Wear of ultra-high molecular weight polyethylene and polytetrafluoroethylene in a hip simulator: A dose-response study of protein concentration

, , , , &
Pages 365-369 | Published online: 08 Jul 2009
 

Abstract

Charnley's laboratory wear studies of non-gamma sterilized polytetrafluoroethylene (PTFE) and polyethylene (PE) found that the PTFE to PE wearrate ratio of 250:1 was much higher than the in-vivo wear ratio of 20:1. Tests of PTFE and PE in our laboratory showed a wear ratio of 150:1, using bovine serum as the lubricant and 190:1 with water as the lubricant. Our hypothesis was that the wear-rates of PTFE and PE cup materials were related to the concentration of protein in the serum. We studied the wear behavior of PTFE and PE cups in varied protein concentrations, using 4 femoral head sizes to validate the clinical range reported by Charnley. The PTFE wear-rates increased with increasing protein concentration and conversely, PE wear-rates decreased with increasing protein concentration. This inverse relationship made it possible to bring the wear ratio closer to the desired clinical wear ratio. We found that the clinically relevant PTFE/PE wear ratio corresponded to 3-10 mg/mL of protein in bovine serum.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.