Sample transport by pneumatic tube system alters results of multiple electrode aggregometry but not rotational thromboelastometry

, , , &
Pages 454-461 | Received 07 Jun 2012, Accepted 31 Jul 2012, Published online: 29 Aug 2012
 

Abstract

Pneumatic tube systems (PTS) present a convenient way for blood sample transport in medical facilities. Associated preanalytical interference in various tests is largely unknown. Implementing point-of-care coagulation management at our institution, we investigated multiple electrode aggregometry (MEA) and rotational thromboelastometry (ROTEM) after PTS transportation. Whole blood samples from patients undergoing general or trauma surgery were analysed by MEA after collection (baseline, ‘0 × PTS’) and sent on a predefined PTS track (n = 12). MEA was repeated after samples travelled the track 4 (‘4 × PTS’), 8 (‘8 × PTS’) and 12 times (‘12 × PTS’) and compared with stationary controls analysed at the same time. Samples for ROTEM (n = 6) were analysed after collection and travelling the track 12 times. An acceleration detector recorded g-forces on the PTS track. At ‘0 × PTS’ no significant differences in MEA results were detected. Values were significantly lower for transported samples compared with controls (‘4 × PTS’ to ‘12 × PTS’, p < 0.001). Furthermore, MEA results of PTS samples were significantly decreased for ‘4 × PTS’ to ‘12 × PTS’ compared to baseline (p < 0.001). Except for the clotting time in EXTEM PTS transport did not significantly alter results for investigated ROTEM parameters, compared with baseline and stationary controls. Acceleration detector readout revealed alternating g-forces between −6.3 and +5.9 during transport. PTS transport caused invalid results in MEA while only one ROTEM parameter was found to be affected in this study. Variable acceleration during transport provides a potential reason for platelet activation. The authors recommend sample transport by hand or the device to be placed patient-side when MEA is performed.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access
  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart
* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.