Comparison of Central Corneal Thickness Measured by Hand-Held and Desk-Mounted Ultrasound Pachymeters in Glaucoma Patients

, &
Pages 268-271 | Received 10 Jun 2013, Accepted 20 Aug 2013, Published online: 10 Jan 2014
 

ABSTRACT

Purpose: The purpose of this study was to compare the accuracy and reproducibility of central corneal thickness measured by hand-held and desk-mounted ultrasound pachymeter in glaucoma patients under treatment. Methods: Prospective study of 65 glaucoma patients. Central corneal thickness was measured by two ultrasound pachymeters: the hand-held, portable PachPen (Accutome, Lynwood, WA), and the desk-mounted PacScan 300 (Sonomed, Lake Success, NY). Results: The mean ± SD central corneal thickness was 526.5 ± 44.8 µm and 530.0 ± 44.7 µm for the hand-held and desk-mounted pachymeters, respectively (p = 0.15). Linear regression analysis revealed a slope of 0.97 with Pearson correlation coefficient of 0.96. Bland-Altman analysis showed a mean difference of measurements by both pachymeters of 3.22 µm with two standard deviations = 9.51. Both instruments showed high intraobserver correlation: 0.972 for the hand-held pachymeter and 0.993 for the desk-mounted pachymeter. Conclusion: Central corneal thickness measurements were comparable with the use of hand-held and desk-mounted ultrasound units in glaucoma patients with good intraobserver reproducibility.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 530.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.