62
Views
1
CrossRef citations to date
0
Altmetric
Case Series

Optimizing visualization in enhanced depth imaging OCT in healthy subjects and patients with retinal pigment epithelial detachment

, , , , , & show all
Pages 1915-1920 | Published online: 21 Nov 2012

Abstract

Background

This study’s objective was to optimize the visualization of three different spectral-domain optical coherence tomography (SD-OCT) display modalities and evaluate enhanced depth imaging (EDI) by comparing the maximum depth of assessment in conventional versus inverted cross-sectional OCT images in healthy subjects and in patients with retinal pigment epithelial detachment (PED).

Methods

Cross-sectional SD-OCT conventional and inverted images were obtained with the HRA2 (Heidelberg Retina Angiograph II, Heidelberg Engineering, Heidelberg, Germany). Horizontal as well as vertical sections in three different display modes were blinded for evaluation by three independent, experienced graders for maximal imaging depth of the deep ocular fundus layers.

Results

The mean imaging depth as measured from the inner segment/outer segment (IS/OS) to the outer choroid of all 14 healthy subjects was 197 ± 44 μm vs 263 ± 56 μm for conventional vs EDI scans: in black/white mode, it was significantly lower (P < 0.001) than in white/black mode (249 ± 42 μm vs 337 ± 71 μm) and color/heat mode (254 ± 48 μm vs 354 ± 73 μm). The mean imaging depth of all 14 study eyes with PED was 240 ± 78 μm vs 345 ± 100 μm for conventional vs EDI scans in black/white mode, and was significantly lower (P < 0.001) than in white/black mode (393 ± 104 μm vs 464 ± 126 μm) and in color/heat mode (373 ± 106 μm vs 453 ± 114 μm). In each display modality of healthy subjects and of patients with PED, EDI scans showed a significantly higher imaging depth than the corresponding conventional scans.

Conclusion

White/black and color/heat modes allow increased imaging depth, compared to black/white mode using both conventional or EDI OCT scans in healthy subjects or patients with PED. EDI obtained with HRA2 significantly improves the imaging depth, compared to conventional OCT scans.

Introduction

Optical coherence tomography (OCT) is becoming more relevant for ophthalmic diagnostics and is increasingly used for therapy.Citation1Citation10 Imaging speed and quality have greatly improved over the last few years.Citation11Citation15

Over the past two decades, the focus has been predominantly on diseases with retinal or vitreoretinal pathology, such as diabetic retinopathy, macular edema, age-related macular degeneration, vitreomacular traction syndrome, and macular holes, imaging of the choroid has gained increasing attention. Ocular diseases of the posterior pole, such as myopia, pseudoxanthoma elasticum, retinal pigment epithelial detachment, central serous retinopathy, Vogt–Koyanagi–Harada disease, and retinal dystrophies, were shown to have alterations of the choroid that could be visualized and evaluated using the noninvasive, enhanced depth imaging (EDI) OCT technique.Citation16Citation21

Increased penetration depth for a better evaluation of the choroid can be achieved either by using laser sources with a higher wavelength compared to the commonly used Spectralis® OCT with 870 nm (Heidelberg Engineering, Heidelberg, Germany) or, alternatively, to manually inverting the image obtained in vivo or using the recently available automatic EDI-module in the Spectralis® OCT HRA2.Citation19,Citation22Citation25 Measurements of the choroid obtained with enhanced depth-imaging protocols using a variety of OCT devices have a high reproducibility among healthy patients.Citation26,Citation27

However, subretinal pathologies with significant signal attenuation due to thickening or high absorption of the emitted light, as for example in pigment epithelial detachment (PED), are still a challenge for OCT imaging.

The purpose of our study was to optimize the visualization of three different SD-OCT display modalities of conventional versus inverted (EDI) cross-sectional OCT images by comparing the maximum depth of assessment in healthy subjects with that in patients with PED.

Methods

Patients

A consecutive series of 14 eyes of healthy subjects, mean age 31 ± 5 years, and 14 eyes diagnosed with PED due to AMD, mean age 74 ± 10 years, were included in this study. Exclusion criteria for healthy subjects were any retinal pathologies or media opacities and for patients with PED previous focal laser treatment of the macula, PED due to choroidal or suprachoroidal tumors, or significant media opacities. After informed consent, a full retinal examination was performed by a retinal specialist, followed by OCT imaging. All research was conducted in accordance with institutional guidelines and board approval, and conformed to the tenets of the World Medical Association Declaration of Helsinki.

SD-OCT images

All patients were imaged with conventional and enhanced depth-image cross-sectional scans by a Heidelberg Spectralis® OCT (Heidelberg Engineering) with an excitation wavelength of 870 nm and a scanning speed of 40,000 A-scans per second. The axial resolution was ~6 μm, compared to a lateral resolution of 14 μm. EDI was performed by positioning the SD-OCT closer to the eye in order to generate an inverted image on the top of the computer display. A possible physical explanation for EDI may be a decreased delay in light wavelengths returning from the depicted subretinal structures that are further away from the zero delay line; this is detected by the interferometer and compared to the light wavelengths of the reference arm, the final result of which is a better signal intensity of the imaged deeper layers of the posterior fundus.Citation25 A horizontal and vertical OCT image was obtained through the fovea, each comprised of 15 averaged scans, using the automatic-averaging and eye-tracking features of the Heidelberg Spectralis® OCT. Acquisition of the cross-sectional scans was obtained together with an infrared, real-time fundus image, which allowed point-to-point correlation between the fundus images and the OCT cross-sectional scans. The same modality with a horizontal and vertical cross-sectional scan was used to obtain conventional scans for comparison.

Image analysis

Quantitative analysis of the obtained horizontal and vertical SD-OCT cross-sectional scans of each study eye was performed using Image J software (NIH, available at http://rsb.info.nih.gov). Before evaluating the images, the inverted scans were processed to obtain upright images allowing blinded evaluation. The largest possible distance from the inner-segment/outer-segment band (IS/OS) to the farthest discernible structure was measured in each conventional and inverted image in the black/white mode, white/black mode, and color/heat mode display modalities () of each horizontal and vertical cross-sectional scan obtained. All measurements of each eye were successively obtained and graded by three blinded, independent, experienced graders on 22-inch TFT displays (Flatron; maximal resolution 1680 × 1050 pixels, brightness 250 cd/m2; LG Corp, Seoul, South Korea) using the same measurement scale. Mean values from each graders’ measurements of the horizontal and vertical scans were calculated for each of the three display modalities.

Figure 1 Vertical cross-sectional scan of the left eye through the fovea of a 66-year-old female patient displaying a central fibrovascular pigment epithelial detachment. The display modalities are black/white mode in (A) (conventional) and (B) (enhanced depth imaging); white/black mode in (C) (conventional) and (D) (enhanced depth imaging); and color/heat mode in (E) (conventional) and (F) (enhanced depth imaging). Inverted scans (B, D, and F) show an increased imaging depth of the choroid in all three different display modalities, compared to the conventional scans.

Figure 1 Vertical cross-sectional scan of the left eye through the fovea of a 66-year-old female patient displaying a central fibrovascular pigment epithelial detachment. The display modalities are black/white mode in (A) (conventional) and (B) (enhanced depth imaging); white/black mode in (C) (conventional) and (D) (enhanced depth imaging); and color/heat mode in (E) (conventional) and (F) (enhanced depth imaging). Inverted scans (B, D, and F) show an increased imaging depth of the choroid in all three different display modalities, compared to the conventional scans.

Statistical analysis

Data were collected and analyzed using SPSS Version 17.0 (IBM Corporation, Armonk, NY). A P-value of <0.05 was considered statistically significant. The interobserver reproducibility values were evaluated based on the intraclass correlation coefficient according to Shrout and Fleiss.Citation28 Nonparametric analyses (Wilcoxon signed-ranks test) were applied and adjusted for multiple testing by applying the Bonferroni adjustment.

Results

The overall interobserver reliability for each measurement and for each display modality between all three independent graders was very high and statistically significant, with a Kronbach’s alpha of 0.888 (0.822–0.950; P < 0.001).

The mean imaging depth of all 14 healthy subjects was 197 ± 44 μm vs 263 ± 56 μm for conventional vs EDI scans in black/white mode, and was significantly lower (P < 0.001) than in the white/black mode, with 249 ± 42 μm vs 337 ± 71 μm, and than in the color/heat mode, with 254 ± 48 μm vs 354 ± 73 μm ( and ).

Figure 2 Vertical cross-sectional scan of the right eye through the fovea of a 27-year-old male healthy subject. The display modalities are black/white mode in (A) (conventional) and (B) (enhanced depth imaging); white/black mode in (C) (conventional) and (D) (enhanced depth imaging); and color/heat mode in (E) (conventional) and (F) (enhanced depth imaging). Inverted scans (B, D, and F) show an increased imaging depth of the choroid in all three different display modalities, compared to the conventional scans.

Figure 2 Vertical cross-sectional scan of the right eye through the fovea of a 27-year-old male healthy subject. The display modalities are black/white mode in (A) (conventional) and (B) (enhanced depth imaging); white/black mode in (C) (conventional) and (D) (enhanced depth imaging); and color/heat mode in (E) (conventional) and (F) (enhanced depth imaging). Inverted scans (B, D, and F) show an increased imaging depth of the choroid in all three different display modalities, compared to the conventional scans.

Figure 3 Mean imaging depth of conventional compared to enhanced depth imaging (EDI) optical coherence tomography scans in three different display modalities of all 14 healthy subjects with pigment epithelial detachment obtained by three independent graders.

Figure 3 Mean imaging depth of conventional compared to enhanced depth imaging (EDI) optical coherence tomography scans in three different display modalities of all 14 healthy subjects with pigment epithelial detachment obtained by three independent graders.

Fourteen eyes with PED met the inclusion criteria for this group. Of those, five had serous PEDs and eight had fibrovascular PEDs, as determined by OCT findings. Imaging artifacts and disruption of OCT signal increased with the height and amount of nonserous content of the PED. The mean imaging depth of all 14 eyes with PED was 240 ± 78 μm vs 345 ± 100 μm for conventional vs EDI scans in black/white mode, which was significantly lower (P < 0.001) than in white/black mode (393 ± 104 μm vs 464 ± 126 μm) and in color/heat mode (373 ± 106 μm vs 453 ± 114 μm). In each display modality of healthy subjects and patients with PED, EDI scans showed a significantly higher imaging depth than did the corresponding conventional scans.

A statistically significant different maximum depth of imaging was found between conventional and EDI scans and their particular display modes by applying the Friedman test for nonparametrical variables, with χ2 = 158.724, P = 0.000 for healthy subjects and χ2 = 138.190, P < 0.001 for patients with PED. Post hoc analysis with the Bonferroni adjustment (P < 0.017) of the Wilcoxon signed-ranks test revealed a significant difference between conventional and EDI scans for healthy subjects and patients with PED, respectively, in each evaluated mode: black/white mode (Z = −5.545, P < 0.001 and Z = −5.208, P < 0.001); white/black mode (Z = −5.458, P < 0.001 and Z = −4.620, P < 0.001); and color/heat mode (Z = −5.570, P < 0.001 and Z = −4.739, P < 0.001) ( and ).

Figure 4 Vertical cross-sectional scan of the left eye through the fovea of a 79-year-old female patient displaying a central mainly serous pigment epithelial detachment. The display modalities are black/white mode in (A) (conventional) and (B) (enhanced depth imaging); white/black mode in (C) (conventional) and (D) (enhanced depth imaging); and color/heat mode in (E) (conventional) and (F) (enhanced depth imaging). Inverted scans (B, D, and F) show an increased imaging depth through the subretinal serous fluid compared to the conventional scans, shown in (C’ and D’).

Figure 4 Vertical cross-sectional scan of the left eye through the fovea of a 79-year-old female patient displaying a central mainly serous pigment epithelial detachment. The display modalities are black/white mode in (A) (conventional) and (B) (enhanced depth imaging); white/black mode in (C) (conventional) and (D) (enhanced depth imaging); and color/heat mode in (E) (conventional) and (F) (enhanced depth imaging). Inverted scans (B, D, and F) show an increased imaging depth through the subretinal serous fluid compared to the conventional scans, shown in (C’ and D’).

Figure 5 Mean imaging depth of conventional compared to enhanced depth imaging (EDI) optical coherence tomography scans in three different display modalities of all 14 study eyes with pigment epithelial detachment obtained by three independent graders.

Figure 5 Mean imaging depth of conventional compared to enhanced depth imaging (EDI) optical coherence tomography scans in three different display modalities of all 14 study eyes with pigment epithelial detachment obtained by three independent graders.

A statistically significant difference was also observed between black/white and white/black mode, as well as between black/white and color/heat mode. This was true for both types of conventional and EDI scans for healthy subjects (Z = −5.620, P < 0.001 and Z = −5.620, P < 0.001) as well as for patients with PED (Z = −5.495, P < 0.001 and Z = −5.645, P < 0.001). It was also true for EDI scans of healthy subjects (Z = −5.449, P < 0.001 and Z = −5.558, P < 0.001) and 1 for patients with PED (Z = −5.495, P < 0.001 and Z = −5.645, P < 0.001).

Discussion

In our study, the depth of obtained OCT scans was, as expected, larger in EDI mode than in conventional imaging mode. This was true for all display modalities and was observable in healthy individuals and in patients with PED.

OCT EDI was introduced in 2008 and has become increasingly popular for imaging the choroid.Citation25 Enhanced depth images of patients with PED allow us to examine the internal structure of PED presenting with serous or fibrovascular content.Citation19

OCT signal strength decreases as light penetrates tissue and is back-reflected or scattered. In addition, using spectral-domain OCT, signal strength inherently decreases on the OCT image as the point of interest moves away from the zero delay line (typically placed at the top of the OCT image).Citation25,Citation29,Citation30

PED, with its increased retinal and/or choroidal thickness, presents a challenge for imaging deeper structures based upon the physical principles mentioned above. Therefore, in addition to assessing visualization in healthy subjects, we have chosen eyes with PED to evaluate and optimize the maximum imaging depth of different modes in conventional and EDI imaging.

The observed increase of imaging depth for inverted vs conventional OCT scans, not only in healthy subjects but also in patients with PED, supports the concept of enhancing signal sensitivity by inverting the image and intentionally focusing upon the deeper layers.

Interestingly, we found an increased imaging depth in white/black mode or color/heat mode compared to black/white mode, both in healthy subjects and in patients with PED. This might be explained by varying distinct software-related contrast discrimination between the evaluated display modalities (black/white, white/black and color/heat) of the same cross-sectional scan. The observed display modality differences showed similar tendencies in conventional and inverted OCT scans.

Our results from the second group of patients are restricted to the specific pathology of PED. The heterogenous internal structures, including blocking phenomena in overlying retinal layers, can be subdivided into serous and fibrovascular PED, which results in varying imaging depths with differing decreases in signal intensity. This heterogeneity can be seen in the relatively large confidence intervals of mean measurements within a particular modality (see ). Nonetheless, there was a convincing consistency among the three independent graders (high interobserver reliability of 0.914). In our study, six out of 14 eyes had serous PED and eight had fibrovascular PED, with significant decrease in signal intensity observed predominantly in the fibrovascular subtype that did not allow for imaging the full thickness of the choroid. Therefore, alternative OCT devices with a better penetration through the retinal pigment epithelium using laser sources with higher wavelengths may be more suitable for choroidal imaging in the setting of PED.Citation31Citation33

To achieve statistical reliability, the maximum imaging depth for each of the 28 eyes included in the study was measured by three independent graders for each of the three display modalities (scanning horizontally and vertically, cross-sectionally, and by type – conventional or EDI), resulting in 36 measurements per eye and 1008 measurements for all 28 eyes included.

In conclusion, the two display modalities – white/black mode and color/heat mode – allow increased imaging depth, compared to the black/white mode using the HRA2 for both conventional and EDI OCT scans in patients with PED. Nonetheless, deeper-structure EDI still presents a challenge. EDI obtained with HRA2 significantly improves the imaging depth, compared to conventional OCT scans.

Disclosure

The authors report no conflicts of interest in this work, including any commercial interest in any of the materials and methods used in this study.

References

  • HuangDSwansonEALinCPOptical coherence tomographyScience1991254117811811957169
  • LandaGSuEGarciaPMSeipleWHRosenRBInner segment-outer segment junctional layer integrity and corresponding retinal sensitivity in dry and wet forms of age-related macular degenerationRetina20113136437021221051
  • NeubauerASPriglingerSGThielMJMayCAWelge-LussenUCSterile structural imaging of donor cornea by optical coherence tomographyCornea20022149049412072724
  • QiuKLZhangMZThe effect of image quality on retinal nerve fiber layer measurement in Stratus OCTAm J Ophthalmol2011151384 author reply 384–38521251500
  • ReznicekLKerntMHaritoglouCKampikAUlbigMNeubauerASIn vivo characterization of ischemic retina in diabetic retinopathyClin Ophthalmol20105313521311655
  • BinderSFalkner-RadlerCIHaugerCMatzHGlittenbergCFeasibility of intrasurgical spectral-domain optical coherence tomographyRetina2011311332133621273942
  • BrownDMKaiserPKMichelsMRanibizumab versus verteporfin for neovascular age-related macular degenerationN Engl J Med20063551432144417021319
  • ClarkMEMcGwinGJrNeelyDAssociation between retinal thickness measured by spectral-domain optical coherence tomography (OCT) and rod-mediated dark adaptation in non-exudative age-related maculopathyBr J Ophthalmol2011951427143221289019
  • FungAELalwaniGARosenfeldPJAn optical coherence tomography-guided, variable dosing regimen with intravitreal ranibizumab (Lucentis) for neovascular age-related macular degenerationAm J Ophthalmol200714356658317386270
  • WitkinAJVuongLNSrinivasanVJHigh-speed ultrahigh resolution optical coherence tomography before and after ranibizumab for age-related macular degenerationOphthalmology200911695696319410953
  • GorczynskaISrinivasanVJVuongLNProjection OCT fundus imaging for visualising outer retinal pathology in non-exudative age-related macular degenerationBr J Ophthalmol20099360360918662918
  • HoJWitkinAJLiuJDocumentation of intraretinal retinal pigment epithelium migration via high-speed ultrahigh-resolution optical coherence tomographyOphthalmology201111868769321093923
  • SerbecicNBeutelspacherSCAboul-EneinFCKircherKReitnerASchmidt-ErfurthUReproducibility of high-resolution optical coherence tomography measurements of the nerve fibre layer with the new Heidelberg Spectralis optical coherence tomographyBr J Ophthalmol20119580481021097787
  • SrinivasanVJAdlerDCChenYUltrahigh-speed optical coherence tomography for three-dimensional and en face imaging of the retina and optic nerve headInvest Ophthalmol Vis Sci2008495103511018658089
  • Wolf-SchnurrbuschUECeklicLBrinkmannCKMacular thickness measurements in healthy eyes using six different optical coherence tomography instrumentsInvest Ophthalmol Vis Sci2009503432343719234346
  • FongAHLiKKWongDChoroidal evaluation using enhanced depth imaging spectral-domain optical coherence tomography in Vogt-Koyanagi-Harada diseaseRetina20113150250921336069
  • FujiwaraTImamuraYMargolisRSlakterJSSpaideRFEnhanced depth imaging optical coherence tomography of the choroid in highly myopic eyesAm J Ophthalmol200914844545019541286
  • Ari YaylaliSErbilHHSalarSKarakurtYOptical coherence tomography findings in pseudoxanthoma elasticumEur J Ophthalmol20102039740119882544
  • SpaideRFEnhanced depth imaging optical coherence tomography of retinal pigment epithelial detachment in age-related macular degenerationAm J Ophthalmol200914764465219152869
  • ImamuraYFujiwaraTMargolisRSpaideRFEnhanced depth imaging optical coherence tomography of the choroid in central serous chorioretinopathyRetina2009291469147319898183
  • YeohJRahmanWChenFChoroidal imaging in inherited retinal disease using the technique of enhanced depth imaging optical coherence tomographyGraefes Arch Clin Exp Ophthalmol20102481719172820640437
  • PovazayBBizhevaKHermannBEnhanced visualization of choroidal vessels using ultrahigh resolution ophthalmic OCT at 1050 nmOpt Express2003111980198619466083
  • LeeECde BoerJFMujatMLimHYunSHIn vivo optical frequency domain imaging of human retina and choroidOpt Express2006144403441119516592
  • EsmaeelpourMPovazayBHermannBThree-dimensional 1060-nm OCT: choroidal thickness maps in normal subjects and improved posterior segment visualization in cataract patientsInvest Ophthalmol Vis Sci2010515260526620445110
  • SpaideRFKoizumiHPozzoniMCEnhanced depth imaging spectral-domain optical coherence tomographyAm J Ophthalmol200814649650018639219
  • RahmanWChenFKYeohJPatelPTufailADa CruzLRepeatability of manual subfoveal choroidal thickness measurements in healthy subjects using the technique of enhanced depth imaging optical coherence tomographyInvest Ophthalmol Vis Sci2011522267227121087970
  • BranchiniLRegatieriCVFlores-MorenoIBaumannBFujimotoJGDukerJSReproducibility of choroidal thickness measurements across three spectral domain optical coherence tomography systemsOphthalmology201211911912321943786
  • ShroutPEFleissJLIntraclass correlations: uses in assessing rater reliabilityPsychol Bull19798642042818839484
  • AnLWangRKUse of a scanner to modulate spatial interferograms for in vivo full-range Fourier-domain optical coherence tomographyOptics letters2007323423342518059954
  • WangRKFourier domain optical coherence tomography achieves full range complex imaging in vivo by introducing a carrier frequency during scanningPhys Med Biol2007525897590717881807
  • EigenwilligCMKleinTWieserWBiedermannBRHuberRWavelength swept amplified spontaneous emission source for high speed retinal optical coherence tomography at 1060 nmJ Biophotonics2011455255821780301
  • KleinTWieserWEigenwilligCMBiedermannBRHuberRMegahertz OCT for ultrawide-field retinal imaging with a 1050 nm Fourier domain mode-locked laserOpt Express2011193044306221369128
  • EsmaeelpourMPovazayBHermannBMapping choroidal and retinal thickness variation in type 2 diabetes using three-dimensional 1060-nm optical coherence tomographyInvest Ophthalmol Vis Sci2011525311531621508108