Design of a water coupling bolus with improved flow distribution for multi-element superficial hyperthermia applicators

, , , , &
Pages 554-565 | Received 13 Mar 2009, Accepted 17 Jun 2009, Published online: 24 Oct 2009

Abstract

A water bolus used in superficial hyperthermia couples the electromagnetic (EM) or acoustic energy into the target tissue and cools the tissue surface to minimise thermal hotspots and patient discomfort during treatment. Parametric analyses of the fluid pressure inside the bolus computed using 3D fluid dynamics simulations are used in this study to determine a bolus design with improved flow and surface temperature distributions for large area superficial heat applicators. The simulation results are used in the design and fabrication of a 19 × 32 cm prototype bolus with dual input-dual output (DIDO) flow channels. Sequential thermal images of the bolus surface temperature recorded for a step change in the circulating water temperature are used to assess steady state flow and surface temperature distributions across the bolus. Modelling and measurement data indicate substantial improvement in bolus flow and surface temperature distributions when changing from the previous single input–single output (SISO) to DIDO configuration. Temperature variation across the bolus at steady state was measured to be less than 0.8°C for the DIDO bolus compared to 1.5°C for the SISO water bolus. The new DIDO bolus configuration maintains a nearly uniform flow distribution and low variation in surface temperature over a large area typically treated in superficial hyperthermia.

Introduction

Moderate temperature hyperthermia in the range of 40–44°C has been demonstrated to be a promising adjuvant to radiation and chemotherapy treatments of superficial tissue disease Citation[1–4]. Moreover, preclinical data has demonstrated a significant increase in thermal enhancement ratio (response with heat/response without heat) when heat and radiation therapies are delivered simultaneously Citation[5–7]. This led to the development of new combination applicators capable of simultaneous therapy Citation[8–13]. One such applicator developed by this group for concurrent microwave heating and brachytherapy treatment of large area surface disease such as chest wall recurrence of breast cancer has already been reported Citation[8],Citation[13–16]. This prototype combination applicator is a conformal vest with the following layers: thermal mapping catheters for tissue surface thermometry, a temperature stabilising water bolus, a conformal microwave array (CMA) applicator with 915 MHz dual concentric conductor (DCC) aperture antennas Citation[17], a dielectric spacer to maintain distance between the tissue surface and high-dose rate (HDR) brachytherapy sources, HDR brachytherapy catheters and an inflatable air bladder for securing the applicator to the contoured patient anatomy Citation[8],Citation[14],Citation[16]. The water bolus is a crucial component of the combination superficial hyperthermia applicator. There are numerous papers that discuss the influence of the water coupling bolus on tissue surface temperature inhomogeneity and tissue specific absorption rate (SAR) during hyperthermia treatment Citation[18–24]. Guidelines for using a water bolus during superficial hyperthermia have been proposed Citation[22] and several water bolus designs have been reported for use with different microwave hyperthermia devices Citation[14],Citation[20],Citation[23],Citation[25]. The water bolus used with multi-element electromagnetic applicators must be thin (0.5–2 cm depending on frequency) or the bolus layer will support spurious surface waves and volume oscillations that distort the specific absorption rate (SAR) pattern inside the tissue Citation[21],Citation[25–29]. The water circulating through the bolus provides a finite thickness homogeneous dielectric interface that couples the electromagnetic (EM) waves into the high dielectric constant tissue. Besides EM coupling, the temperature-controlled deionised water circulating inside the bolus cools the skin surface and minimises spatial variation in tissue surface temperature during the hour-long hyperthermia treatments.

In our superficial hyperthermia treatments of chest wall recurrence, the target tissue (typically 10–15 mm deep including skin) is elevated to 40–45°C using a 915 MHz multi-element array applicator to heat at depth, and a thin bolus with circulating 42.5°C water is used to improve uniformity of skin surface temperature Citation[27–30]. The fluid dynamics of the water bolus used with the DCC array have been shown to play a crucial role in regulating the surface temperature distribution Citation[31]. As the size of multi-element array applicators increases to treat larger areas, the ability to generate uniform flow across the bolus and thereby maximise surface temperature homogeneity becomes increasingly difficult, especially for simple bolus designs with small inlet and outlet ports. Larger bolus sizes exhibit pressure loss along the flow channels which produces non-uniform flow distribution unless the bolus design is improved. Regions inside the water bolus with poor fluid flow have low thermal transfer to skin due to slow convective heat exchange with the circulating water. In addition, the longer transit time of water in low flow regions of the bolus allows increased absorption of energy from the microwave antennas. As a result, surface temperature increases in low flow regions, resulting in hot spots on the skin during hyperthermia treatment Citation[31].

Thus, it is important to maintain a nearly uniform flow distribution across the bolus to regulate tissue surface temperature homogeneity during hyperthermia treatment. Maintaining uniform flow distribution across a bolus with perforated inlet and outlet flow channels and a finite thickness open cell filter foam (bolus active area) like the one used in our hyperthermia clinic requires a nearly constant lateral pressure distribution inside the flow channels to maintain monotonic pressure drop across the bolus active area. In this paper we present 3D computational fluid dynamics simulations and experimental measurements that investigate the flow dynamics of the water bolus and propose an improved water bolus design for a 19 × 32 cm hyperthermia applicator intended to heat large-area superficial disease.

Methods

Fluid flow inside a thin layer of rectangular water bolus with a 19 × 32 cm active area suitable for coupling a 3 × 6 DCC microwave antenna array to tissue was analysed as a manifold fluid network Citation[32–34]. Parametric analyses of the bolus design variables were studied using 3D computational fluid dynamics simulations and pressure distributions inside the bolus were used to design a prototype 19 × 32 cm conformal water bolus with improved fluid flow and surface temperature distribution.

Single inletsingle outlet (SISO) water bolus

shows the schematic of a water bolus with single inlet and single outlet flow channels used in the combination CMA applicator Citation[8]. Layers of flexible medical grade PVC material (3 mm thickness) were heat-sealed together around the edges to form a reusable and highly conformal bolus bag. Deionised water was circulated through the inlet and outlet channels of the bolus which contained 6 mm thick open cell filter foam. Numerical simulations and experimental measurements of SAR inside tissue-equivalent phantoms have shown the optimum bolus thickness for 915 MHz DCC apertures to be between 5–10 mm Citation[27–29]. The thin layer of filter foam maintained a finite thickness coupling for optimal SAR deposition from the microwave antenna array and a constant distance between the HDR brachytherapy sources and tissue surface Citation[15]. The 1-cm wide inlet water channel on one side of the bolus-distributed water into the porous filter foam (bolus active area) through 3-mm wide openings spaced at 10 mm intervals. The water flowing through the porous medium was collected on the opposite side of the bolus using a 0.635 cm diameter (1/4″ ID) tubing with approximately 3 mm diameter perforations at 10 mm interval. Water circulation was maintained using a peristaltic pump so that water inflow and outflow rates through the bolus are identical.

Figure 1. Schematic illustration of the bolus with single inlet and single outlet (SISO) flow channels. The 3 × 6 DCC aperture antenna array with coaxial connectors is overlaid to indicate the 19 × 32 cm bolus area.

Figure 1. Schematic illustration of the bolus with single inlet and single outlet (SISO) flow channels. The 3 × 6 DCC aperture antenna array with coaxial connectors is overlaid to indicate the 19 × 32 cm bolus area.

Improved water bolus design

The SISO bolus configuration of is a reverse manifold fluid network Citation[32–34]. Water flowing through the bolus flow channels experiences frictional loss at the rigid wall resulting in pressure drop along the flow path. At each opening in the inlet manifold, part of the water flowing through the orifice reduces fluid momentum resulting in pressure increase Citation[35–36]. Friction and momentum effects inside the inlet manifold have opposing effects causing variation in pressure along the fluid path Citation[35–36]. On the other hand, in the outlet channel, frictional losses along the fluid path and increasing fluid momentum due to confluence of fluid streams from successive orifices combine to reduce pressure along the fluid path Citation[32–36]. A nearly uniform pressure distribution inside the flow channels will maintain a monotonic pressure drop and uniform fluid flow distribution across the bolus. Simple design techniques such as tapering inlet/outlet flow channels to compensate for change in fluid flow velocity due to pressure drop along the fluid path Citation[37], or changing the individual orifice size Citation[38] do not prove sufficient to maintain uniform flow distribution over a range of fluid flow rates possible with this size bolus. Thus, it is desirable to construct a bolus bag that yields satisfactory performance over the desired range of flow rates with acceptably low variation in fluid velocity across the bolus.

Design variables

In order to minimise variation in fluid pressure within the flow channels and maintain a nearly uniform flow distribution across the bolus active area, flow channels with larger cross-sectional area were investigated. The following parameters were analysed to improve the fluid dynamics of the 19 × 32 cm rectangular bolus: inlet channel width (Ds), diameter of the outlet channel (Dr), width of the skip seals (a0) in the inlet channel and diameter of the holes (d0) in the outlet channels. identifies the bolus design variables and the 19 × 32 cm active bolus area of the 3 × 6 DCC array. The length of the inlet and outlet channels was fixed by the dimension of the applicator array. Conformity to the body surface, comfort and physical dimensions of the CMA applicator were considered in determining the following limits for the bolus design variables: DsCitation[1],Citation[3] cm, Dr ≤ 1.27 cm. The upper limit on outlet channel diameter (Dr) was set to match the dimension of a readily available half-inch diameter vacuum-rated tubing with a bend radius ≤5 cm.

The upper limit on Ds was chosen in order to fit patient anatomy while the lower limit was based on previous bolus design which is known to provide minimally adequate flow Citation[31]. For the new water bolus design, the orifice dimensions in the inlet and outlet channels (a0, d0) were increased from 3 to 5 mm with 10 mm spacing. Larger openings in the outlet channel reduce flow resistance inside the return tubing due to increase in flow cross-section. On the inlet channel they reduce flow resistance and also the mechanical stress on the narrow skip seals between each orifice. Increasing orifice width any further while maintaining 10-mm spacing interval (>50% opening) was observed to deteriorate flow distribution due to increased fluid momentum from the first few orifices along the fluid path.

Fluid dynamics simulations

Mathematical model

Steady state fluid flow inside the water bolus is studied by coupling the incompressible Navier-Stokes equation in the flow channels with Brinkman's equation for porous media inside the open cell filter foam. Free flow of water through the bolus inlet and outlet channels and orifices during steady state is studied using incompressible Navier-Stokes and continuity equations, given by Citation[39],

The Navier-Stokes equation (1a) represents conservation of momentum and the continuity equation (1b) represents conservation of mass of the circulating water with constant dynamic viscosity (η), density (ρ), velocity vector, u and pressure, p. Steady state fluid flow through the porous filter foam in the bolus active area is studied using Brinkman's and continuity equations Citation[40],respectively where ε and κ are the porosity and permeability of the filter foam. Steady state pressure and velocity of the circulating water is studied by solving equations (1–2) using the following boundary conditions:at the rigid walls of the bolus,at the exit port of the outlet channel and continuity of fluid flow at the interface between free flow and flow through a porous medium. Simulations were carried out for laminar inflow of Q [L/min] through the inlet port.

Simulations

An equivalent 3D model of the rectangular bolus was created to study fluid flow dynamics through the water bolus. shows a SISO bolus model used in the simulation. During water circulation, the RF welded inlet channel and skip seals in the PVC bolus bag take a cylindrical shape. Thus, the inlet channel of width Ds and skip seals of width a0 were modelled as cylinders with diameters, 2Ds/π and 2a0/π respectively. The length of the inlet channel and skip seals were 34 cm and 0.5 cm respectively in the model. The outlet tubing was modelled as a 0.34 cm long cylinder with inner diameter, Dr and wall thickness equal to 0.3175 cm (1/8″). The holes on the outlet tubing connecting the porous medium were modelled as 0.3175 cm long cylinders with diameter, d0 spaced every 10 mm. The porous medium containing open cell filter foam is modelled as 19 × 32 cm cuboid, of thickness t = 6 mm.

Figure 2. Numerical model of a SISO bolus illustrating the peripheral inlet and outlet flow channels and central ‘active area’ of the bolus available for coupling a hyperthermia applicator to tissue surface.

Figure 2. Numerical model of a SISO bolus illustrating the peripheral inlet and outlet flow channels and central ‘active area’ of the bolus available for coupling a hyperthermia applicator to tissue surface.

Fluid dynamics simulations of the bolus steady state flow were carried out for the following cases: (a) SISO bolus of with a0 = 3 mm, d0 = 3 mm, Dr = 0.635 cm (1/4″) and Ds = 1 cm–existing design Citation[14], (b) SISO bolus with wider flow channels; a0 = 5 mm, d0 = 5 mm, Dr = 1.27 cm and Ds varying between 1 and 3 cm, and (c) DIDO water bolus with wider flow channels; a0 = 5 mm, d0 = 5 mm, Dr = 1.27 cm and Ds varying between 1 and 3 cm. The wider cross section flow channels in models (b) and (c) were investigated for their potential to improve flow distribution across the bolus. The SISO bolus models of (a)-(b) are reverse manifold fluid networks Citation[32–34]. The DIDO bolus configuration of model (c) combines the reverse and parallel manifold fluid networks Citation[32–34]. summarises the fluid dynamics simulations studied in this work.

Table I.  Fluid dynamics models simulated for possible flow distributions across a 19 × 32 cm water bolus.

Comsol Multiphysics (Comsol, Burlington, MA), a finite element-based simulation software, was used to solve the combined Navier-Stokes and Brinkman Equations (1–2) for the boundary conditions in Equation (3). Steady state pressure and velocity distributions in the numerical simulations were calculated for 2 L/min total inflow rate. The DIDO bolus simulations were carried out for 2 L/min total inflow that was equally split between the two inlet ports of the bolus. The following values were used in the simulations: p0 = 0, ρ = 991.46 kg/m3, and η = 6.3e-4 kg/m/s for 42.5°C water, ε = 0.85 and κ = 3.3e-7 m2 for 15–25 ppi open cell filter foam Citation[40]. As the flow dynamics in the bolus active area is studied in terms of internal fluid pressure, normalised pressure defined as,was computed in the central plane (z = t) of the bolus active area for all models. In (4), Pmin and Pmax refer to the minimum and maximum values of steady state pressure in the mid plane (z = t) of the bolus active area. Let and be the normalised pressure in Equation (4) evaluated at y = y1 and y = y2 where y1 and y2 denote the start and end locations of the bolus active area containing the open cell filter foam. The location of the filter foam is indicated by the dotted lines in . When and , a monotonic pressure gradient exists across the bolus leading to nearly uniform flow distribution. The fluid dynamics of the bolus models were assessed based on the normalised pressure curves, and .

Experimental evaluation of bolus fluid dynamics

Results of the simulations were used to finalise the design of a new conformal DIDO water bolus (19 × 32 cm). A prototype following this design was fabricated by Bionix Development Corp (Paoli, PA) as a part of a collaborative research project with Duke University. Since it is difficult to measure flow or pressure inside the thin water bolus, thermal image sequences of the bolus surface were used to assess steady state flow and surface temperature distributions of the bolus under forced convection. Temperature-controlled water was circulated through the bolus bags lying on a flat surface using a peristaltic pump (Cole-Parmer, Vernon Hills, IL). For the prototype DIDO bolus, water was fed to both inlet channels using a Tee splitter in the supply tubing and water was pulled from both ends of the outlet tube using a Tee combiner on the return side connected to the peristaltic pump. The bolus thickness during water circulation was maintained at 6 mm and temperature distribution on the bolus surface was monitored from above using an infrared (IR) camera (Mikron 7500, Mikron Infrared, Oakland, NJ) with ±2% measurement accuracy. The temperature of the water flowing through the bolus was suddenly switched to a temperature-controlled water bath that was 5°C above the initial temperature. Thermal image sequences of the bolus surface were recorded using the IR camera before and after this step change in circulating water temperature at a steady state fluid flow rate of 2 L/min. Thermographic images of the bolus surface temperatures were used to assess the steady state fluid flow dynamics and quantify surface temperature variation of the SISO and DIDO water boluses.

Results

Fluid dynamics simulations

Case a

shows the steady state velocity field computed in the mid plane of the SISO bolus of for 2 L/min inflow. The arrows in denote the direction and magnitude of the water velocity at steady state. shows the normalised pressure curves, and computed at the beginning and end of the active area. In and water enters the inlet channel at x = L (L = 32 cm) from above and exits via the port on the outlet channel located at the bottom right (x = L), and x = 0 corresponds to the closed end of the inlet and outlet channels. Friction along the flow path and narrow cross section of the flow channels in the SISO bolus of resulted in significant pressure drop () with maximum fluid momentum through few orifices near the inlet and outlet ports Citation[33],Citation[35]. Thus, the water entering the inlet manifold disperses through the numerous openings at a non-linearly decreasing flow rate along the fluid path (). Once inside the bolus, water suction through the holes in the outlet tubing with narrow flow cross-section causes a non-uniform flow distribution with preferential volume flowing through the holes nearest to the outlet port at bottom right ().

Figure 3. Steady state velocity field in the central plane of the SISO bolus model (a) of , calculated for 2 L/min flow rate.

Figure 3. Steady state velocity field in the central plane of the SISO bolus model (a) of Figure 1, calculated for 2 L/min flow rate.

Figure 4. Normalised pressure above and below the 19 × 32 cm bolus active area computed for the SISO bolus model of .

Figure 4. Normalised pressure above and below the 19 × 32 cm bolus active area computed for the SISO bolus model of Figure 1.

Case b

shows the normalised pressure curves of the SISO bolus model with wider flow channels; DsCitation[1],Citation[3] cm, Dr = 1.27 cm, d0 = 5mm and a0 = 5 mm with equi-spaced orifices in the flow channels. In , increasing the inlet channel width (Ds) increased the surface area of the channel and reduced pressure drop due to friction along the fluid path Citation[32]. Subsequently, it improved the pressure distribution in the inlet channel and above the bolus active area, . From it can be observed that the effect of fluid momentum on internal fluid pressure is negligible compared to frictional effect along the fluid path. However, pressure variation on the bolus return side, increased with increase in inlet channel width, Ds. This is because, on the return side, friction from the tube wall and increase in fluid momentum due to confluence of numerous fluid streams act together to reduce the pressure along the flow path Citation[32–36]. Sufficiently larger diameter tubing is required on the return side to minimise variation in fluid pressure in the outlet channel and subsequently inside the bolus active area. Furthermore, since it is a closed fluid network, the flow cross-section of the outlet channel (Dr) should be larger than that of the inlet channel in order to maintain nearly uniform flow distribution across the bolus Citation[32–33],Citation[35]. Thus, the variation in pressure on the exit side of the bolus active area, increased in , as Ds was increased.

Figure 5. Normalised pressure above and below the 19 × 32 cm bolus active area computed for the SISO bolus model b of .

Figure 5. Normalised pressure above and below the 19 × 32 cm bolus active area computed for the SISO bolus model b of Table I.

Case c

shows the normalised pressure curves computed for the DIDO bolus model c in . Similar to the observations made for the SISO bolus, normalised fluid pressure across the active area improved in as the inlet channel width (Ds) and return tubing diameter (Dr) were increased compared to the SISO bolus model, (a) of . In addition to the use of wider flow channels, water inflow from opposing directions and suction from both ends of the outlet tubing effectively reduced the pressure variation inside the flow channels and across the bolus active area. For values of Ds greater than 1 cm, pressure distribution above the active area, in approached unity. On the return side of the bolus, reduction in fluid path length due to suction from both ends of the return tubing reduced the friction and fluid momentum effects on fluid pressure. Thus, variation in pressure on the return side of the bolus active area in significantly reduced for the DIDO bolus models.

In an effort to reduce variation in fluid pressure away from the exit ports, without increasing the upper limit on Dr, simulations were carried out for the DIDO water bolus with non-uniform orifice spacing on the return tubing. Subsequently, fluid flow through a DIDO bolus with Ds = 3 cm and d0 = 5 mm, a0 = 5 mm was simulated for Dr = 1.27 cm (half an inch) and variable spacing of holes in the return tubing. Logarithmic spacing was chosen to counteract the non-linear increase in pressure on the bolus return side (). The holes in the outlet channel were symmetrically located about the centre (x = L/2; L = 32 cm) with closer spacing towards the centre (x = L/2) and sparser holes near the tube ends (x = 0, L) of the return tubing as shown in .

Figure 6. Normalised pressure above () and below () the 19 × 32 cm bolus active area computed in the mid plane of the DIDO bolus model c of .

Figure 6. Normalised pressure above () and below () the 19 × 32 cm bolus active area computed in the mid plane of the DIDO bolus model c of Table I.

shows the steady velocity field in the mid plane (z = t) of the DIDO bolus with log-spaced holes in the outlet channel. indicates nearly monotonic flow distribution across the bolus. The solid curves in denote the normalised fluid pressure , computed on either sides of the active area of the DIDO bolus with log-spaced return holes. The low variation in fluid pressure across the filter foam in maintained a monotonic pressure drop and nearly uniform fluid flow across the 19 × 32 cm active area. shows the steady state velocity profiles in the central plane of the DIDO water bolus with logarithmically spaced orifices and larger diameter return tubing (Dr = 1.27 cm) computed parallel to the centre of each row of the 3 × 6 DCC array for 2 and 4 L/min flow rates. Simulation results in indicate uniform fluid distribution across the bolus, and a proportional increase in velocity at higher flow rate.

Figure 7. Steady state velocity field in the mid plane of the DIDO bolus with Ds = 3 cm, a0 = 5 mm, d0 = 5 mm, Dr = 1.27 cm and logarithmically spaced holes in the outlet channel.

Figure 7. Steady state velocity field in the mid plane of the DIDO bolus with Ds = 3 cm, a0 = 5 mm, d0 = 5 mm, Dr = 1.27 cm and logarithmically spaced holes in the outlet channel.

Figure 8. Steady state velocity field inside the active area of the DIDO bolus of . Velocity profiles (a)-(c) were calculated for the cross-sectional cuts in the bolus mid plane parallel to the centre of each row of the 3 × 6 DCC array for 2 and 4 L/min flow rates as shown in (d).

Figure 8. Steady state velocity field inside the active area of the DIDO bolus of Figure 7. Velocity profiles (a)-(c) were calculated for the cross-sectional cuts in the bolus mid plane parallel to the centre of each row of the 3 × 6 DCC array for 2 and 4 L/min flow rates as shown in (d).

New prototype water coupling bolus

shows the prototype PVC DIDO water bolus bag fabricated based on the numerical simulation outcome. The prototype water bolus in was fabricated with the design values of the DIDO bolus model in . The flow channels and skip seals on the inlet side of the PVC water bolus in were sealed by RF welding. PVC tubing with 5 cm bending radius, sufficiently rigid to remain open under negative pressure during water circulation and still conform to contoured body surface, was used for the return tubing. Vinyl medical micro tubings (Scientific Commodities, Lake Havasu City, AZ) were sealed to the patient contacting side of the water bolus for tissue surface temperature monitoring and feedback power control of the DCC antenna array during treatment.

Experimental confirmation of bolus flow dynamics

shows the propagation of the isothermal contour on the surface of the SISO bolus of as the temperature of the circulating water was switched at steady state flow. The time rate of propagation of the isothermal contour in indicates non-uniformity in fluid velocity with preferential volume flowing through the holes near the inlet and outlet ports of the flow channels which is in qualitative agreement with the simulation (). Temperature variation across the SISO bolus at thermal steady state prior to switching the water bath was measured as 1.5°C with the highest variation on the left side of the bolus where the flow is much slower.

Figure 9. Prototype DIDO water coupling bolus with dual opposing inlet and outlet flow channels and logarithmically spaced orifices in the return tubing fabricated based on the fluid dynamics simulation outcome; design values: Ds = 3 cm, a0 = 5 mm, d0 = 5 mm, Dr = 1.27 cm.

Figure 9. Prototype DIDO water coupling bolus with dual opposing inlet and outlet flow channels and logarithmically spaced orifices in the return tubing fabricated based on the fluid dynamics simulation outcome; design values: Ds = 3 cm, a0 = 5 mm, d0 = 5 mm, Dr = 1.27 cm.

shows the isothermal contours of the DIDO bolus of . The time rate of propagation of the isothermal contour in indicates a nearly uniform water velocity with acceptably low variation across the bolus active area. The low variation in fluid flow across the DIDO bolus maintained a low variation in bolus surface temperatures (<0.8°C). Since the water flowing through the DIDO bolus was split using a Tee splitter on the inlet side and combined using a Tee combiner on the bolus return side, the propagation of the thermal contour in is slow compared to the flow pattern of the SISO bolus in . It should be noted that the measured flow patterns in observed on the bolus surface is due to convective and conductive heat transfer of the temperature-controlled circulating water. Thus, the flow pattern measurements of are only a qualitative measure of the bolus flow dynamics, unlike the fluid velocity fields computed in the simulations. The discrepancy in the simulated () and measured () DIDO bolus flow dynamics is acceptably low and indicates the difference between an ideal numerical model and real world. Nonetheless, the steady state flow pattern and surface temperature distributions across the 19 × 32 cm bolus are superior for the new DIDO water bolus compared to the previously used SISO bolus.

Figure 10. Propagation of isothermal contour on the bolus surface for a step change in the circulating water temperature captured for the (A) SISO and (B) prototype DIDO water boluses at steady state fluid flow and 2 L/min flow rate.

Figure 10. Propagation of isothermal contour on the bolus surface for a step change in the circulating water temperature captured for the (A) SISO and (B) prototype DIDO water boluses at steady state fluid flow and 2 L/min flow rate.

Discussion

The SISO water bolus of initially designed for use with multi-element DCC array applicator is adequate to maintain sufficient flow and temperature uniformity across smaller SISO bolus bags. The velocity and pressure distributions of the SISO bolus in for a 19 × 32 cm active area indicate non-uniform flow for larger size applicators. The tissue beneath the low flow regions of the SISO bolus experiences poor thermal transfer with the slow circulating water leading to thermal hot spots when irradiated with EM energy Citation[31]. As applicator size increases, the pressure gradient across the bolus increases, resulting in increasingly non-uniform flow and inhomogeneous tissue surface temperature distributions, unless the flow channels are optimised to operate at higher flow rates. Several papers on the guidelines for using a water bolus Citation[22],Citation[25] and surface temperature distributions maintained by different bolus configurations Citation[18–21], Citation[23–27] describe the importance of the water coupling bolus in hyperthermia treatments. A detailed analysis of the convective heat transfer across the SISO bolus on tissue surface temperature distribution when irradiated by the DCC microwave antenna array is reported in Citation[31]. The analysis revealed the limitations of the SISO water bolus configuration for large applicator arrays and higher flow rates. Thus, we studied steady state bolus flow dynamics for varying values of bolus design variables and flow configurations (SISO and DIDO) to aid in the fabrication of conformal water bolus with nearly uniform flow and temperature distribution for use with large multi-element hyperthermia applicators.

Pressure curves of the SISO bolus model with wider flow channels in indicate a reduced variation in and an increased variation in as the channel width (Ds) was increased. The increase in flow cross-section on the inlet channel due to increase in Ds reduced the pressure drop along the fluid flow path and laterally above the active area Citation[32]. On the return side, variation in increased with increase in Ds because the flow cross-section of the outlet channel (Dr) was not large enough to minimise the friction and fluid momentum effects acting on the fluid pressure Citation[33–35]. Being a reverse manifold fluid network Citation[33–35], sufficiently larger diameter tubing is required on the return side of the SISO bolus to minimise variation in fluid pressure in the outlet channel and subsequently inside the active bolus area.

Due to practical design limitations on the return tubing diameter, an alternate flow configuration with dual opposing feed and suction ports was studied in an attempt to minimise pressure variation along the flow channels of the 19 × 32 cm bolus. This DIDO bolus configuration is a combination of reverse and forward manifold fluid flow networks Citation[33–35]. In the DIDO bolus models investigated in , superposition of equal volume fluid flowing from opposite directions compensated the pressure drop along the flow path. The variation in pressure along the inlet channel lowered with increasing inlet channel width (Ds) as observed in . On the outlet side, suction from both ends of the return tubing reduced the fluid path length towards the exit ports. This reduction in path length reduced the frictional loss and fluid momentum effects acting on fluid pressure Citation[33],Citation[35], resulting in a significantly lower variation in () compared to the SISO bolus (). The variation in the pressure curves of () resulted in a progressively decreasing flow rate through the perforated filter foam with lower flow towards the centre (x = L/2) compared to the ends (x = 0, x = L). The decrease in flow towards the centre of the bolus active area was minimised by changing to a logarithmic spacing of holes on the return tube without having to increase the outlet channel diameter (Dr). Simulation data of and of the DIDO bolus with logarithmically spaced holes in the outlet channel indicated nearly uniform fluid flow across the 19 × 32 cm bolus with and (solid curves in ), and proportional increase in fluid velocity over the desired range of flow rates.

The steady state flow pattern of the SISO water bolus with narrow flow channels in indicate non-uniform flow with a relatively higher flow on the right and reduced flow on the left. Empirically derived steady state flow dynamics of the SISO water bolus showed qualitative agreement with the simulation data of (arrows). Regions with lower flow in resulted in surface temperature variation as high as 1.5°C across the SISO bolus. The isothermal contours of the prototype DIDO water bolus of indicated a nearly uniform flow across the bolus with acceptably low variation in . The surface temperature variation across the 19 × 32 cm DIDO bolus was less than 0.8 C at steady state, almost 50% improvement in thermal homogeneity. The steady state fluid dynamics of the prototype DIDO water bolus in is in qualitative agreement with the simulation data of . Compared to the SISO water bolus, measured variations in steady state fluid flow and surface temperature across the 19 × 32 cm bolus are acceptably low for the new DIDO water bolus. Nearly uniform bolus flow distribution at higher flow rates is expected to improve convective heat transfer across the tissue surface and minimise surface thermal heterogeneities during hyperthermia treatment of large area tissue disease using the DCC array.

Conclusion

A new conformal water bolus configuration allowing higher flow rates without turbulence and improved fluid flow distribution over a large surface area was studied for large multi-element superficial hyperthermia applicators. Numerical simulations indicate that the bolus flow distribution is characterised solely by internal fluid pressure along the flow channels and nearly uniform bolus flow distribution can be maintained by minimising variation in the internal fluid pressure across the bolus active area. Nearly uniform flow distribution across the bolus leads to nearly uniform convective heat transfer from the circulating temperature-controlled water. This enables the water coupling bolus to effectively regulate tissue surface temperature inhomogeneity during hyperthermia treatment, particularly as the applicator size is increased to treat large area superficial disease.

In this effort we have shown that water bolus with nearly uniform flow and low variation in surface temperature distribution can be designed for large superficial hyperthermia applicators using wider dual opposing flow channels as in . Steady state fluid flow patterns of the SISO and prototype DIDO water boluses show significant improvement in flow distribution, in agreement with fluid dynamics simulation data. With the improvement in flow distribution, surface temperature variation over the 19 × 32 cm active area of the DIDO bolus was reduced below 0.8°C. Bolus flow pattern measurements indicate that the new DIDO water bolus configuration with larger diameter return tubing is capable of maintaining a nearly uniform flow distribution and effectively homogenising surface temperatures under large array applicators used in hyperthermia treatment of chest wall recurrence of breast cancer.

Acknowledgements

This effort was supported by NIH Grants R44 CA104061 and RO1 CA70761. The authors would like to express their appreciation of the work of Edward Markowitz and Tamara Masters at Bionix Development Corp for their contribution to bolus fabrication. The authors also thank Roberto Berruti from Ditta Schinetti, Torino, Italy for his suggestions on manifold feeds.

Declaration of interest: The author reports no conflicts of interest. The author alone is responsible for the content and writing of the paper.

References

  • Jones EL, Oleson JR, Prosnitz LR, Samulski TV, Vujaskovic ZV, Yu D, Sanders LL, Dewhirst MW. Randomized trial of hyperthermia and radiation for superficial tumors. J Clin Oncol 2005; 23: 3079–3085
  • Kapp DS. Efficacy of adjuvant hyperthermia in the treatment of superficial recurrent breast cancer: Confirmation and future directions. Int J Radiat Oncol, Biol Phys 1996; 35: 1117–1121
  • Sneed PK, Stauffer PR, McDermott MW, Diederich CJ, Lamborn KR, Prados MD, Chang S, Weaver KA, Spry L, Malec MK, et al. Survival benefit of hyperthermia in a prospective randomized trial of brachytherapy boost +/− hyperthermia for glioblastoma multiforme. Int J Radiat Oncol, Biol Phys 1998; 40: 287–295
  • Vernon CC, Hand JW, Field SB, Machin D, Whaley JB, van der Zee J, van Putten WLJ, van Rhoon GC, van Dijk JDP, Gonzalez-Gonzalez D, et al. Radiotherapy with or without hyperthermia in the treatment of superficial localized breast cancer: Results from five randomized controlled trials. Int J Radiat Oncol, Biol Phys 1996; 35: 731–744
  • Dewey WC, Freeman ML, Raaphorst GP, Clark EP, Wong RSL, Highfield DP, Spiro IJ, Tomasovic SP, Denman DL, Coss RA. Cell biology of hyperthermia and radiation. Radiation Biology in Cancer Research, RE Meyn, HR Withers. Raven Press, New York 1980; 589–621
  • Overgaard J. Simultaneous and sequential hyperthermia and radiation treatment of an experimental tumor and its surrounding normal tissue in vivo. Int J Radiat Oncol Biol Phys 1980; 6: 1507–1517
  • Overgaard J. The current and potential role of hyperthermia in radiotherapy. Int J Radiat Oncol Biol Phys 1989; 16: 535–549
  • Juang T, Stauffer PR, Neuman DG, Schlorff J. Multilayer conformal applicator for microwave heating and brachytherapy treatment of superficial tissue disease. Int J Hyperthermia 2006; 22: 527–544
  • Moros EG, Straube WL, Klein EE, Yousaf M, Myerson RJ. Simultaneous delivery of electron beam therapy and ultrasound hyperthermia utilizing scanning reflectors: A feasibility study. Int J Radiat Oncol Biol Phys 1995; 31: 893–904
  • Myerson RJ, Straube WL, Moros EG, Emami BN, Lee HK, Perez CA, Taylor ME. Simultaneous superficial hyperthermia and external radiotherapy: Report of thermal dosimetry and tolerance to treatment. Int J Hyperthermia 1999; 15: 251–266
  • Novak P, Moros EG, Straube WL, Myerson RJ. SURLAS: A new clinical grade ultrasound system for sequential or concomitant thermoradiotherapy of superficial tumors: Applicator description. Med Phys 2005; 32: 230–240
  • Straube WL, Klein EE, Moros EG, Low DA, Myerson RJ. Dosimetry and techniques for simultaneous hyperthermia and external beam radiation therapy. Int J Hyperthermia 2001; 17: 48–62
  • Stauffer PR, Schlorff JL, Juang T, Neuman DG, Johnson JE, Maccarini PF, Pouliot J. Progress on system for applying simultaneous heat and brachytherapy to large-area surface disease. Proc. of SPIE, TP Ryan, 2005; 5698: 82–96
  • Juang T, Neuman DG, Schlorff JL, Stauffer PR. Construction of a conformal water bolus vest applicator for hyperthermia treatment of superficial skin cancer. International Conference of the IEEE Engineering in Medicine and Biology Society, 2004. IEEE Press, San Francisco 2004
  • Taschereau R, Stauffer PR, Hsu IC, Schlorff JL, Milligan AJ, Pouliot J. Radiation dosimetry of a conformal heat-brachytherapy applicator. Technology in Cancer Research and Treatment 2004; 3: 347–358
  • Arunachalam K, Craciunescu OI, Maccarini PF, Schlorff JL, Markowitz E, Stauffer PR. Progress on thermobrachytherapy surface applicator for superficial tissue disease, Energy-based Treatment of Tissue and Assessment V. Proceedings SPIE 2009; 7181: 71810C–71810C-12
  • Rossetto F, Diederich CJ, Stauffer PR. Thermal and SAR characterization of multielement dual concentric conductor microwave applicators for hyperthermia: A theoretical investigation. Med Phys 2000; 27: 745–753
  • Lee ER, Kapp DS, Lohrbach AW, Sokol JL. Influence of water bolus temperature on measured skin surface and intradermal temperatures. Int J Hyperthermia 1994; 10: 59–72
  • Chan KW, McDougall JA, Chou CK. FDTD simulations of Clini-Therm applicators on inhomogeneous planar tissue models. Int J Hyperthermia 1995; 11: 809–820
  • Chou CK, McDougall JA, Chan KW, Luk KH. Evaluation of captive bolus applicators. Med Phys 1990; 17: 705–709
  • De Bruijne M, Samaras T, Bakker JF, van Rhoon GC. Effects of water bolus size, shape and configuration on the SAR distribution pattern of the Lucite cone applicator. Int J Hyperthermia 2006; 22: 15–28
  • Van Der Gaag ML, De Bruijne M, Samaras T, Van Der Zee J, Van Rhoon GC. Development of a guideline for the water bolus temperature in superficial hyperthermia. Int J Hyperthermia 2006; 22: 637–656
  • Sherar MD, Clark H, Cooper B, Kumaradas J, Liu FF. A variable microwave array attenuator for use with single-element waveguide applicators. Int J Hyperthermia 1994; 10: 723–731
  • Sherar MD, Liu FF, Newcombe DJ, Cooper B, Levin W, Taylor WB, Hunt JW. Beam shaping for microwave waveguide hyperthermia applicators. Int J Radiat Oncol Biol Phys 1993; 25: 849–857
  • Diederich CJ, Stauffer PR, Bozzo D. An improved bolus configuration for commercial multielement ultrasound and microwave hyperthermia systems. Med Phys 1994; 21: 1401–1403
  • Gelvich EA, Mazokhin VN. Contact flexible microstrip applicators (CFMA) in a range from microwaves up to short waves. IEEE Trans Biomed Eng 2002; 49: 1015–1023
  • Newman DG, Diederich CJ, Stauffer PR, Bozzo D. SAR pattern perturbations from resonance effects in water bolus layers used with superficial microwave hyperthermia applicators. Int J Hyperthermia 2002; 18: 180–193
  • Rossetto F, Stauffer PR. Effect of complex bolus-tissue load configurations on SAR distributions from dual concentric conductor applicators. IEEE Trans Biomed Eng 1999; 46: 1310–1319
  • Rossetto F, Stauffer PR, Manfrini V, Diederich CJ, Biffi GG. Effect of practical layered dielectric loads on SAR patterns from dual concentric conductor microstrip antennas. Int J Hyperthermia 1998; 14: 513–534
  • Stauffer PR. Thermal therapy techniques for skin and superficial tissue disease. A Critical Review, Matching the Energy Source to the Clinical Need, Vol. CR75, TP Ryan. SPIE Optical Engineering Press, Bellingham, WA 2000; 327–367
  • Birkelund Y, Jacobsen S, Arunachalam K, Maccarini P, Stauffer PR. Flow patterns and heat convection in a rectangular water bolus for use in superficial hyperthermia. Phys Med Biol 2009; 54: 3937–3953
  • Baijura RA. A model for flow distribution in manifolds. J Eng power 1971; 93: 7–12
  • Johnson ST. Analysis of manifold fluid flow networks for air and liquid flow through modular electronics. Seventeenth IEEE SEMI-THERM symposium. IEEE, 2001
  • Datta AB, Majumdar AK. Flow distribution in parallel and reverse flow manifolds. Int J Heat Fluid Flow 1980; 2: 253–262
  • Acrivos A, Babcock BD, Pigford RL. Flow distribution in manifolds. J Chem Eng Science 1959; 10: 112–124
  • Baijura RA, Jones EH. Flow distribution manifolds. ASME J Fluids Eng 1976; 98: 654–666
  • Dow WM. The uniform distribution of a fluid flowing through a perforated pipe. J Applied Mech, Trans ASME 1950; 72: 431–438
  • Kubo T, Ueda T. On the characterization of divided flow and confluent flow in headers. Bull Jap Soc Mech Eng 1969; 12: 802–809
  • Zienkiewicz OC, Taylor RL, Nithiarasu P. The finite element method for fluid dynamics, 6th edition. Elsevier, OxfordUK 2005
  • Bars ML, Worster MG. Interfacial conditions between a pure fluid and a porous medium: Implications for binary alloy solidification. J Fluid Mech 2006; 550: 149–173
  • Givler RC, Altobelli SA. A determination of effective viscosity for the Brinkman-Forchheimer flow model. J Fluid Mech 1994; 258: 355–370

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.