

In glaucoma patients, mean applanation intraocular pressure was 15.70 mmHg for the worst eye and 14.93 mmHg for the best eye (P = 0.078, , ). Mean dynamic contour tonometry intraocular pressure was 20.63 mmHg for the worst eye and 18.19 mmHg for the best eye. This difference was statistically significant (P < 0.05, , ). Mean ocular pulse amplitude was 2.76 mmHg for the worst eye and 2.68 mmHg for the best eye (P = 0.501, , ). Corneal thickness was 516.60 μm for the worst eye and 507.07 μm for the best eye (P > 0.05, , ).

Figure 1 The mean applanation IOP was 15.70 mmHg for the worst eye and 14.93 mmHg for the best eye (P = 0.078).

Abbreviation: IOP, intraocular pressure.
Figure 1 The mean applanation IOP was 15.70 mmHg for the worst eye and 14.93 mmHg for the best eye (P = 0.078).

Figure 2 The mean DCT IOP was 20.63 mmHg for the worst eye and 18.19 mmHg for the best eye (P < 0.05).

Abbreviations: DCT, dynamic contour tonometry; IOP, intraocular pressure.
Figure 2 The mean DCT IOP was 20.63 mmHg for the worst eye and 18.19 mmHg for the best eye (P < 0.05).

Figure 3 The mean OPA was 2.76 mmHg for the worst eye and 2.68 mmHg for the best eye (P = 0.501).

Abbreviation: OPA, ocular pulse amplitude.
Figure 3 The mean OPA was 2.76 mmHg for the worst eye and 2.68 mmHg for the best eye (P = 0.501).

Figure 4 Corneal thickness was 516.60 μm for the worst eye and 507.07 μm for the best eye (P < 0.05).

Abbreviation: CCT, central corneal thickness.
Figure 4 Corneal thickness was 516.60 μm for the worst eye and 507.07 μm for the best eye (P < 0.05).

Table 2 Applanation tonometry in glaucoma patients (mmHg)

Table 3 Dynamic contour tonometry intraocular pressure (mmHg)

Table 4 Ocular amplitude pulse (mmHg)

Table 5 Central corneal thickness (μm)

Discussion

Some studies report that thinner corneas are more likely to be associated with glaucoma probably because of underestimation of intraocular pressure measurement by applanation tonometry. The Ocular Hypertension Treatment Study showed that decreased corneal thickness is associated with a higher risk of development of glaucoma in patients with ocular hypertension.Citation3 Studies regarding elasticity and other properties of the cornea are still unclear. The pulsatility of intraocular pressure and its correlation with glaucoma also remains unclear. Some authors have reported different results for ocular pulse amplitude measurement.

Weizer et al showed that increased ocular pulse amplitude seems to correlate with less severe glaucoma and with increased central corneal thickness.Citation4 Punjabi et al published a study comparing intraocular pressure and ocular pulse amplitude in different types of glaucoma using dynamic contour tonometry and concluded that ocular pulse amplitude was higher in patients with ocular hypertension.Citation5 Romppainen et al found that subjects with ocular hypertension showed significantly higher ocular pulse amplitude values (3.6 ± 1.3 mmHg) than healthy eyes (3.1 ± 1.4 mmHg) and eyes with low-tension glaucoma (2.9 ± 1.4 mmHg). After trabeculectomy, the values were significantly lower (2.4 ± 1.3 mmHg) than in healthy eyes.Citation6 Stalmans et al studied the ocular pulse amplitude in normal tension and primary open-angle glaucoma and concluded that ocular pulse amplitude was reduced in patients with primary open-angle glaucoma and normal tension compared with healthy controls. Ocular pulse amplitude is influenced by intraocular pressure, but not by corneal thickness.Citation7 Sullivan-Mee et al in a very similar study of patients with asymmetric glaucoma demonstrated that dynamic contour tonometry intraocular pressure was significantly higher in eyes with a higher AGIS score than in eyes with a lower AGIS score (16.3 versus 15.5 mmHg, P = 0.004), but applanation intraocular pressure was not significantly different in the same eyes (14.5 mmHg versus 14.4 mmHg, P = 0.56).

These findings suggest that dynamic contour tonometry intraocular pressure is more related to the extension of damage of glaucoma than is applanation intraocular pressure. The most likely explanation for these results is that applanation intraocular pressure systematically underestimates intraocular pressure compared with dynamic contour tonometry intraocular pressure.Citation8

In our study, ocular pulse amplitude was similar in both asymmetric eyes and in controls. However, dynamic contour tonometry intraocular pressure was significantly different between the worst and best eye in glaucoma patients. We found a 2.44 mmHg higher dynamic contour tonometry intraocular pressure in the worst eye than in the best eye, despite the same applanation intraocular pressure. However, no significant dynamic contour tonometry difference was observed in the controls.

Perhaps these results can be explained in light of the underestimation of applanation tonometry intraocular pressure compared with dynamic contour tonometry intraocular pressure. Further, some as yet unknown corneal factors could be responsible for this difference between worst and best eyes with glaucoma. This difference could represent a new risk factor for glaucoma prognosis, onset, or progression.

Conclusion

No differences in ocular pulse amplitude were found between the groups or between the worst and the best eye. In spite of the lack of difference in ocular pulse amplitude, dynamic contour tonometry intraocular pressure was 2.44 mmHg higher in the worst eye than in the best eye in glaucoma patients, even with the same applanation intraocular pressure. Further studies are needed to confirm if this difference is related to glaucoma progression and/or a worse prognosis, and whether dynamic contour tonometry intraocular pressure should be considered as a new risk factor.

Disclosure

The authors report no conflicts of interest in this work.

References

  • HagerALogeKSchroederBFüllhasMOWiegandWEffect of central corneal thickness and corneal hysteresis on tonometry as measured by dynamic contour tonometry, ocular response analyzer, and Goldmann tonometry in glaucomatous eyesJ Glaucoma2008177604
  • BoehmAGWeberAPillunatLEKochRSpoerlEDynamic contour tonometry in comparison to intracameral IOP measurementsInvest Ophthalmol Vis Sci20084962472247718316699
  • BrandtJDBeiserJAKassMAGordonMOCentral corneal thickness in the Ocular Hypertension Treatment Study (OHTS)Ophthalmology2001108101779178811581049
  • WeizerJSAsraniSStinnettSSHerndonLWThe clinical utility of dynamic contour tonometry and ocular pulse amplitudeJ Glaucoma200716870070318091458
  • PunjabiOSHoHKKniestedtCBostromAGStamperRLLinSCIntraocular pressure and ocular pulse amplitude comparisons in different types of glaucoma using dynamic contour tonometryCurr Eye Res2006311085186217050277
  • RomppainenTKniestedtCBachmannLMStürmerJOcular pulse amplitude: a new biometrical parameter for the diagnose of glaucoma?Ophthalmologe20071043230235 German17323044
  • StalmansIHarrisAVanbellinghenVZeyenTSieskyBOcular pulse amplitude in normal tension and primary open angle glaucomaJ Glaucoma200817540340718703952
  • Sullivan-MeeMHalversonKDQuallsCClinical comparison of Pascal dynamic contour tonometry and Goldmann applanation tonometry in asymmetric open-angle glaucomaJ Glaucoma200716869469918091457