Ear Oximeter Calibration Using Alveolar Po2Measurements

, &
Pages 37-43 | Received 28 May 1971, Accepted 12 Nov 1971, Published online: 08 Jul 2009
 

Abstract

An indirect method of calibrating an ear oximeter using alveolar Po2measurements has been tested in 10 normal subjects. An oximeter calibration curve was constructed for each individual subject by comparing the output of an ear oximeter with arterial oxygen saturation calculated from alveolar Po2and PCO2during 10 to 20 seconds of hypoxia produced by breathing low oxygen mixtures in the supine position. Hypoxia is terminated by having the subject breathe oxygen, and it is assumed that arterial saturation reaches 100% and that discrepancies between end-expired and arterial gas tensions due to ventilation-perfusion imbalance, impaired diffusion, and right-to-left shunts are negligible under the conditions of calibration. The accuracy of the indirect calibration method was tested by comparing it with direct calibration using arterial blood samples. It is concluded that the indirect calibration method gives a reasonably accurate measure of arterial oxygen saturation in the range of 25–100%.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 200.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.