Microbiological monitoring of ultrapure dialysis fluid in a hemodialysis center in Alexandria, EgyptFootnoteFootnote

, , , &
Pages 523-527 | Received 13 Feb 2018, Accepted 27 Apr 2018, Published online: 17 May 2019

1 Introduction

Microbiological contamination of dialysate is an issue that attracted attention since the 1990s. This is partly due to the extensive use of sodium bicarbonate as a dialysate buffer, which helps microbial proliferation, and also the use of high flux dialysers that allow back filtration of microbial contaminants.Citation1 Microbial contaminants most frequently found in dialysis water are bacteria and their degradation products, such as endotoxins, peptidoglycans and bacterial DNA (bDNA). Fungi, viruses and protozoa can also be encountered occasionally.Citation2 Endotoxins are heat-stable lipopolysaccharides (LPS) and the major cell wall components of Gram-negative bacteria.Citation3 The molecular mass of LPS ranges between 2000 and 20,000 Da. LPS can be transferred through membranes with large pore size by back filtration/diffusion from the dialysis fluid to the blood compartment.Citation4 If the dialyser membrane is tight, bacteria are not able to pass from the dialysate compartment into the blood compartment. However, endotoxins, can permeate and cause an inflammatory state.Citation5

Exposure to microbial contaminants is clearly associated with short-term complications, ranging from pyrogenic reactions to septicaemia.Citation6,Citation7 Furthermore, microbial contamination of dialysis fluid is an important cause of chronic inflammation and MIA syndrome (malnutrition, inflammation, and atherosclerosis) among end-stage renal disease patients. Long-term HD patients suffer from arteriosclerosis, and β2-microglobulin-associated amyloidosis with complicating spinal stenosis, carpal tunnel syndrome, bone cysts and arthropathy. To overcome the above- mentioned short and long-term complications of microbiologically contaminated dialysis fluid, the use of ultrapure dialysis fluid (UPDF) which has a lower microbial load was recommended. This has been accompanied by an improvement in the inflammatory state of HD patient.Citation8

The Association for the Advancement of Medical Instrumentation (AAMI) and the European Pharmacopoeia (EP) established upper limits for microbial contamination of UPDF as containing <0.1 colony forming unit (CFU)/ml and <0.03 endotoxin unit (EU)/ml using LAL assay.Citation9 To produce highly purified (ultrapure) water, a system is required based on having a second reverse osmosis (RO) module and/or an electrochemical deionizer (DI) placed in series.Citation10 There are no international standards for fungi in dialysis fluids, however Fresenius Medical Care (FMC) recommended the level of fungi to be <10 CFU/ml in dialysis water and dialysis fluid, and <1 CFU/ml in UPDF.Citation11

2 Aim of the work

This study aimed at evaluating microbial contamination of dialysate and UPDF through estimation of heterotrophic plate count, total coliforms and fungi as well as endotoxin detection.

3 Material and methods

This cross-sectional study was carried out during a 6-months period, on a total of 100 dialysis fluid samples randomly collected from a private HD center in Alexandria, Egypt. The HD unit consists of one water treatment unit and two rooms for HD. In this unit, water passes through a RO filter and a mixed bed DI, and then it is stored in a tank of 1000 Liter capacity. The finally treated water is pumped into two rooms of HD through a system of polyvinyl chloride (PVC) pipes to reach every HD machine. Chlorination is done every month for maintenance of water treatment unit and the bacterial filter (0.2 µm) is substituted every month. The total number of HD machines in both HD rooms is eight machines. The first room is back to back to the water treatment unit and contains three machines. On the other hand, there is a long distance (about 20 m) between the water treatment unit and the second room which contains five machines. Most of the machines were commercial type A, the remaining were type B which is relatively newer version with some technological advances. The dialysate concentrate used in this unit is a sealed sterile capsulated bicarbonate powder along with the complementary solution A for bicarbonate based HD (referred to as A-component). Chemical disinfection of machines is done after every session using 0.2% peracetic acid solution 30 min at room temperature.

3.1 Sampling

A total of eighty UPDF (infusate) samples were collected from all online hemodiafiltration (HDF) sessions. In addition, twenty pure dialysis fluid samples (distributed as 10 dialysis water and 10 final dialysate samples) were collected on weekly basis. Samples were aseptically collected in sterile wide-mouthed glass containers (1000 ml containers for infusate samples, 900 ml each, and 500 ml containers for dialysis fluid samples, 300 ml each).Citation12

Thirty UPDF samples were collected from room (1) and 50 samples from room (2), out of which, 10 samples were from Type B machine.

Before collecting the final treated water samples (dialysis water), the faucets were wiped with alcohol, left to dry and then water was left to flow for one minute to flush any residuals in the connections and pipes. The final dialysate and the infusate fluid samples were collected from sampling ports by allowing the fluid to flow for half a minute then aseptically collected in glass containers. The samples for endotoxin analysis were collected aseptically by one ml depyrogenated sterile disposable syringes.

All the collected samples were labeled, transported and delivered to the laboratory as soon as possible to be analyzed within 1–2 h of collection. In case of expected delay, samples were stored at 4 °C and analyzed within 24 h maximum. Each sample was accompanied by a sheet including: sampling date, time of collection, sampling site and type of machine. All samples (dialysis water, final dialysate and infusate) were subjected to membrane filtration method, the membranes were plated onto the following media: Reasoner's 2A (R2A) agar (Oxoid CM0906), Eosine Methylene Blue agar (EMB) (Levine CM0069) and Sabouraud dextrose agar (SDA) (Oxoid CM0041) supplemented with 50 mg chloramphenicol/liter). R2A, a low nutrient medium used for the recovery of stressed heterotrophic bacteria found in high purity water, was used as recommended for HPC, and EMB was used for detection of total coliforms while SDA was used for fungal cultivation. R2A and SDA plates were incubated at room temperature (25 °C) for at least 5 days and up to 7 days, while EMB plates were incubated at 37 °C for 24 h. After the incubation period, colonies were counted and reported as CFU per ml of sample. LAL assay for endotoxin measurement was done by gel clot method, with Pyrotell (Associates of Cape Cod, Inc.) of sensitivities 0.25 EU/ml and 0.03 EU/ml for pure and ultrapure water samples respectively.

During the sessions in which samples were collected observation of any febrile/ pyrogenic reactions in patients were recorded.

3.2 Statistical analysis

Data were fed to the computer and analyzed using IBM SPSS software package version 20.0. Qualitative data were described using number and percent. Quantitative data were described using range (minimum and maximum), mean, standard deviation and median. Significance of the obtained results was judged at the 5% level.

4 Results

During the time of sample collection, none of the HD patients encountered any febrile/pyrogenic reaction.

According to the European standards for HD fluids and Fresenius Medical Care (FMC) guidelines Citation11,Citation13 the final treated water and the final dialysate were considered unacceptable if one or more of the following criteria were found: HPC > 100 CFU/ml, presence of coliforms, endotoxin level > 0.25 EU/ml and fungi > 10 CFU/ml. On the other hand, the infusate was considered unacceptable if one or more of the following criteria were found: HPC > 0.1 CFU/ml, presence of coliforms, endotoxin level > 0.03 EU/ml or fungi > 1 CFU/ml.

According to the above standards, TC and fungi parameters were found acceptable in all samples (100.0% each). Out of 80 infusate samples, 63 samples (78.8%) were acceptable when examined for HPC alone. The percentage of acceptability decreased to 35.0% when examined by LAL alone and further decreased to 28.8% when collectively evaluated by all parameters ().

Table 1 Distribution of the acceptability of the 100 HD fluids samples according to the examined microbiological parameters.

The mean count of HPC was recorded for the infusate samples as 0.09 ± 0.11 CFU/ml, and for dialysate and dialysis water samples as (16.8 ± 2.39, and 23.0 ± 2.54 CFU/ml, respectively). Nearly equal mean counts of fungi were present in infusate, dialysate and dialysis water samples (0.05 ± 0.09, 0.04 ± 0.02 and 0.05 ± 0.04 CFU/ml respectively), and all were within acceptable levels ().

Concerning the two types of HD machines included in the study, the mean count of fungi of infusate samples from Type A machines was higher than that from Type B (0.05 ± 0.10, 0.02 ± 0.01 CFU/ml) respectively. However, the difference was statistically insignificant (p = 0.329). On the other hand, infusate samples from Type A machines had significantly higher mean HPC counts than those from Type B (0.09 ± 0.12, 0.02 ± 0.02 CFU/ml) respectively (z = -3.60, p = 0.000).

Out of 70 samples from Type A machines, 18 (25.7%) were acceptable and 52 (74.3%) were unacceptable for endotoxin by LAL testing. On the other hand, all the 10 samples from Type B machine were acceptable by LAL, and this difference between both machines as regards LAL testing was found to be statistically significant (p = 0.000) .

Table 2 Distribution of infusate samples according to HD machine type and LAL acceptability.

Concerning the 2 rooms of HD, infusate samples from Type A machines were compared in both rooms. There was an insignificant difference between Room 1 and Room 2 as regards HPC and fungal counts (HPC: 0.08 ± 0.06 CFU/ml and 0.11 ± 0.14 CFU/ml respectively, fungal count: 0.07 ± 0.13 CFU/ml and 0.04 ± 0.08 CFU/ml respectively). However, Room 1 showed higher percentage of acceptable infusate samples from Type A machines by LAL test (53.3%) than Room 2 (5%) with statistical significance (χ2 = 20.96, p = 0.000).

About 80% of samples from Type A machines with a total working hours <5000 or 5000–10,000 were acceptable for HPC compared to 66.7% of samples from machines with >10,000 h. However, this difference was statistically insignificant. On the other hand, when examined by LAL, the percentages of acceptability decreased to 51.7% and 22.5% for the machines with total working hours <5000 and 5000–10,000 respectively, compared to 11.1% for those with >10,000 h. This difference was statistically significant ((χ 2 = 8.6, p = 0.013) ().

Table 3 Acceptability of samples from Type A machines according to HPC and LAL parameters and its relation to the total working hours of dialysis machines.

Out of the 80 infusate samples, 63 samples were acceptable by HPC, and only 23 of these were also acceptable by LAL (). The observed agreement between both parameters (HPC and LAL) was 44%.

Table 4 The relationship between LAL and HPC parameters among the 80 examined infusate samples.

5 Discussion

The quality of the dialysate is of paramount importance in assuring patient safety. To help prevent injury to HD patients from contaminants in the dialysate, standards have been developed for the quality of the dialysate as well as the water concentrates and devices used in its preparation Citation13.

Several studies reported the presence of microbiological contaminants in dialysate Citation14Citation18; however, data on ultrapure contamination are infrequently available. In the present study, although the acceptability for infusate was poor regarding both HPC and LAL assay parameters, the respective acceptability for treated water as well as dialysate was almost 100.0%. This can be explained by the more astringent standard for the infusate. This high standard was not feasible to be accomplished by the older machine technology (type A) which was of less technological advances. Whereas recent advances in machine technology (type B) could compensate for any potential defect in water treatment unit raising the acceptability of the infusate again to 100.0%.

In this study, the observed discrepancy between LAL assay and HPC parameter agrees with reports of Klein et al.Citation14 Kulander et al.Citation15 and Bland et al.Citation16 who found no correlation between bacterial growth and endotoxin concentrations in water or dialysate samples. They proposed that low levels of bacterial growth in the dialysate might be associated with high endotoxin concentration, since endotoxin is released mainly after death of microorganisms. Ledebo and Nystard recommended that endotoxin levels should always be measured in addition to bacterial counts because they give a different yet complementary picture of the microbiological quality of the fluid.Citation17

Regarding final dialysate, the present study revealed 100.0% acceptability for HPC, while El-Koraie et al. reported lower compliance results (51.7% and 86.1% acceptability at two HD units in Alexandria).Citation18 The AAMI/ISO standards recommend culturing conditions that more closely resemble the conditions under which contaminating bacteria exist in the systems used to prepare dialysate. These new culturing conditions have been shown to increase the yield of bacteria obtained for a given water or dialysate sample, thus providing better information on the level of contamination in the system.Citation9,Citation19 The above mentioned culturing conditions have been implemented in the current study. In addition, the MF method is also recommended by the AAMI/ISO standards and has been implemented in this study for detecting low levels of contamination.

There are few reports concerning the extent of fungal contamination of dialysis fluid. Figel et al.Citation20 isolated black fungi from the municipal supply and in treated water for dialysis and dialysate. Most of the international standards do not include fungi as indicators of quality for dialysis, except in Sweden, where the current standards for fungi are 10 CFU/ml.Citation20 In this piece of work, no fungal elements above the standard level were recorded.

In this study, TC were not isolated in any of the samples, where dialysate concentrate used in this unit is a sealed sterile capsulated bicarbonate powder (ready-made Bicart capsules). However, in a previous study conducted in Alexandria, TC were isolated from 4% of examined samples, where reconstituted bicarbonate powder was used, which is more liable for contamination. In that same study, one sample was contaminated with E. coli, but neither TC nor E. coli were revealed from the unit that was using ready-made Bicart capsules.Citation18 Again, Zunino et al.Citation21 encountered TC in 0.5% of samples whereas Arvanitidou et al.Citation22 reported much higher results: 12.3% for TC and 8.6% for E. coli.

Water purification by a double RO module is recommended, because microbiological water quality after purification with a single RO does not often comply with the reference quality level.Citation2,Citation4 The presence of a treated water-storage tank, as in the current study, which may not be frequently drained and disinfected, is a recognized factor contributing to persistent bacterial contamination and to biofilm formation.Citation6,Citation9 Favero et al. stated that storage tanks should not be used unless provision were made for frequent draining and adequate disinfection.Citation23 Concerning the distribution system, high grade stainless steel is the only material resistant to biological corrosion.Citation10 However, PVC is usually used because of its low cost, as in this studied dialysis unit.

In the present study, infusate samples were collected from two dialysis rooms, the first (room 1) is back to back to the water treatment unit, while for the second room, long distribution tubes were encountered (about 20 m long). This could explain the reason for the much higher percentage of acceptable LAL infusate samples at room 1 than room 2 (53.3% and 5.0% respectively). In addition, HPC mean counts of infusate samples collected from dialysis machines at room 2 were higher than those at room 1 (0.11 and 0.08 CFU/ml respectively), although did not reach a significant level. These data indicate that dialysis fluid pathways should be configured in a loop, be as short as possible, have no dead-ends, and contain the minimum number of pipe fittings and joints.Citation9,Citation19 Otherwise, biofilm formation is frequently encountered in such points, allowing the shedding of bacteria in low counts and the continuous release of their degradation products, as the endotoxins.

Regarding the finally treated water, the present study revealed a 100.0% HPC acceptability. There was a significant association between the quality of final dialysate and that of treated water. Similarly, a study conducted by El-Koraie et al.Citation18 showed an association between the final dialysate and each treated water and concentrates in the two examined HD units. In contrast, Baumbauer et al.Citation24 found no correlation between the level of contamination of dialysate and the water processing method or type of concentrate.

Arvantidou et al.Citation22 suggested that the increasing working hours of dialysis machines correlated more with the quality of dialysate, as it becomes more difficult to clean or disinfect. In accordance with those workers, the present study has shown significant decrease of acceptability for LAL parameter with the increase of the total working hours of the Type A machines.

Pertaining to HD machines, this study demonstrated significant differences between Type A and Type B machines in the mean counts of HPC and in LAL test results of infusate samples. This might be related to the fact that Type B machine is of newer generation which incorporates recent technology regarding continuous flow of dialysis fluid, the heating system and avoidance of closed circuits and loops. Of note, that the back filter in the Type B machine is a double filter bath (diasafe), whereas in the Type A machine it is only single (foreclean filter). New machines have incorporated several recommended features including low volumes of hydraulic circuit, high dialysate rates, global disinfection and non stagnant zones.Citation10

Fortunately, despite the detection of endotoxin levels that were more than the permissible one (0.03 EU/ml), there were no encountering of any pyrogenic reactions among patients on HDF sessions. This could be considered as potential limitation, as the main aim of this study did not address clinical signs for endotoxemia and inflammatory reactions, which can be further investigated in future research. However, this issue was previously tackled in assessing endotoxin level in HD patients and its role in inducing inflammation.Citation25 Di Iorio et al. reported after a one year study of using ultrafilter, C-reactive protein and pro-inflammatory cytokines decreased while anti-inflammatory ones increased, and hemoglobin levels were improved. They speculated that ultrapure water can ameliorate the inflammatory status and improve survival of dialysis patients.Citation26

The present study has shown 44% agreement between HPC and LAL parameters performed to assess the quality of the infusate. The same percentage of agreement was reported by El-Koraie et al. in Alexandria.Citation18 On further analysis of our findings, HPC showed 21.2% unacceptability, which increased to 65.0% and 71% by LAL and by both HPC and LAL respectively. In accordance with these findings, El-Koraie et al. revealed that HPC showed 40.0% unacceptability, which increased to 88.0% and 92.0% by LAL and by both HPC and LAL respectively.Citation18

6 Conclusion

This study highlighted the potential risk and safety concern that might be associated with the use of online HDF if to be generalized without implementing the international microbiological standards for ultrapure dialysate quality. Again, the ability to reach a 100.0% acceptability of the infusate using the more advanced newer HDF machines can raise the hope of implementing such technique with a high safety margin. It is also evident that long distribution tubes in the HD unit, as well as long working hours of HD machines, are of paramount importance in determining the quality of infusate used in online HDF.

Conflicts of interest

All authors declare that they have no conflicts of interest in relation to the present manuscript.

Acknowledgements

The authors take full responsibility for the content of the article.

Declaration of interest

None.

Funding source

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Notes

Peer review under responsibility of Alexandria University Faculty of Medicine.

Available online 18 May 2018

References

  • S.M.GordonC.W.OettingerL.A.BlandPyrogenic reactions in patients receiving conventional, high-efficiency, or high-flux hemodialysis treatments with bicarbonate dialysate containing high concentrations of bacteria and entotoxinJ Am Soc Nephrol2199214361444
  • G.J.HandelmanP.A.MegdalS.K.HandelmanBacterial DNA in Water and Dialysate: Detection and Significance for Patient OutcomesBlood Purif27120098185
  • M.CaroffD.KaribianJ.M.CavaillonN.Haeffner-CavaillonStructural and functional analyses of bacterial lipopolysaccharidesMicrobes Infect42002915926
  • R.VanholderE.VanhaeckeN.VeysEndotoxin transfer through dialysis membranes—small-pore versus large-pore membranesNephrol Dial Transplant71992333339
  • C.TielemansC.HussonT.SchurmansEffects of ultrapure and non-sterile dialysate on the inflammatory response during in vitro hemodialysisKidney Int49236199643
  • P.BrunetY.BerlandWater quality and complications of haemodialysisNephrol Dial Transplant152000578580
  • C.TielemansN.A.HoenichN.W.LevinG.LonnemannM.S.FaveroH.SchifflAre standards for dialysate purity in hemodialysis insufficiently strict?Semin Dial142001328336
  • I.LedeboUltrapure dialysis fluid – how pure is it and do we need it?Nephrol Dial Transplant2220072023
  • Association for the Advancement of Medical Instrumentation (AAMI) [homepage on the Internet]. Water treatment equipment for hemodialysis applications. Arlington, VA, USA; 2000: 2–32. <http://dialysiswatersolution.com/regulations-and%20guidelines/ansiaami%20/rd-62-water-for-dialysis/>.
  • R.Perez-GarciaP.O.C.Rodriguez-BenitezWhy and how to monitor bacterial contamination of dialysate?Nephrol Dial Transplant152000760764
  • Fresenius Medical Care [homepage on the Internet]. Quality of water for haemodialysis and dialysis fluids; 2001. <http://www.vision-fmc.com/download/C-CG1001.pdf>.
  • E.L.PenneL.VisserM.A.van den DorpelMicrobiological quality and quality control of purified water and ultrapure dialysis fluids for online hemodiafiltration in routine clinical practiceKidney Int7662009665672
  • European Directorate for the Quality of Medicines. Purified water. European Pharmacopoeia 6.3, Strasbourg; 2009. p. 4344–6.
  • E.KleinT.PassG.B.HardingR.WrightC.MillionMicrobial and endotoxin contamination in water and dialysate in central United StatesArtif Organs1419908594
  • L.KulanderU.NisbethB.G.DanielssonO.ErikssonOccurrence of endotoxin in dialysis fluid from 39 dialysis unitsJ Hosp Infect2419932937
  • L.A.BlandJ.C.OliverM.J.ArduinoC.W.OettingerS.K.McAllisterM.S.FaveroPotency of endotoxin from bicarbonate dialysate compared with endotoxin from Escherichia coli and Shigella flexneriJ Am Soc Nephrol5199416341637
  • I.LedeboR.NystrandDefining the microbiological quality of dialysis fluidArtif Organs2319993743
  • A.F.El-KoraieW.A.HazzahA.A.AbbassS.A.El-ShazlyBacteriological monitoring of dialysis fluid in 2 hemodialysis units in AlexandriaEgypt. Saudi Med J288200712341238
  • Association for the Advancement of Medical Instrumentation. Quality of dialysis fluid for haemodialysis and related therapies. ANSI/AAMI/ISO 11663; 2009.
  • I.C.FigelP.R.DalzotoI.C.PimentelMicrobiological quality of water and dialysate from haemodialysis units in Southern BrazilRev Inst Adolfo Lutz São Paulo74120156670
  • P.ZuninoL.BeltránL.ZunioMicrobiological quality of hemodialysis water in a three-year multicenter study in UruguayJ Nephrol152002374379
  • M.ArvanitidouS.SpaiaN.AsklepidisEndotoxin concentration in treated water of all hemodialysis units in Greece and inquisition of influencing factorsJ Nephrol12119993237
  • M.S.FaveroN.J.PetersenL.A.CarsonW.W.BondS.H.HindmanGram negative water bacteria in hemodialysis systemsHealth Lab Sci121975321334
  • R.BaumbauerM.SchauerW.K.JungV.DaumJ.VienkenContamination of dialysis water and dialysate. A survey of 30 centersASAIO J40199410121016
  • A.El-KoraieY.NagaA.SaaranN.FarahatW.A.HazzahEndotoxins and inflammation in hemodialysis patientsHemodial Int172013
  • B.Di IorioL.Di MiccoD.BruzzeseUltrapure dialysis water obtained with additional ultrafilter may reduce inflammation in patients on hemodialysisJ Nephrol3062017795801