Correlation between epithelial thickness in normal corneas, untreated ectatic corneas, and ectatic corneas previously treated with CXL; is overall epithelial thickness a very early ectasia prognostic factor?

, &
Pages 789-800 | Published online: 23 May 2012

Video abstract

Point your SmartPhone at the code above. If you have a QR code reader the video abstract will appear. Or use:

http://dvpr.es/JszOGf

Purpose

To determine and correlate epithelial corneal thickness (pachymetric) measurements taken with a digital arc scanning very high frequency ultrasound biomicroscopy (HF UBM) imaging system (Artemis-II), and compare mean and central epithelial thickness among normal eyes, untreated keratoconic eyes, and keratoconic eyes previously treated with collagen crosslinking (CXL).

Methods

Epithelial pachymetry measurements (topographic mapping) were conducted on 100 subjects via HF UBM. Three groups of patients were included: patients with normal eyes (controls), patients with untreated keratoconic eyes, and patients with keratoconic eyes treated with CXL. Central, mean, and peripheral corneal epithelial thickness was examined for each group, and a statistical study was conducted.

Results

Mean, central, and peripheral corneal epithelial thickness was compared between the three groups of patients. Epithelium thickness varied substantially in the keratoconic group, and in some cases there was a difference of up to 20 μm between various points of the same eye, and often a thinner epithelium coincided with a thinner cornea. However, on average, data from the keratoconic group suggested an overall thickening of the epithelium, particularly over the pupil center of the order of +3 μm, while the mean epithelium thickness was on average +1.1 μm, compared to the control population (P = 0.005). This overall thickening was more pronounced in younger patients in the keratoconic group. Keratoconic eyes previously treated with CXL showed, on average, virtually the same average epithelium thickness (mean −0.7 μm, −0.2 μm over the pupil center, −0.9 μm over the peripheral zone) as the control group. This finding further reinforces our novel theory of the “reactive” component of epithelial thickening in corneas that are biomechanically unstable, becoming stable when biomechanical rigidity is accomplished despite persistence of cornea topographic irregularity.

Conclusion

A highly irregular epithelium may be suggestive of an ectatic cornea. Our results indicate that the epithelium is thinner over the keratoconic protrusion, but to a much lesser extent than anticipated, and on average epithelium is thicker in this group of patients. This difference appears to be clinically significant and may become a screening tool for eyes suspected for ectasia.

Introduction

Importance of corneal epithelium imaging

The contribution of the corneal epithelium to the refractive power of the cornea, and thus ocular refraction, cannot be ignored. Studies have shown that epithelial refractive power alone is an average of 1.03 D (range 0.55–1.85 D) over the central 2 mm diameter zoneCitation1 and 0.85 D (range 0.29–1.60 D) at the 3.6 mm diameter zone.Citation1

Knowledge of the specific epithelium thickness distribution may prove beneficial in close call clinical judgments and aid in the safe screening of a candidate for excimerlaser corneal refractive surgery. The reason for this is that the epithelium does not have homogeneous depth (mean 53.4 ± 4.6 μm) over the Bowman’s layer.Citation2 The epithelium tends to compensate for stromal surface irregularities, ie, by being thicker over the “valleys” and thinner over the “hills”, where the stroma protrudes. Central epithelial thickness in normal eyes has been reported to be thicker than those of keratoconic eyes (mean difference +2.1 μm), and the epithelium is thinner superiorly than inferiorly in normal eyes (mean difference −1.4 μm).Citation2 It has been therefore suggested that the epithelium attempts to minimize abrupt changes in stromal thickness. As a result, the cornea may project a smoother topography and an alleviated wavefront error map, compared to what the underlying anterior stroma alone would show.

In the specific case of subclinical corneal ectasia, if the epithelium is thinner over the ectatic area, profile maps may help a clinician who would otherwise be reading a total corneal pachymetry map to identify patients in whom corneal ablative procedures are contraindicated.Citation3 However, if the epithelium is thicker centrally, an assessment based only on total corneal pachymetry – topography, with no knowledge of the specific epithelial depth – may result in an incorrect assessment of keratoconic progression.

Epithelium thickness irregularities can also be detected in contact lens wearers, who are likely to exhibit epithelium wrappingCitation4,Citation5 or arcuate lesions,Citation6 thus leading to incorrect assessment of their true refraction. Epithelium properties have also been investigated in eyes of patients with diabetes mellitus.Citation7,Citation8 In addition, a large number of refractive errors after laser refractive cornea surgery such as LASIK may be explained by epithelial factors.

For the reasons described above, devices providing precise and accurate in vivo measurements of epithelial thickness distribution over the entire cornea are very important.

Optical-based anterior segment imaging technologies

While the most advanced and established corneal imaging systems are mainly optical, they cannot provide visualization of epithelial thickness, ie, they cannot discern epithelial from stromal components within the cornea because corneal layers have minimal refractive index changes in order to prevent internal reflections.

The two exceptions are anterior segment optical coherence tomography (AS OCT)Citation9Citation11 and confocal microscopy.Citation12,Citation13 To a lesser degree, optical pachymetryCitation14 and through focusing confocal microscopyCitation15 are also capable of discerning epithelium. All these methods, however, are limited to specific points, ie, spot measurements or a single meridian B-scan of the cornea, thus are not suitable for providing a full area visualization of the epithelial thickness map.

Spectral domain (SD) OCTCitation16 signal processing employs Fourier transformation of the optical spectrum of a low coherence interferometer, enabling excellent axial resolution (how close together along the direction of beam propagation two points can be distinguishable from each other) down to 5 μm. Lateral (or transverse) resolution (ability to distinguish two points located normal to each other along the direction of propagation) is limited by ocular optics, the numerical aperture of the illuminating beam, and the number of A-scans employed in the construction of the B-scan.

Currently, there are a number of SD OCT systems commercially available with anterior segment imaging capabilities. These include the Visante (Carl Zeiss Meditec AG, Jena, Germany), Optovue RTVue (Optovue Inc, Fremont, CA), OCT/SLO Combination Imaging System (Optos Inc, Marlborough, MA), and SS-1000 CASIA (Tomey Corporation, Nishi-Ku, Nagoya, Japan).Citation17

The high resolution and precision offered by these systems enables local visualization and segmentation of anterior corneal and epithelial boundaries in a specific meridional scan. Typically, such systems provide corneal pachymetric maps, as well as high resolution cross-sectional images across a meridian of choice. In many of these systems, meridional scan images can be used to calculate epithelial thickness via the caliper tool.Citation18,Citation19 However, to produce epithelial topography (thickness profile) continuously over an area covering the entire cornea, dedicated software to interpolate thickness profile calculated from each meridian is required.Citation20 To the best of our knowledge, this is not offered commercially for any clinical imaging system.

Ultrasound-based anterior segment imaging technologies

Real-time in vivo epithelial mapping is possible with high frequency ultrasound biomicroscopy (HF UBM) systems.Citation21Citation23 The major advantage of ultrasound over optically based systems is that they are not hindered by optically opaque intervening ocular structures. Ultrasound systems, however, require fluid coupling to the eye, and thus necessitate more effort by the examiner and a higher degree of compliance from the patient.

The inception of technology for use in ocular imaging dates back to the 1950s with the work of Mundt and HughesCitation24 (A-scan) and Baum and GreenwoodCitation25 (B-scan). A-scan systems are mainly contact (and thus require local corneal anesthesia), whereas B-scan systems involve scanning. The latter are advantageous over the former, which suffer from variability largely related to decentration, oblique incidence which may cause spuriously thicker measurements, possible corneal compression by the probe, and risk of epithelium scarring, corneal injury, and infection. B-scans require mechanical scanning of the acoustic transducer and setting pixel intensity in proportion to echo amplitude along each line of sight. However, for many years the axial resolution of these systems had been limited because the center frequency had been restricted to the vicinity of 10 MHz.

The evolution of conventional B-scanning, termed UBM, appeared much later.Citation26 Coupled with the introduction of high frequencies in the range of 35–50 MHz, as well as improvements in scanning, data processing, and display, UBM enabled substantial improvements in both axial and lateral resolution down to 40 μm or less. The first UBM system which allowed meaningful clinical application of corneal imaging was developed by Foster and colleagues in the early 1990s, resulting in the ultrasound biomicroscope manufactured by Zeiss-Humphrey Instruments (San Leandro, CA).Citation27 Almost simultaneously, Silverman and colleagues independently developed a digital UBM arc scanner system emphasizing the processing of raw echo data in a noncontact immersion method.Citation21 This system was eventually commercialized as the Artemis, which has received United States Food and Drug Administration approval and is commercially available.

The Artemis system uses a lithium niobate transducer operating on a broad band 50 MHz very high frequency. The cornea is offset from the probe as the patient places the eye over a disposable soft rimmed eye cup, and the area is then filled by sterile normal saline solution at room temperature. The ultrasound transducer and scan mechanism are submerged in deionized water (27°C–33°C) within the body of the scanner and separated from the eye by the transparent membrane provided by the disposable eye cup. In the raw data (A-scans) the local maxima correspond to the interfaces where there is a localized chance of impedance, such as the epithelium–immersion water, Bowman’s interface–anterior stroma, and inferior stroma (including Descemet’s membrane)–aqueous humor. The system claims resolution of 21 μm, whereas the precision of measurement varies according to position within the cornea, with 0.5 μm at the center and less than 1.3 μm peripherally.Citation28 Examples of B-scan images obtained via the Artemis II are presented in .

Figure 1 B-scan reconstruction, as obtained from the HF UBM system showing epithelium and corneal thickness measurements via the caliper tool. We observe the anterior and posterior cornea, as well as the cornea–epithelium interface. (A) Full scan, scale 8.4 μm/pixel. (B) Detail.

Figure 1 B-scan reconstruction, as obtained from the HF UBM system showing epithelium and corneal thickness measurements via the caliper tool. We observe the anterior and posterior cornea, as well as the cornea–epithelium interface. (A) Full scan, scale 8.4 μm/pixel. (B) Detail.

Visualization of epithelium mapping can be quite valuable in the decision making process as well as in the short-and long-term postoperative assessment of excimer-laser corneal refractive surgery procedures. It is also valuable in the assessment of crosslinking (CXL) postoperative epithelial healing, and keratoconus (KCN) and pellucid marginal degeneration screening. Despite its large potential, epithelium mapping is underreported amongst researchers and underused as a diagnostic tool by clinicians, compared to total corneal pachymetry (eg, mapping the whole cornea). This is because current optically-based corneal imaging systems have limited capability to discern epithelial from stromal components within the cornea over the entire corneal area.

The scope of our work is to explore the efficacy of a high frequency arc scanning UBM system, namely the Artemis II + superior (Artemis Medical Technologies Inc, Vancouver, British Columbia, Canada) in the task of clinical epithelial corneal mapping in normal corneas, and keratoconic corneas either nontreated or treated with collagen CXL.

Materials and methods

Epithelial evaluation via 3-D mapping (surface pachymetric measurements) was conducted with the Artemis II + superior. This study was approved by the Ethics Committee. Informed consent for all testing was obtained from each volunteer. To screen for ocular abnormalities each patient underwent a complete ocular examination, including manifest refraction, autorefraction and keratometry measurements. (Auto Refkeratometer Speedy-K Nikon Ophthalmic Instruments Miyagi, Japan), and tomography (Pentacam HR, Oculus Optikgerate GmbH, Wetzlar, Germany). This study included three groups of patients: those with normal eyes who served as controls (n = 33), patients with untreated keratoconus (n = 26), and patients with keratoconus treated at least 6 months prior to diagnosis with CXL and simultaneous partial, topography-guided photoreactive keratectomy (n = 42) (). Normal eyes were defined as eyes with corneas that did not have any ectasia criteria, based on cornea tomographic measurements, and noted as normal on the keratoconus screening software available on the Pentacam HR. Keratoconic eyes were defined as eyes diagnosed with keratoconus based on the same criteria. The average postoperative time elapsed for the group of CXL ectatic eyes was 28.9 ± 12.4 months (range 6 months to 6 years).

Table 1 Age characteristics of patients

HF UBM measurements

Ultrasound imaging data were stored and processed via the Zeus V 1.0 (Build 11.780) software, licensed from Artemis Medical Technologies, Inc. The software assumes a constant speed of sound of 1640 m/s throughout the ocular media. From each eye, the scan sequence consisted of four meridional B-scans at 45° intervals, namely horizontal, 45°, vertical, and 135°. Each scan sweep took about 0.25 seconds and consisted of 256 scan lines (pulse echo vectors). During the acquisition of each scan, raw data were converted (in near real time) to a B-scan displayed on the computer screen. Each B-scan reveals information regarding centration, ranging, and eye movements that may have occurred during the scan sweep. The examiner either accepted or chose to repeat a particular meridional sweep before proceeding to the next. At least three full scans per meridian were stored. The digitized ultrasound data were then transformed via Zeus software, which includes autocorrelation of back surface curvatures to center and aligns the meridional scans.

During acquisition, the center of rotation of the system was adjusted until it was coaxial with the corneal center, as seen from the internal camera. Following fixation and centering, total acquisition time was in the order of a few minutes. Of the two options offered, we selected the corneal analysis to be “centered on the middle of the scan”, and we did not select “circle-filter to smooth the map”.

From the group of stored scans, the examiner selected one scan (the most representative, most complete and centered acquisition) from each meridian to be included in the thickness profile report. A linear polar/radial interpolation function was used to interpolate among the selected scans to produce an 8 mm diameter map.

Descriptive statistics, comparative statistics, and linear regression were performed via Microsoft Excel 2010 (Microsoft Corp, Redmond, WA) and Origin Lab version 8 (OriginLab Corp, Northampton, MA). Statistical analysis was performed via the Origin Lab and IBM SPSS statistics tool (v.19.0).

Results

Precision dependence on meridional scan selection

One of the first issues to be investigated was the influence of meridional scan selection and number of acquisition scans on the precision of produced thickness measurements. As stated above, the same examiner selected the most representative scan to be included in the report. Currently the software does not offer the option of averaging of a number of same-meridian scans. We wanted to investigate how the selection among apparently equally well-suited choices of scans (for the same meridian) affects the precision of the reported values.

For this reason one male participant (normal eye) was subjected to a more extensive study, and a total of 20 meridional scans (ie, five scans per meridian) were stored. All possible different combinations of scan selections were considered to produce corneal pachymetry reports as shown in . Results from all possible combinations (n = 540) were subsequently tabulated and statistically analyzed ().

Figure 2 Corneal report produced by the Zeus software showing total corneal, epithelial, and stromal thickness pachymetry maps over 8 mm diameter.

Notes: The subject’s eye is normal. We observe the overall thicker epithelium over the pupil center.
Figure 2 Corneal report produced by the Zeus software showing total corneal, epithelial, and stromal thickness pachymetry maps over 8 mm diameter.

Table 2 Mean central corneal and epithelium thickness, as measured by HF UBM for the same patient and the same eye, as a result of different meridian selections acquired during a single examination session

demonstrates the correlation between central corneal thickness versus central epithelial thickness for the same point corresponding to the same data whose descriptive statistics appear in . shows the histogram graphs for central corneal thickness and central epithelial thickness for the same eye, as produced by the 540 different meridional combinations.

Figure 3 Central epithelial thickness versus central corneal thickness of the same data points as produced by 540 different combinations of select meridional scans.

Note: All data correspond to the same eye.
Figure 3 Central epithelial thickness versus central corneal thickness of the same data points as produced by 540 different combinations of select meridional scans.

Figure 4 Central corneal (A) and central epithelial (B) thickness histogram plots of the same data points as produced by 540 different combinations of select meridional scans.

Note: All data correspond to the same eye.
Figure 4 Central corneal (A) and central epithelial (B) thickness histogram plots of the same data points as produced by 540 different combinations of select meridional scans.

Data validation and collection

Our study of epithelial thickness maps was based on the concept of data validation. Of the 33 control group patients, 11 were female and 22 were male. Of the 26 untreated keratoconic patients, 7 were female and 19 were male. Of the 42 CXL-treated patients, 13 were female and 29 were male. For some patients, eyes were categorized into different groups (if, for example only one eye was treated with CXL). The bias towards the male population is consistent with our clinical experience in treating keratoconic patients.Citation29Citation31

The mean epithelial thickness over an area of 8 mm diameter and central epithelial thickness (where the center is defined as the pupil center) was recorded for each eye from the output of the corneal report, such as shown in . In order to estimate variation between the center and the periphery, the average of four points was computed for each eye and reported as Epithelium @ Periphery (). and summarize these results and show descriptive statistics.

Figure 5 The four transitional zone points used to calculate peripheral epithelium thickness on each eye.

Note: Epithelium thickness from these points was averaged, and is referred to as peripheral epithelium thickness.
Figure 5 The four transitional zone points used to calculate peripheral epithelium thickness on each eye.

Table 3 Descriptive statistics for mean (over an area of 8 mm diameter), center, and peripheral epithelium thickness measurements per each study group

Table 4 Descriptive statistics performed on the epithelial thickness measurements among the three groups

Discussion

Epithelium imaging – precision and accuracy

It is evident from our results in that, even when acquisition is carefully carried out and all indicators for a successful mapping exist, there is a ± 3–4 μm uncertainty on both corneal and epithelial thickness measurements. The latter is true not only on the nasal, temporal, superior, and inferior points which we examined, but also on the pupil center. While this figure might only be ±0.7% of the corresponding corneal thickness, it corresponds to a 5%–8% uncertainty for the corresponding epithelial thickness.

Our results indicate that there was a statistically significant correlation between central corneal thickness and central epithelium thickness (R = 0.554), as measured for the same patient for the same corneal point with different meridian B-scans selection (). However, this correlation was not very strong, suggesting that the epithelium does not necessarily follow equally proportional thickness increments as central corneal thickness. More than 30% of the total sample size was responsible for this correlation.

Regarding accuracy, the comparison of the same eye pachymetry measurements subjected to both AS OCT and Pentacam HR system revealed that there was at least a 10 μm difference between HF UBM and OCT, and a few microns difference between HF UBM and the Pentacam HR. The corneal thickness accuracy, as expressed by the paired correlations between HF UBM, AS OCT, and Pentacam HR, has been studied extensively by our group and has been submitted for publication.Citation31

Considering that during the specific acquisition particular care was taken to produce well-centered scans, and the patient was well-educated on the system (ie, the specific case almost fits the description of a best case), we will estimate in all of our following results that the HF UBM epithelium thickness readings have a ± 10% or ± 5 μm uncertainty. For the rest of our study, we will accept these figures as representative of all reported epithelium thickness measurements.

Epithelium imaging – control group

We observed from the results in that average mean corneal epithelial thickness for the control population was 50.9 ± 3.7 μm (range 43–58 μm). This is in agreement with previously reported values in pertinent publications, which range from 48 ± 5 μmCitation34 to 59.9 ± 5.9 μm.Citation3,Citation32Citation35

In this group of control patients, the epithelium was slightly thicker at the pupil center by an average of 1.3 μm (52.3 ± 4.9 μm, range 42–60 μm), when compared to the mean thickness. Mean thickness was computed by a small contribution of the area over the pupil center (ie, the thicker part of the epithelium) and by a larger contribution of the peripheral area (where the epithelium is thinner). Therefore, it is no surprise that while the epithelial center was, on average, thicker by only 1.3 μm compared to the mean, on several occasions it was thicker compared to the nasal, temporal, inferior, or superior points by up to 10 μm. This conclusion is also supported by the fact that the periphery epithelium thickness value of 49.7 μm was closer to the mean (50.9 μm) than to the pupil center (52.3 μm).

We note that the standard deviation of the measurements (±3–4 μm) is comparable to the accuracy and precision of the instrument, as established by our investigation, and thus epithelial thickness variations of ±4–4 μm, as it is the case, might be observed differently even on the same eye. An example of a control patient who demonstrated a thicker epithelium at the pupil center is shown in . In one instance the central epithelium was elevated by 9 μm (51 μm–42 μm), while in a subsequent examination of the same eye, the difference between the same points was recorded as only 6 μm (54 μm–48 μm).

Figure 6 Epithelial thickness maps of the same (control) patient, same eye, as produced by two consequtive acquisitions. Both maps demonstrate a thicker epithelium over the pupil center. Different acquisitions (A and B) of the same eye may produce epithelium maps which may vary by an estimated ± 4 μm.

Figure 6 Epithelial thickness maps of the same (control) patient, same eye, as produced by two consequtive acquisitions. Both maps demonstrate a thicker epithelium over the pupil center. Different acquisitions (A and B) of the same eye may produce epithelium maps which may vary by an estimated ± 4 μm.

Epithelium imaging – untreated KCN group

The group of untreated KCN patients showed an overall thickening, pronounced at the pupil center. Results over the whole KCN group indicated that on average, the epithelium at the pupil center was 3.1 μm thicker than the mean epithelial thickness (over the entire cornea) (). Simultaneously, the epithelium at the periphery was on average −2.1 μm thinner compared to the epithelium mean, and −5.2 μm thinner compared to the central epithelium.

Figure 7 Epithelium thickness across the three study groups, at the periphery, mean, and pupil center.

Abbreviation: KCN, keratoconus.
Figure 7 Epithelium thickness across the three study groups, at the periphery, mean, and pupil center.

The statistical significance of overall epithelial thickness was supported by descriptive statistics (run by both Origin Lab and SPSS, shown in .). The three-paired ANoVA test indicated P = 0.084 for the mean epithelium values, 0.28 for the epithelium at the periphery, and 0.004 for the epithelium thickness at the pupil center. Furthermore, the two-tailed t-test indicated statistical significance (P = 0.002) for the central epithelium between the control and the KCN group, as well as between the CXL-treated KCN and the untreated KCN group (P = 0.0008). This was in contrast to the null statistical insignificance (P = 0.9) between the control and the CXL-treated groups, further supporting our theory.

We encountered many examples such as shown in . It is noteworthy that there is a small area (located temporally–inferiorly), where the epithelium is much thinner. Although not reported on the map, the color scale helps us to estimate that the epithelium over an ectatic area in another example (,) was approximately 35 μm, while at the center was almost 60 μm. This is the area of the corneal protrusion, as we can observe from the thinner cornea. In a striking resemblance, the epithelium demonstrated similar characteristics in a pellucid marginal degeneration patient, ie, an area of thin epithelium over the corneal protrusion ().

Figure 8 Corneal and epithelial thickness maps of a KCN patient.

Note: A significantly thicker epithelium over the pupil center is observed.
Abbreviation: KCN, keratoconus.
Figure 8 Corneal and epithelial thickness maps of a KCN patient.

Figure 9 Example of localized thin epithelium (lower segment), over an area of a significantly thinner stroma.

Figure 9 Example of localized thin epithelium (lower segment), over an area of a significantly thinner stroma.

Figure 10 Examples of localized thinner epithelium over protruding corneas. (A and B) KCN patients, (C) a PMD patient.

Abbreviations: KCN, keratoconus; PMD, pellucid marginal degeneration.
Figure 10 Examples of localized thinner epithelium over protruding corneas. (A and B) KCN patients, (C) a PMD patient.

We believe that the overall thicker epithelium in KCN patients is a result of a reactive process; the epithelium may thicken in less “rigid” corneas due to being more susceptible to mechanical variations produced by one or a combination of the factors below:

  1. intraocular pressure variations

  2. eye rubbing

  3. heart driven blood pulse, translating in a pulsation on the cornea (personal communication: John Marshall, PhD)

  4. blinking mechanism.

To further investigate this hypothesis, and considering that keratoconic untreated corneas may become more biomechanically stable with advancing age, we separately examined the five younger (age 20.9 ± 3.5 years) and the five elder (age 33.1 ± 3.8 years) patients from the untreated KCN group.

The results indicate that while the periphery epithelium was identical, compared to the elder group, the average mean epithelium thickness was larger in the younger group by a small margin (+0.6 μm) and the central epithelium was statistically thicker (+2.5 μm) (). In other words, the elder group had on average a smoother and slightly thinner epithelium than the younger group.

Table 5 Mean, center, and peripheral epithelium thickness, as examined separately for the five younger and the five elder patients from the KCN group

The CXL treated KCN patients had epithelium thickness distributions that were similar to the control group, rather than the keratoconic group. While in some instances the epithelium was quite normal (), most often the epithelium map presented local variations (). This particular finding further supports our novel theory of “reactive” epithelial hyperplasia in biomechanically unstable corneas. It appears that in biomechanically stabilized irregular corneas the epithelium becomes “nonreactive” and conforms to more normal thickness, despite the highly irregular underlying stroma contour.

Figure 11 Examples of very smooth epithelium of a patient subjected to CXL.

Abbreviation: CXL, crosslinking.
Figure 11 Examples of very smooth epithelium of a patient subjected to CXL.

Figure 12 Examples of epithelium (patients subjected to CXL and partial topography guided PRK) demonstrating localized variations.

Abbreviations: CXL, crosslinking; PRK, photorefractive keratectomy.
Figure 12 Examples of epithelium (patients subjected to CXL and partial topography guided PRK) demonstrating localized variations.

Conclusions

In our study, an overall thicker epithelium with large variations was observed in untreated KCN patients. This is in contrast to previously published work and current understanding. On many occasions the epithelium was significantly thinner over areas of corneal ectasia, in agreement with previous observations.

However, patients subjected to CXL treatment did not show a statistically significant trend towards a thicker “reactive” epithelium, supporting our hypothesis of epithelial hyperplasia in biomechanically unstable corneas.

We feel this may be a significant new finding. In an era when most AS OCT devices will soon be able to offer detailed and accurate cornea epithelial maps, this theory may provide a very sensitive, pre-ectasia clinical parameter in corneas that have become biomechanically unstable. This “reactive” epithelial hypertrophy may precede any of the standard topographic or tomographic cornea findings. For example, “suspect” corneas due to thinner structure, high astigmatism, or even sanguine relation to a known KCN patient may be differentiated in regard to their ectasia potential based on the average cornea epithelial thickness.

The “normalization” of average epithelial thickness in long term follow-up of CXL treated ectasia cases further supports this theory. Further studies are needed to validate these results.

Disclosure

Dr Aslanides is a consultant of Artemis II.

References

  • SimonGRenQKervickGNParelJMOptics of the corneal epitheliumRefract Corneal Surg19939142508481372
  • ReinsteinDZArcherTJGobbeMSilvermanRHColemanDJEpithelial thickness in the normal cornea: three-dimensional display with very high frequency ultrasoundJ Refract Surg200824657158118581782
  • ReinsteinDZArcherTGobbeMStability of LASIK in corneas with forme fruste or suspect keratoconus, where keratoconus was excluded by epithelial thickness mappingPresented at: American Academy of Ophthalmology Annual MeetingNovember 9, 2008Atlanta, GA
  • WangJFonnDSimpsonTLSorbaraLKortRJonesLTopographical thickness of the epithelium and total cornea after overnight wear of reverse-geometry rigid contact lenses for myopia reductionInvest Ophthalmol Vis Sci2003444742474614578394
  • LadagePMYamamotoKRenDHEffects of rigid and soft contact lens daily wear on corneal epithelium, tear lactate dehydrogenase, and bacterial binding to exfoliated epithelial cellsOphthalmology20011081279128811425688
  • HoldenBAStephensonAStrettonSSuperior epithelial arcuate lesions with soft contact lens wearOptom Vis Sci200178191211233339
  • LarssonLIBourneWMPachJMStructure and function of the corneal endothelium in diabetes mellitus type I and type IIArch Ophthalmol19961149148540858
  • ShenoyRKhandekarRBialasiewiczAAl MuniriACorneal endothelium in patients with diabetes mellitus: a historical cohort studyEur J Ophthalmol200919336937519396780
  • HaqueSSimpsonTJonesLCorneal and epithelial thickness in keratoconus: a comparison of ultrasonic pachymetry, Orbscan II, and optical coherence tomographyJ Refract Surg200622548649316722488
  • LiYTanOHuangDNormal and keratoconic corneal epithelial thickness mapping using Fourier-domain optical coherence tomographyProc SPIE2011796579650810.1117/12.878567.
  • FengYSimpsonTLComparison of human central cornea and limbus in vivo using optical coherence tomographyOptom Vis Sci20058241641915894917
  • EckardAStaveJGuthoffRFIn vivo investigations of the corneal epithelium with the confocal Rostock Laser Scanning Microscope (RLSM)Cornea200625212713116371768
  • Møller-PedersenTVogelMLiHFPetrollWMCavanaghHDJesterJVQuantification of stromal thinning, epithelial thickness, and corneal haze after photorefractive keratectomy using in vivo confocal microscopyOphthalmology19971043603689082257
  • PérezJGMéijomeJMJalbertISweeneyDFEricksonPCorneal epithelial thinning profile induced by long-term wear of hydrogel lensesCornea20032230430712792471
  • LiHFPetrollWMMøller-PedersenTMaurerJKCavanaghHDJesterJVEpithelial and corneal thickness measurements by in vivo confocal microscopy through focusing (CMTF)Curr Eye Res1997162142219088737
  • ChenTCCenseBPierceMCSpectral domain optical coherence tomography: ultra-high speed, ultra-high resolution ophthalmic imagingArch Ophthalmol2005123121715172016344444
  • GrewalDSBrarGSJainRGrewalSPSComparison of Scheimpflug imaging and spectral domain anterior segment optical coherence tomography for detection of narrow anterior chamber anglesEye20112560361121336254
  • WangJFonnDSimpsonTLJonesLThe measurement of corneal epithelial thickness in response to hypoxia using optical coherence tomographyAm J Ophthalmol2002133331531911860966
  • SinSSimpsonTLThe repeatability of corneal and corneal epithelial thickness measurements using optical coherence tomographyOptom Vis Sci200683636036516772894
  • PrakasamRKSchwiedeMHützWWGuthoffRFStachsOCorneal responses to eye rubbing with spectral domain optical coherence tomographyCurr Eye Res2012371253222029687
  • ReinsteinDZSilvermanRHColemanDJHigh-frequency ultrasound measurement of the thickness of the corneal epitheliumRefract Corneal Surg1993953853878241045
  • ReinsteinDZSilvermanRHTrokelSLColemanDJCorneal pachymetric topographyOphthalmology199410134324388127563
  • DenoyerAOssantFArbeilleBVery-high-frequency ultrasound corneal imaging as a new tool for early diagnosis of ocular surface toxicity in rabbits treated with a preserved glaucoma drugOphthalmic Res20084029830818506111
  • MundtGHHughesWFUltrasonics in ocular diagnosisAm J Ophthalmol19564248849813362453
  • BaumGGreenwoodIThe application of ultrasonic locating techniques to ophthalmology. II. Ultrasonic slit-lamp in the ultrasonic visualization of soft tissuesArch Ophthalmol195860263279
  • PavlinCJHarasiewiczKSherarMDFosterFSClinical use of ultrasound biomicroscopyOphthalmology1991982872952023747
  • PavlinCJSherarMDFosterFSSubsurface ultrasound microscopic imaging of the intact eyeOphthalmology1990972442502326015
  • ReinsteinDZSilvermanRHRaevskyTArc-scanning very high-frequency digital ultrasound for 3D pachymetric mapping of the corneal epithelium and stroma in laser in situ keratomileusisJ Refract Surg20001641443010939721
  • KanellopoulosAJComparison of sequential vs same-day simultaneous collagen cross-linking and topography-guided PRK for treatment of keratoconusJ Refract Surg2009259S812S81819772257
  • KanellopoulosAJLong term results of a prospective randomized bilateral eye comparison trial of higher fluence, shorter duration ultraviolet A radiation, and riboflavin collagen cross linking for progressive keratoconusClin Ophthalmol201269710122275813
  • KanellopoulosAJAslanidesIMAsimellisGClinical correlation between corneal pachymetry measurements by Scheimflug-image cornea tomography (Oculyzer II-Pentacam HR), anterior segment optical coherence tomography and high-frequency scanning ultrasound biomicroscopy (submitted Jan 2012)
  • WangJThomasJCoxIRollinsANoncontact measurements of central corneal epithelial and flap thickness after laser in situ keratomileusisInvest Ophthalmol Vis Sci2004451812181615161844
  • WirbelauerCPhamDTMonitoring corneal structures with slit-lamp-adapted optical coherence tomography in laser in situ keratomileusisJ Cataract Refract Surg2004301851186015342046
  • HaqueSFonnDSimpsonTJonesLCorneal and epithelial thickness changes after 4 weeks of overnight corneal refractive therapy lens wear, measured with optical coherence tomographyEye Contact Lens20043018919315499246
  • Møller-PedersenTLiHFPetrollWMCavanaghHDJesterJVConfocal microscopic characterization of wound repair after photorefractive keratectomyInvest Ophthalmol Vis Sci1998394875019501858