Dynamic intracochlear pressure measurement during cochlear implant electrode insertion

, , , &
Pages 860-865 | Received 20 Apr 2019, Accepted 30 Jun 2019, Published online: 12 Jul 2019
 

Abstract

Background: Electrode insertion into the cochlea can cause significant pressure changes inside the cochlea with assumed effects on the cochlea’s functionality regarding residual hearing. Model-based intracochlear pressure (ICP) changes were performed statically at the cochlear helix.

Aims/objectives: The aim of this study was to observe dynamic pressure measurements during electrode insertion directly at the cochlear implant electrode.

Material and methods: The experiments were performed in an uncurled cochlear model that contained a volume value equivalent to a full cochlea. A microfibre pressure sensor was attached at one of two positions on a cochlear implant electrode and inserted under different insertional conditions.

Results: We observed the ICP increase depending on the insertional depth. A sensor-position-specific pressure change is insertional-depth dependent. Interval insertion did not lead to a lower peak insertional ICP.

Conclusions and significance: In contrast to the static pressure-sensor measurement in the artificial model’s helix, a dynamic measurement directly at the electrode shows the pressure profile to increase based on the insertional depth. A mechanical traumatic relevance of the observed pressure values cannot be fully excluded.

Chinese abstract

背景:电极插入耳蜗可引起耳蜗内压力的显著变化, 并对残余听力方面的耳蜗功能产生了所设想的影响。在耳蜗螺旋线处静态进行基于模型的耳蜗内压(ICP)变化。

目的:本研究的目的是观察电极直接插入耳蜗植入电极时的动态压力测数。

材料和方法:实验是在一个未弯曲的耳蜗模型中进行的, 该模型的体积与全耳蜗的相当。将微纤维压力传感器安装在人工耳蜗植入电极的两个位置之一, 并在不同的插入条件下插入。

结果:随着插入深度的增加, 耳蜗内压升高。传感器位置特定的压力变化取决于插入深度。间隔插入并没有导致较低的峰值插入ICP。

结论与意义:与人工模型螺旋线中的静态压力传感器测量相反, 直接在电极上进行的动态测量显示, 根据插入深度, 压力会增加。不能完全排除所观察到的压力值可能与机械创伤相关。

Disclosure statement

No potential conflict of interest was reported by the authors.

Additional information

Funding

This study was supported by Advanced Bionics, Stäfa, Switzerland.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.