Pages 588-594 | Received 08 Jan 2014, Accepted 11 Jun 2015, Published online: 12 Aug 2015
 

Abstract

Introduction. Automated hematology analyzers dilute patient erythrocytes with an isoosmotic diluent before quantitating the erythrocyte mean cell volume (MCV). However, if patient plasma osmolality differs from the diluent, water will cross the erythrocytes membrane and establish a new equilibrium across the membrane. Since the new equilibrium is reached before the measurement of the MCV, the measured MCV may not reflect the true MCV in vivo. Aim. Calculation of the theoretical change in MCV at changed P-Sodium/P-Osmolality and to investigate if the automated blood cell counter Sysmex XE 2100 measures MCV correctly in hypo- and hyperosmolality and hypo-and hypernatremia. In addition, to examine whether the theoretically calculated change in MCV corresponds with the experimentally determined MCV change. Method. Theoretical calculation of the MCV inaccuracy at hypo- and hypernatremia, as well as at hypo- and hyperosmolality. Experimental studies with comparison of MCV measured at Sysmex XE 2100 to MCV found by using the manual measured packed cell volume method. Results and conclusion. Measurement of MCV in hypo- and hypernatremia patients using the automated blood cell counter Sysmex XE 2100 resulted in inaccurate MCV. The experimental results also revealed a strong correlation between P-Osmolality/P-Sodium and MCV inaccuracy (R2 = 0.70/0.85) similar to the theoretically calculated MCV inaccuracy. We suggest using mean cellular Hb (MCH) instead of MCV, mean corpuscular Hb concentration (MCHC) and B-Erythrocyte volume fraction (EVF). Alternatively, we suggest standardizing the measured MCV to a normal P-Sodium e.g. 140 mmol/L to estimate the in vivo MCV.

Acknowledgements

A special thanks to senior chief biomedical laboratory technologist Lisbeth Mikkelsen at Department of Medical Research, Holstebro Hospital and Aarhus University, Holstebro, Denmark for measuring the P-Osmolality samples.

Declaration of interest: The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 200.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.