919
Views
35
CrossRef citations to date
0
Altmetric
Research Article

Analytical performance, agreement and user-friendliness of five C-reactive protein point-of-care tests

, , , , , , , & show all
Pages 627-634 | Received 21 Feb 2013, Accepted 31 Aug 2013, Published online: 14 Oct 2013
 

Abstract

Background. Point-of-care (POC) C-reactive protein (CRP) testing is increasingly used in primary care to assist general practitioners (GPs) in the diagnostic workup for various complaints. The present study compares analytical performance, agreement and user-friendliness of five of these POC CRP tests. Methods. The following five POC CRP tests were evaluated: Afinion and NycoCard Reader II (both Alere), Eurolyser Smart 700/340 (Eurolyser), QuikRead go and QuikRead 101 (both Orion Diagnostica). Results were compared with those of a standard immunoturbidimetric method performed on a routine analyzer (Olympus AU 2700, Beckman Coulter). Analytical performance and agreement with the laboratory standard for the five different POC tests were analyzed. Subsequently, user-friendliness of the POC tests was assessed. Results. Within-day CVs varied from 2.6% (QuikRead go) to 19.4% (Eurolyser Smart 700/340) for low CRP values (< 20 mg/L), and 1.1% (QuikRead go) to 17.5% (Eurolyser Smart 700/340) for high values (> 100 mg/L). Between-day CVs varied from 4.6% (Afinion) to 30.5% (Eurolyser Smart 700/340) for low values and 4.0% (QuikRead go) to 18.0% (Eurolyser Smart 700/340) for high values. With high CRP values (> 100 mg/L) agreement with the laboratory standard systematically decreased for all POC tests. Regarding user-friendliness Afinion and Eurolyser Smart 700/340 were judged easiest to operate. Conclusions. Analytical performance, agreement, and user-friendliness of the POC CRP tests varied considerably, yet overall four devices showed adequate analytical performance and agreement.

Acknowledgements

We thank the staff of the medical diagnostic center of Utrecht (Saltro Diagnostics) particular Zakia El Hachami and Atika Talhaoui and the GPs and GPAs who volunteered in performing analyses for this evaluation.

Declaration of interest: The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 200.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.