A short biodegradable helical spiral ureteric stent provides better antireflux and drainage properties than a double-J stent

, , , &
Pages 129-133 | Received 02 Sep 2010, Accepted 20 Oct 2010, Published online: 12 Jan 2011
 

Abstract

Objective. The use of ureteric stents is a standard treatment for the relief of ureter blockages for benign or malignant reasons. The most common stent design in clinical use is a double-J stent with coiled ends to avoid stent displacement. However, there are a number of complications associated with stent use. A double-J stent design bypasses the ureterovesical junction, enables bladder pressure reflection to the renal pelvis and causes vesicoureteral reflux (VUR). This may result in scarring and renal failure. Material and methods. An animal model was used to investigate whether VUR can be avoided in stented ureters using a short biodegradable partial helical spiral stent design that leaves the ureterovesical junction intact. Materials and methods. Eight female pigs were used. Ureters on the left side were stented using a short helical spiral SR-PLGA stent (group A) and ureters on the right side using double-J stents (group B). Simulated voiding cystoureterography and standard intravenous urography examinations were performed on all eight animals at 4 weeks and on the remaining four animals at 8 weeks. Results. An SR-PLGA single coiled partial stent demonstrated superior drainage properties to a double-J stent at 4 weeks (p = 0.020). A marked but not statistically significant difference in favour of a SR-PLGA stent was also observed at 8 weeks (p = 0.102). A statistically significant difference was observed in VCUG findings in favour of group A at immediate postoperative control as well as in the 4 and 8 week follow-up studies (p = 0.011, p = 0.010, p = 0.046, respectively). Conclusion. A self-expandable, SR-PLGA partial ureteric stent presented with superior drainage and antireflux properties compared to a double-J stent. The reflux commonly related to double-J stent use can be minimized by using a partial ureteric stent design.

Declaration of interest: The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access
  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart
* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.