Wear of a sequentially annealed polyethylene acetabular liner

RSA measurements in 19 hips

, , &
Pages 470-473 | Received 10 Sep 2013, Accepted 09 Jun 2014, Published online: 20 Aug 2014

Abstract

Background and purpose — We previously reported on a randomized controlled trial (RCT) that examined the effect of adding tobramycin to bone cement after femoral stem migration. The present study examined femoral head penetration into both conventional and highly crosslinked polyethylene acetabular liners in the same group of RCT patients, with a minimum of 5 years of postoperative follow-up.

Patients and methods — Linear penetration of the femoral head into an X3 (Stryker) crosslinked polyethylene (XLPE) liner was measured in 18 patients (19 hips) using radiostereometric analysis (RSA). Femoral head penetration was also measured in 6 patients (6 hips) with a conventional polyethylene liner (CPE), which served as a control group.

Results — The median proximal femoral head penetration in the XLPE group after 5.5 years was 0.025 mm with a steady-state penetration rate of 0.001 mm/year between year 1 and year 5. The CPE liner showed a median proximal head penetration of 0.274 mm after 7.2 years, at a rate of 0.037 mm/year.

Interpretation — The Trident X3 sequentially annealed XLPE liner shows excellent in vivo wear resistance compared to non-crosslinked CPE liners at medium-term implantation. The rate of linear head penetration in the XLPE liners after > 5 years of follow-up was 0.001 mm/year, which is in close agreement with the results of previous studies.

It is well established that crosslinked polyethylene (XLPE) as a bearing surface in total hip replacements undergoes substantially less wear than conventional polyethylene bearings (Bragdon et al. Citation2007a, Calvert et al. Citation2009). Concerns regarding reduced mechanical properties of XLPE, such as susceptibility to oxidation (Currier et al. Citation2007), fracture (Furmanski et al. Citation2009, Citation2011), and delamination (Bradford et al. Citation2004) ushered in a second generation of crosslinked polyethylene liners (Digas et al. Citation2007, Glyn-Jones et al. Citation2008a).

Stryker Orthopaedics (Mahwah, NJ) developed a sequential stepwise process involving moderate doses of gamma irradiation followed by annealing (heating to just below melting point) of the polyethylene in order to create their X3 crosslinked polyethylene material (Dumbleton et al. Citation2006). This iterative process forms the crosslinked polyethylene structure necessary for wear resistance while removing the majority of free radicals, which are damaging to the molecular structure. The X3 liners have shown promising mechanical strength and wear resistance in vitro (Dumbleton et al. Citation2006), with limited reports of in vivo follow-up to 5 years (Callary et al. Citation2013, D’Antonio et al. Citation2012). The purpose of the present research was to confirm the low in vivo wear of the X3 crosslinked acetabular liner at the 5-year mark.

Patients and methods

In this study, we examined patients who were enrolled in a randomized controlled trial (RCT) beginning in 2003, which examined femoral stem migration between standard and antibiotic-laden bone cements (Bohm et al. Citation2012). Ethical approval for extending this RCT to include analysis of femoral head penetration was obtained from the University of Manitoba Biomedical Research Ethics Board (B2003:108).

Surgeries were performed between 2003 and 2007 on patients over the age of 65 who required total hip replacement. An Exeter femoral stem with a modular stainless-steel femoral head of varying sizes (28 mm, 32 mm, and 36 mm) was used in all patients. A Trident (Stryker Orthopaedics) conventional polyethylene (CPE) acetabular liner was implanted into patients who were enrolled in the primary RCT between 2003 and April of 2005. The study institution then switched to use of the X3 (Stryker Orthopaedics) crosslinked polyethylene (XLPE) liner based on evidence of improved in vitro wear resistance (Wang et al. Citation2003) and because the bearing material was not the focus of the original RCT. Patients enrolled after April 2005 received the XLPE liner.

Of the original 30 patients (32 hips) recruited for the primary RCT, 25 patients (27 hips) were followed beyond 5 years; 2 patients moved away, 2 patients were lost to follow-up, and 1 patient was in poor health. Of the 27 hips, 8 were implanted with a CPE liner and 19 were implanted with an XLPE liner. In the CPE group, 6 were 28-mm diameter and 2 were 32-mm diameter. The latter were removed from the analysis of CPE head penetration, as previous studies have shown that femoral head size statistically significantly affects linear head penetration (Kesteris et al. Citation1996, Tarasevicius et al. Citation2008). In the XLPE group, 2 were 28-mm diameter, 16 were 32-mm diameter, and 1 was 36-mm diameter. Despite the different femoral head sizes, data were combined in the XLPE group, as some studies have shown that femoral head size does not statistically significantly affect linear head penetration into crosslinked polyethylene (Muratoglu et al. Citation2001, Bragdon et al. Citation2007b, Lachiewicz et al. Citation2009).

The CPE group consisted of 6 women with a median age of 70 (63–73) years at the time of surgery, and a median follow-up of 7.2 (5.8–8) years. The XLPE group consisted of 8 men and 10 women (1 bilateral) with a median age of 68 (63–85) years and a median follow-up time of 5.5 (4.9–6.0) years.

Radiostereometric examinations were performed at 6 weeks, at 6 months, and at 1, 2, 3, and > 5 years postoperatively. The 5-year RSA examinations were performed using a digital radiography system consisting of 2 ceiling-mounted X-ray sources (RAD-92; Varian Medical Systems, Palo Alto, CA) and corresponding digital detectors (Canon CXDI-55C; Canon USA, Lake Success, NY). The uniplanar calibration cage and overall RSA setup remained the same throughout the study. Radiographic analysis was performed with the UmRSA software suite version 6.0 (RSA Biomedical, Umeå, Sweden). Femoral head penetration was analyzed using automated edge-detection algorithms and was measured as the change in proximity between the center of the femoral head and the exterior of the acetabular cup—quantified as both proximal migration and a Pythagorean sum of 3-D motion, commonly termed maximum total point motion (MTPM). The rate of femoral head penetration (clinical wear rate) was calculated as the group median slope of linear regression lines computed for each individual patient (McCalden et al. Citation2009) between the 1-year and the 5-year follow-up periods. The 6-week and 6-month follow-up data were not analyzed for this study, as bedding-in (creep) of the polyethylene usually occurs within the first 6–12 months of use (Glyn-Jones et al. Citation2008b) and cannot be differentiated from wear during the same time using this technique.

We report WOMAC outcome scores obtained for each patient at each follow-up as part of the primary RCT. The WOMAC scores recorded at the 2-year follow-up and the 5-year follow-up are reported for comparison of clinical outcomes between polyethylene groups and follow-up periods.

Statistics

Due to the small sample sizes, non-parametric statistics are presented: medians and 95% confidence intervals (CI) based on ordered rank. 17 double examinations were performed in order to ascertain the clinical precision of the digital X-ray system. Precision was calculated as the width of the CI of double-examination error, a non-parametric version of the method recommended by Derbyshire et al. (Citation2009). Polyethylene groups were compared using Mann-Whitney U-tests, while Wilcoxon signed rank tests were used for comparison within groups. Significance for all statistical tests was set at p < 0.05.

Results

None of the patients who were enrolled in the primary RCT underwent a revision of their hip replacement at the latest follow-up. WOMAC scores were similar between polyethylene groups at the 2-year and 5-year follow-up periods (p = 0.9 and p = 0.5, respectively). A decline in WOMAC scores was noted between the 2-year follow-up and the 5-year follow-up in both groups, but statistical significance was not reached (CPE group: p = 0.06; XLPE group: p = 0.08).

Based on double examinations, the precision of the clinical RSA measurements was 0.06 mm (proximal migration) and 0.20 mm (MTPM). The median femoral head penetration at a median follow-up time of 5.5 years for the XLPE patients was 0.025 mm (proximal migration) and 0.197 mm (MTPM) ( and ). The median rate of femoral head penetration between year 1 and year 5 for the XLPE group was 0.001 mm/year proximally (95% CI: –0.002 to 0.002), and –0.010 mm/year for MTPM (95% CI: –0.003 to 0.001). CPE patients experienced a median femoral head penetration of 0.274 mm (proximal migration) and 0.442 mm (MTPM) at a median time of 7.2 years after surgery ( and ). The median rate of head penetration in this PE group was 0.037 mm/year (95% CI: 0.001–0.005) proximally and 0.046 mm/year (CI: 0.003–0.006 ) for MTPM.

Figure 1. Median femoral head penetration in the proximal direction with 95% CI (vertical error bars). The shaded region indicates the precision limit of the RSA system based on double examinations.

Figure 1. Median femoral head penetration in the proximal direction with 95% CI (vertical error bars). The shaded region indicates the precision limit of the RSA system based on double examinations.

Figure 2. Median and total linear femoral head penetration in all directions (MTPM) with 95% CI (vertical error bars). The shaded region indicates the precision limit of the RSA system based on double examinations.

Figure 2. Median and total linear femoral head penetration in all directions (MTPM) with 95% CI (vertical error bars). The shaded region indicates the precision limit of the RSA system based on double examinations.

Discussion

We found 0.025 mm of proximal penetration into XLPE over a median period of 5.4 years, which is substantially less than the values found by other studies examining crosslinked polyethylene bearings after 5 or more years in vivo: 0.15 mm (Digas et al. Citation2007) and 0.08 mm (Rohrl et al. Citation2007).

The rate of femoral head penetration of the X3 polyethylene bearing in the coronal plane between the 1-year and the 5-year follow-up has previously been reported to be 0.014 mm/year in vitro (Dumbleton et al. Citation2006) and 0.015 mm/year in vivo (D’Antonio et al. Citation2012) with < 0.001 mm/yr in the proximal direction (Callary et al. Citation2013). These reported rates of linear head penetration are within the range of 0.001 mm/year measured in the present study. Furthermore, the rate of total linear head penetration (MTPM) between year 1 and year 5 was –0.010 mm/year, which is almost identical to the mean rate of –0.007 mm/year reported by Callary et al. (Citation2013).

Little to no bedding-in of the polyethylene bearing (due to creep) in the proximal direction was measured within the first postoperative year (). This lack of detectable bedding-in is partially explained by the use of a reference RSA examination done 6 weeks postoperatively. As most creep has been shown to occur within the first 2–3 months of implantation (Glyn-Jones et al. Citation2008), it is possible that our delayed reference examination missed much of the creep process.

This study was limited, as the polyethylene groups were not randomized, therefore possibly introducing selection bias. Stem subsidence, not linear head penetration, was the subject of the primary RCT, which is why the patients in the present study were not randomized and why the control (CPE) group consisted of only 6 subjects. The patients enrolled in this study had a median age of 69 years at the time of surgery, which may have resulted in a reduced clinical wear rate compared to a younger and perhaps more active patient cohort. A bilateral XLPE patient was included in the analysis of linear head penetration, which may have introduced inter-patient bias into the results (Bryant et al. Citation2006). Accordingly, this patient’s data were provisionally removed, resulting in miniscule changes to the median rate of femoral head penetration (a decrease in proximal penetration of 0.006 mm/year and a decrease in MTPM penetration of 0.007 mm/year).

Different femoral head sizes in the PE groups may also have contributed some amount of error to our results. However, a number of studies have shown that femoral head size does not affect head penetration, specifically with highly crosslinked polyethylene bearings (Bragdon et al. Citation2007, Lachiewicz et al. Citation2009, Muratoglu et al. Citation2001). The opposite is true for non-crosslinked polyethylene, where smaller head sizes show reduced linear wear (Kesteris et al. Citation1996, Tarasevicius et al. Citation2008). Accordingly, the 2 patients in the CPE group with 32-mm heads were not included in the calculation of wear.

We believe that the decline in patient-reported outcomes seen in both polyethylene groups between the 2-year and the 5-year follow-up periods was the result of declining health of the patients as they aged and was not a result of poorly-performing total hip replacements. Also, the sample size of this study was too small for us to make statistically meaningful conclusions regarding patient function as it relates to choice of polyethylene.

In summary, the X3 sequentially irradiated and annealed crosslinked polyethylene acetabular liner used in this study showed excellent in vivo wear properties, as the rate of linear head penetration after 5.5 years was 0.001 mm/year in the proximal direction.

TG: radiographic examinations, measurements, analysis of data, and preparation of manuscript. MP: radiographic examinations and preparation of manuscript. TT: interpretation, final review of manuscript, and contribution of surgical cases. EB: lead investigator, lead surgeon, preparation of manuscript.

We are grateful to Stryker Orthopaedics for support in the primary RCT. We also thank Sarah Tran for her work in coordinating patient examinations and Michelle Hofer for obtaining the RSA radiographs.

  • Bohm E, Petrak M, Gascoyne T, Turgeon T. The effect of adding tobramycin to Simplex P cement on femoral stem micromotion as measured by radiostereometric analysis: a 2-year randomized controlled trial. Acta Orthop 2012; 83 (2): 115-20.
  • Bradford L, Baker DA, Graham J, Chawan A, Ries MD, Pruitt LA. Wear and surface cracking in early retrieved highly cross-linked polyethylene acetabular liners. J Bone Joint Surg (Am) 2004; 86 (6): 1271-82.
  • Bragdon CR, Kwon YM, Geller JA, Greene ME, Freiberg AA, Harris WH, et al. Minimum 6-year followup of highly cross-linked polyethylene in THA. Clin Orthop 2007a; (465): 122-7.
  • Bragdon CR, Greene ME, Freiberg AA, Harris WH, Malchau H. Radiostereometric analysis comparison of wear of highly cross-linked polyethylene against 36- vs 28-mm femoral heads. J Arthroplasty (Suppl 2) 2007b; 22 (6:) 125-9.
  • Bryant D, Havey TC, Roberts R, Guyatt G. How many patients? How many limbs? Analysis of patients or limbs in the orthopaedic literature: a systematic review. J Bone Joint Surg (Am) 2006; 88 (1): 41-5.
  • Callary SA, Field JR, Campbell DG. Low wear of a second-generation highly crosslinked polyethylene liner: a 5-year radiostereometric analysis study. Clin Orthop 2013; (471) (11): 3596-600.
  • Calvert GT, Devane PA, Fielden J, Adams K, Horne JG. A double-blind, prospective, randomized controlled trial comparing highly cross-linked and conventional polyethylene in primary total hip arthroplasty. J Arthroplasty 2009; 24 (4): 505-10.
  • Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Van Citters DW. Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups. J Bone Joint Surg (Am) 2007; 89 (9): 2023-9.
  • D’Antonio JA, Capello WN, Ramakrishnan R. Second-generation annealed highly cross-linked polyethylene exhibits low wear. Clin Orthop 2012; (470) (6): 1696-704.
  • Derbyshire B, Prescott RJ, Porter ML. Notes on the use and interpretation of radiostereometric analysis. Acta Orthop 2009; 80 (1): 124-30.
  • Digas G, Karrholm J, Thanner J, Herberts P. 5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis. Acta Orthop 2007; 78 (6): 746-54.
  • Dumbleton JH, D’Antonio JA, Manley MT, Capello WN, Wang A. The basis for a second-generation highly cross-linked UHMWPE. Clin Orthop 2006; (453): 265-71.
  • . Furmanski J, Anderson M, Bal S, Greenwald AS, Halley D, Penenberg B, et al Clinical fracture of cross-linked UHMWPE acetabular liners. Biomaterials 2009; 30 (29): 5572-82.
  • Furmanski J, Kraay MJ, Rimnac CM. Crack initiation in retrieved cross-linked highly cross-linked ultrahigh-molecular-weight polyethylene acetabular liners: an investigation of 9 cases. J Arthroplasty 2011; 26 (5): 796-801.
  • Glyn-Jones S, Isaac S, Hauptfleisch J, McLardy-Smith P, Murray DW, Gill HS. Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty? A double-blind, randomized, and controlled trial using roentgen stereophotogrammetric analysis. J Arthroplasty 2008a; 23 (3): 337-43.
  • Glyn-Jones S, McLardy-Smith P, Gill HS, Murray DW. The creep and wear of highly cross-linked polyethylene: a three-year randomised, controlled trial using radiostereometric analysis. J Bone Joint Surg (Br) 2008b; 90 (5): 556-61.
  • Kesteris U, Ilchmann T, Wingstrand H, Onnerfalt R. Polyethylene wear in Scanhip arthroplasty with a 22 or 32 mm head: 62 matched patients followed for 7-9 years. Acta Orthop Scand 1996; 67 (2): 125-7.
  • Lachiewicz PF, Heckman DS, Soileau ES, Mangla J, Martell JM. Femoral head size and wear of highly cross-linked polyethylene at 5 to 8 years. Clin Orthop 2009; (467) (12): 3290-6.
  • McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty. A randomized controlled trial. J Bone Joint Surg (Am) 2009; 91 (4): 773-82.
  • Muratoglu OK, Bragdon CR, O’Connor D, Perinchief RS, Estok DM, 2nd, Jasty M, et al. Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept. J Arthroplasty (Suppl 1) 2001; 16 (8): 24-30.
  • Rohrl SM, Li MG, Nilsson KG, Nivbrant B. Very low wear of non-remelted highly cross-linked polyethylene cups: an RSA study lasting up to 6 years. Acta Orthop 2007; 78 (6): 739-45.
  • Tarasevicius S, Robertsson O, Kesteris U, Kalesinskas RJ, Wingstrand H. Effect of femoral head size on polyethylene wear and synovitis after total hip arthroplasty: a sonographic and radiographic study of 39 patients. Acta Orthop 2008; 79 (4): 489-93.
  • Wang A, Manley M, Serekian P, Kurtz S, Gsell R, Martell J. Wear and structural fatigue simulation of crosslinked ultra-high molecular weight polyethylene hip and knee bearing applications—Crosslinked and thermally treated ultra-high molecular weight polyethylene for joint replacements, ASTM STP1445. ASTM International, West Conshohecken, PA, 2003: 151-68.