Low accuracy of manual white blood cell count in amniotic fluid

, , , &
Pages 364-366 | Published online: 08 May 2013
 

Abstract

Intra-amniotic infection (IAI) is a common cause of pre-term labour. Manual WBC count on amniotic fluid (AF) has been suggested as a diagnostic test for IAI using a threshold of 50 cells/mm3. However, no validation studies assessing the accuracy of this method have been performed. AF samples were selected for cell count analysis. WBCs were introduced to 47 AF samples. The results from two technologists’ counts were compared with the calculated expected value for WBCs in these samples. Results showed that a comparison between the technologists’ WBC count to the expected WBC count yielded R2 coefficients of 0.62 and 0.78, indicating moderate accuracy. Percentage agreement between the technologists was 67%, indicating low reproducibility. It was concluded that there was moderate correlation between the manual and the expected WBC in the spiked AF samples. Clinicians should be aware of the inaccuracy and imprecision associated with this test when evaluating a patient for IAI.

Declaration of interest: The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.